吸引输尿管通路鞘的安全压力:体外压力向流量的转换

Safe pressures for suction ureteric access sheaths: in vitro conversion of pressures to flow.
PubMed 文献 · BJU Int · 2025-09-25

本体外研究将吸引输尿管通路鞘(sUAS)的负压设置转换为尺寸特异性的操作限值。使用8.7-F输尿管镜和3.9-F工作通道,测量不同器械下灌注入流,以及10/12-、11/13-、12/14-F sUAS在50-300 mmHg负压下的出流,通过线性插值匹配流量。结果显示50-80 mmHg负压对多数sUAS足够,10/12-F需150-300 mmHg。

要点

  • 50-80 mmHg负压对多数sUAS维持流量平衡足够。
  • 10/12-F sUAS因直径小、镜鞘比高需150-300 mmHg负压。
  • 工作通道器械显著减少入流,可能抵消增加灌注压力的效果。

核心结论

  • 低至50-80 mmHg的吸引压力对大多数sUAS病例已足够。
  • 增加灌注压力可对抗过度出流,但引入器械可能抵消此效应。

研究设计与样本

  • 体外实验使用8.7-F输尿管镜(3.9-F工作通道)测量灌注入流和sUAS出流。
  • 出流测量使用10/12-、11/13-、12/14-F sUAS,Neptune 3™吸引,压力50-300 mmHg,间隔30秒。
  • 通过线性插值匹配入流和出流,推导尺寸和器械特异性的保守吸引压力限值。

主要发现

  • 输尿管镜灌注入流随灌溉压力从50 cmH2O增至300 mmHg,范围为30-130 mL/min。
  • 工作通道器械减少入流:1.9-F取石篮平均变化-57.1%(1.6%),270-μm激光纤维-42.1%(2.1%)(vs 无器械)。
  • sUAS负压产生出流从5 mL/min(10/12-F sUAS,50 mmHg吸引)至700 mL/min(12/14-F sUAS,300 mmHg吸引)。
  • 出流与负压呈线性关系(R² 0.94-0.99)。
  • 增加镜鞘比(SSR)减少灌溉出流;完全打开阀门平均减少出流70.8%;40-cm sUAS较50-cm同直径sUAS平均增加出流13.3%。

关键数值与推荐

  • 50-80 mmHg吸引压力对11/13-和12/14-F sUAS维持流量平衡理想。
  • 10/12-F sUAS(SSR 0.87)需150-300 mmHg负压维持流量平衡。

临床怎么用

  • 根据sUAS尺寸选择吸引压力:11/13-和12/14-F可用50-80 mmHg,10/12-F需150-300 mmHg。
  • 使用压力袋增加灌注可能对抗过度出流,但需注意工作通道器械会减少入流。

局限性

  • 本研究为体外实验,结果需临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To convert suction pressure settings into practical, size-specific operating limits for suction ureteric access sheaths (sUAS). MATERIALS AND METHODS: Using an 8.7-F flexible ureteroscope with a standard 3.9-F working channel, we measured irrigation inflow at various pressures with no instrument, 270-μm laser fibre, and 1.9-F basket. Outflow was measured using 10/12-, 11/13-, and 12/14-F sUAS with suction applied via Neptune 3™ (Stryker, Kalamazoo, MI, USA) at 30-s intervals and pressures ranging from 50 to 300 mmHg. Inflow and outflow were matched by linear interpolation deriving conservative size- and instrument-specific suction pressure limits for all evaluated sUAS configurations. RESULTS: Ureteroscope irrigation inflow ranged from 30 to 130 mL/min as irrigation pressure irrigation increased from 50 cmH20 (gravity) to 300 mmHg. Working channel instruments reduced inflow: mean (± 2 × the standard error of the mean) change of -57.1% (1.6%) for the 1.9-F basket and -42.1% (2.1%) for the 270-μm laser fibre vs no instrument. Negative pressure via sUAS generated outflow as slow as 5 mL/min (10/12-F sUAS, 50 mmHg suction) and as fast as 700 mL/min (12/14-F sUAS, 300 mmHg suction). Outflow and negative pressure were linearly related (R2 0.94-0.99). Increased scope-to-sheath ratio (SSR) reduced irrigation outflow. Fully opening the valve significantly reduced outflow by 70.8% on average. The 40-cm sUAS modestly increased outflow by 13.3% on average compared to 50-cm sUAS of same diameter. Suction of 50-80 mmHg was ideal for maintaining flow equilibrium with 11/13- and 12/14-F sUAS. The small diameter and high SSR (0.87) of the 10/12-F sUAS required greater negative pressures (150-300 mmHg) to maintain flow equilibrium. CONCLUSIONS: Suction pressures as low as 50-80 mmHg are sufficient for majority of sUAS cases. Increasing irrigation inflow with pressure bag systems may counter excessive sUAS outflow; however, introducing instruments can negate this effect.

来源:BJU Int,查看原文。版权归原作者所有。

引用

  1. Santucci J, Alberto M, Sathianathen N, et al. Safe pressures for suction ureteric access sheaths: in vitro conversion of pressures to flow. BJU Int. 2025. PMID: 40995799. doi:10.1111/bju.70012

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。