磁技术钬激光体外评估:脉冲、气泡与光纤安全性
体外研究评估磁技术Ho:YAG激光的脉冲波形、气泡特性、功率衰减及光纤断裂风险,使用150 W发生器及200/365/550 μm光纤。结果显示准矩形脉冲、峰值功率随功率增加,生理盐水中初始功率损失13-28%,测试条件下无光纤断裂。
要点
- 磁技术Ho:YAG激光呈准矩形脉冲,脉宽0.7-1.7 ms,峰值功率513.4-1149.7 W。
- 550 μm光纤气泡更宽,200 μm光纤气泡持续时间更长。
- 生理盐水中初始功率损失13-28%,接触BegoStone时更明显。
核心结论
- 磁技术Ho:YAG激光因脉冲轮廓峰值功率受限,在传统Ho:YAG与铥光纤激光之间呈现有前景的折中。
研究设计与样本
- 体外实验使用150 W Ho:YAG发生器(磁技术),分别用200和550 μm光纤测脉冲,200 μm光纤高速摄像评估气泡,200/365/550 μm光纤测功率损失,200 μm光纤测光纤断裂风险。
主要发现
- 脉冲波形为准矩形,脉宽0.7-1.7 ms,平均(标准差)峰值功率513.4(23.3)至1149.7(6.9)W,随功率成比例增加,但每个脉冲内有多个尖峰。
- 550 μm光纤的气泡显著宽于200 μm光纤,而200 μm光纤的气泡持续时间显著更长。
- 生理盐水中所有光纤直径初始功率损失13-28.0%,连续激活至1分钟时功率损失进展,接触BegoStone时功率衰减更明显。
- 在测试的实验条件下(不同激光设置和光纤弯曲半径0.45、0.6、0.9 cm,连续激活5分钟或至断裂),未观察到光纤断裂。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To evaluate the following characteristics of the holmium:yttrium-aluminium-garnet (Ho:YAG) laser with Magneto technology: pulse shape, shape and duration of the laser-induced vapour bubble (VB), power loss over time, and the risk of fibre fracture (RFF). METHODS: A 150 W Ho:YAG generator was characterised in vitro with Magneto technology. The pulse shape was measured using an oscilloscope. The pulse duration was measured to calculate the peak power (PP) using both a 200- and 550-μm laser fibre (LF). VB were evaluated with a 200-μm fibre using a high-speed camera (8000 frames/s). Power loss over time was obtained at the fibre tip before and after 30 s, then 1 min of continuous laser activation in saline solution and in contact with BegoStone® (BEGO GmbH & Co., Bremen, Germany). Power loss was evaluated using LFs with core diameters of 200, 365, and 550 μm. RFF was tested after 5 min of continuous laser activation or until LF fracture with different laser settings and LF bending radii (0.45, 0.6, 0.9 cm) using a 200-μm LF. RESULTS: The Ho:YAG laser with the Magneto technology possesses a quasi-rectangular waveform with extended pulse durations ranging between 0.7 and 1.7 ms and limited PPs between a mean (standard deviation) of 513.4 (23.3) to 1149.7 (6.9) W that increase proportionally with power. However, multiple spikes were observed within each pulse, precluding a perfectly linear plateau. Regarding VB, it was significantly wider with the 550-μm LF compared to the 200-μm LF. Conversely, the VB duration was significantly longer with the 200-μm LF. A consistent initial power loss was observed in normal saline of 13-28.0% across all LF diameters, which progressed during continuous activation of up to 1 min. This power degradation was more pronounced on synthetic BegoStone. No LF fractures were observed under the tested experimental conditions. CONCLUSION: The new Ho:YAG laser with Magneto technology appears to be a promising compromise between conventional Ho:YAG and thulium fibre laser, owing to its pulse profile with a limited PP.
引用
- Martin-Illac A, Chicaud M, Letouche ML, et al. In vitro evaluation of the new holmium:yttrium-aluminium-garnet laser with Magneto technology on behalf of the Progress and Collaboration in Endourology (PACE) group. BJU Int. 2026. PMID: 42610757. doi:10.1111/bju.70425
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。