文献速递泌尿系结石

直接镜内吸引与可弯曲可导航吸引鞘的体外结石清除实验

Experimental stone clearance with in-scope suction and flexible and navigable suction access sheaths.
PubMed 文献 · BJU Int · 2025-07-29

本研究通过台式实验评估直接镜内吸引(DISS)输尿管镜(3.6-Ch和5.1-Ch工作通道)与可弯曲可导航吸引(FANS)输尿管通路鞘对直径<63至2000 μm结石颗粒的清除能力。结果显示3.6-Ch DISS可清除≤250 μm颗粒,5.1-Ch DISS可清除≤500 μm颗粒,而FANS可清除所有测试尺寸(≤2000 μm)的结石颗粒。

要点

  • 3.6-Ch工作通道DISS可清除≤250 μm结石颗粒
  • 5.1-Ch工作通道DISS可清除≤500 μm结石颗粒
  • FANS可清除所有测试尺寸(≤2000 μm)的结石颗粒

核心结论

  • DISS可吸除影响视野的粉尘颗粒(<250 μm),但较大颗粒可能阻塞输尿管镜。
  • FANS能成功清除较大结石碎片,可能有助于提高清石效率。

研究设计与样本

  • 台式实验,使用DISS输尿管镜(Pusen Medical,3.6-Ch和5.1-Ch工作通道)和FANS输尿管通路鞘(ClearPetra;Wellead,11/13-Ch)。
  • 实验组一使用3.6-Ch或5.1-Ch工作通道DISS(配200 μm激光)吸引结石;实验组二使用空5.1-Ch工作通道DISS或11/13-Ch FANS配9.5-Ch输尿管镜(间歇回撤)吸引结石。
  • 每组依次使用逐级增大粒径的结石混合物(<63、63-125、125-250、250-500、500-1000、1000-2000 μm),直至吸引无法完全清除;结石混合物在30分钟内分次引入。

主要发现

  • 3.6-Ch工作通道DISS(含激光纤维)可清除至250 μm的结石。
  • 5.1-Ch工作通道DISS无论是否置入激光纤维均可清除至500 μm的结石。
  • FANS可清除所有测试尺寸(至2000 μm)的结石。

临床怎么用

  • 激光碎石时,DISS可用于吸除<250 μm的粉尘颗粒以改善视野。
  • 对于>250 μm(3.6-Ch)或>500 μm(5.1-Ch)的颗粒,DISS可能造成输尿管镜阻塞,需注意。
  • FANS可清除较大碎片,可能有助于提高较大结石碎片的清除效果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To perform benchtop experiments using direct in-scope suction (DISS) ureteroscopes (Pusen Medical, Zhuhai, China; 3.6-Ch and 5.1-Ch working channels) and flexible and navigable suction (FANS) ureteric access sheaths (ClearPetra; Wellead, Guangzhou, China) to establish their ability to clear stones of different diameters. METHODS: For each experimental group, we conducted a series of experiments using mixes of progressively larger stone particles (<63, 63-125, 125-250, 250-500, 500-1000 and 1000-2000 μm) until suction through the respective lumen failed to achieve complete stone clearance; the first experiment's stone mix contained particles <63 μm; the second experiment's stone mix contained particles <63 μm + 63-125 μm; the third experiment's stone mix contained particles <63 μm + 63-125 μm + 125-250 μm, and so on, up to 2000 μm. The stone mixes were introduced in aliquots over the course of 30 min. In the first experimental group, either a 3.6-Ch or a 5.1-Ch working channel DISS ureteroscope (with a 200-μm laser) was used to aspirate stones. In the second experimental group, either an empty 5.1-Ch working channel DISS ureteroscope or an 11/13-Ch FANS with a 9.5-Ch ureteroscope (intermittently withdrawn) was used to aspirate stones. RESULTS: The use of DISS through a 3.6-Ch working channel (with laser fibre) cleared stones up to 250 μm. DISS through a 5.1-Ch working channel cleared stones up to 500 μm with and without an indwelling laser fibre. FANS cleared all stone sizes tested (up to 2000 μm). CONCLUSIONS: During laser lithotripsy, DISS can aspirate 'dust' particles (<250 μm), which are known to most affect vision. However, particles >250 μm and >500 μm (with 3.6-Ch and 5.1-Ch working channels) may block the ureteroscope. The use of FANS, through successful clearance of larger fragments, may facilitate clearance of larger fragments.

来源:BJU Int,查看原文。版权归原作者所有。

引用

  1. Menzies-Wilson R, Ruiken T, Turney B, et al. Experimental stone clearance with in-scope suction and flexible and navigable suction access sheaths. BJU Int. 2025. PMID: 40734453. doi:10.1111/bju.16849