超脉冲铥光纤激光与脉冲调制钬激光在软性输尿管镜中的前瞻性随机对照试验
两组零结石率无显著差异(51.3% vs 53.6%,P=1.00),但TFL光纤降解率更低(2.8% vs 21.3%,P=.02),术者满意度及噪声负担均更优。
要点
- TFL与HoYAG的零结石率相当(51.3% vs 53.6%,P=1.00)。
- TFL光纤降解率显著低于HoYAG(2.8% vs 21.3%,P=.02)。
- TFL在操控性、可操作性及粉末化性能上术者满意度更优(均P<.001)。
核心结论
- TFL与高功率脉冲调制HoYAG的结石清除率等效,但TFL在光纤耐用性、术者满意度和术中噪声方面具有明显优势,支持TFL作为软性输尿管镜优选平台。
主要发现
- 零结石率两组相当(51.3% vs 53.6%,P=1.00;<2 mm:66.7% vs 85.7%,P=.14;<4 mm:82.1% vs 89.3%,P=.50)。
- TFL光纤降解率显著低于HoYAG(2.8% vs 21.3%,P=.02)。
- 术者满意度在操控性、可操作性及粉末化性能上均 favoring TFL(所有P<.001)。
- TFL在所有噪声负担测量指标上均报告更低(P<.001)。
关键数值与推荐
- 主要终点为术后CT零结石率,组间无显著差异(51.3% vs 53.6%,P=1.00)。
- 光纤降解率TFL组2.8%,HoYAG组21.3%,差异有统计学意义(P=.02)。
- 术者满意度及噪声负担各项比较均P<.001, favoring TFL。
临床怎么用
- 对于<2 cm肾结石行软性输尿管镜碎石,TFL可提供与高功率HoYAG相当的结石清除率,同时减少光纤损耗并改善术者操作体验与噪声环境。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate the clinical performance of the SOLTIVE SuperPulsed thulium fiber laser (TFL) compared to pulse-modulated Holmium:YAG (HoYAG) in flexible ureteroscopy for renal stones <2 cm, with a focus on efficiency, ergonomics, and surgeon experience, given the limited comparative U.S. METHODS: We performed a single-center prospective randomized controlled trial (September, 2022-September, 2024) of patients ≥18 years with stone burden <2 cm undergoing ureteroscopic lithotripsy. Participants were randomized 1:1 to SOLTIVE TFL or pulse-modulated HoYAG (Lumenis Pulse 120 H with MOSES 2.0). The primary endpoint was zero stone-free rate (SFR) on postoperative CT. Secondary endpoints included operative metrics, reintervention, fiber degradation, and surgeon-reported satisfaction and noise encumbrance. Statistical significance was set at P<.05. RESULTS: Eighty-three patients were analyzed (HoYAG: 47; TFL: 36). Groups were similar in demographics, stone characteristics, and operative time. SFRs were comparable (Zero-SFR: 51.3% vs 53.6%, P = 1.00; <2 mm: 66.7% vs 85.7%, P = .14; <4 mm: 82.1% vs 89.3%, P = .50). Fiber degradation was significantly lower with TFL (2.8% vs 21.3%, P = .02). Surgeon satisfaction favored TFL for handling, maneuverability, and dusting performance (all P<.001). TFL also yielded lower reported noise encumbrance across all measures (P<.001). CONCLUSION: While SFR was equivalent, SOLTIVE TFL showed clear benefits in fiber durability, surgeon satisfaction, and intraoperative noise profile compared to high-power pulse-modulated HoYAG. These results support TFL as a preferred flexible ureteroscopy platform. Multicenter trials are needed to confirm generalizability.
引用
- Vergamini LB, Ito W, Whiles BB, et al. Silent Satisfaction and Improved Laser Ergonomics With TFL: A Prospective Randomized Controlled Trial Comparing SuperPulse Thulium Fiber Laser Versus High-power HoYAG With Moses for Flexible Ureteroscopy. Urology. 2026. PMID: 41651229. doi:10.1016/j.urology.2026.01.047