文献速递前列腺增生与排尿功能障碍

人工尿道括约肌压力调节球囊的时间依赖性材料疲劳:力学与微观分析

Time Dependent Material Fatigue of the Artificial Urinary Sphincter Pressure-regulating Balloon: A Mechanical and Microscopic Analysis.
PubMed 文献 · Urology · 2025-12-19

研究评估人工尿道括约肌(AUS)装置年龄对压力调节球囊(PRB)力学与微观性质的影响,对32个既往植入和1个全新的PRB进行拉伸测试,并对7个年龄均匀分布的PRB子集进行应力软化、松弛测试及扫描电镜(SEM)观察。结果显示PRB年龄与刚度呈中度正线性相关(r=0.68,P<.01),与壁厚呈弱负相关(r=-0.38,P=.03),7个标本中4个最老标本表面出现微裂纹,提示PRB在终末失效前已发生渐进性材料降解。

要点

  • PRB年龄与刚度呈中度正相关(r=0.68,P<.01),与壁厚呈弱负相关(r=-0.38,P=.03)。
  • 7个SEM标本中4个最老标本表面出现微裂纹,提示微观结构缺陷。
  • 日内重复使用未见PRB明显弱化,与测试时间或装置年龄无关。

核心结论

  • PRB在终末失效前随年龄发生渐进性材料降解,表现为刚度增加、壁厚减少和微观结构缺陷形成。
  • 年龄相关PRB疲劳可能是装置性能下降的潜在原因,支持对反复压力性尿失禁的症状患者更早评估PRB更换。

研究设计与样本

  • 对32个既往植入和1个全新AUS的PRB进行拉伸测试。
  • 对7个年龄均匀分布的PRB子集进行应力软化、松弛测试及扫描电镜(SEM)观察。
  • 数据分析采用线性回归、Pearson相关和重复测量方差分析。

主要发现

  • AUS装置年龄范围为2个月至17.6年(中位7.2年)。
  • PRB年龄与刚度呈中度正线性相关(r=0.68,P<.01)。
  • PRB壁厚与PRB年龄呈弱负相关(r=-0.38,P=.03)。
  • 日内重复使用未见PRB明显弱化,与测试时间或装置年龄无关。
  • SEM显示7个PRB标本中4个表面存在微裂纹,见于4个最老标本。

临床怎么用

  • 对反复压力性尿失禁的症状患者,应考虑年龄相关PRB疲劳并更早评估PRB更换。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To assesses the influence of artificial urinary sphincter (AUS) device age on pressure-regulating balloon (PRB) mechanical and microscopic properties. The AUS is a fluid-filled hydraulic system comprised of a urethral cuff, mechanical pump, and PRB. Material fatigue prior to terminal failure of the PRB can reduce its ability to pressurize the urethral cuff, leading to recurrent urinary incontinence. METHODS: Tensile testing was performed on PRBs from 32 previously implanted and 1 new AUS. Stress softening and relaxation testing, as well as scanning electron microscopy (SEM), were conducted on a subset of 7 PRBs evenly distributed in age. Data were analyzed using linear regression, Pearson correlation, and repeated measures ANOVA. RESULTS: AUS device age ranged from 2 months to 17.6 years (median 7.2 years). There was a moderate positive linear correlation between PRB age and stiffness (r = 0.68, P <.01). PRB wall thickness displayed a weak negative correlation with PRB age (r = -0.38,P = .03). There was no noticeable weakening of the PRB with repeated use over the course of a day, regardless of the time of testing or device age. After SEM, 4 of the 7 PRB specimens revealed fine surface cracks present in their microstructure, which were evident in the 4 oldest specimens. CONCLUSION: PRBs exhibit progressive material degradation with age prior to terminal failure, including increased stiffness, decreased wall thickness, and formation of microstructural defects. These findings underscore considering age-related PRB fatigue as a potential cause of declining device performance and support earlier evaluation of PRB replacement in symptomatic patients with recurrent stress urinary incontinence.

来源:Urology,查看原文。版权归原作者所有。

引用

  1. Salvino MJ, Schroeder TE, Grimaud LW, et al. Time Dependent Material Fatigue of the Artificial Urinary Sphincter Pressure-regulating Balloon: A Mechanical and Microscopic Analysis. Urology. 2025. PMID: 41423089. doi:10.1016/j.urology.2025.12.031