新型温控软性输尿管镜系统在软性输尿管镜术中监测与控制肾内温度的在体评估
研究开发了新型温控软性输尿管镜系统PT-Scope,并在猪肾结石模型中评估其监测与控制肾内温度的效果。结果显示PT-Scope测温与经皮肾造瘘测温一致,温控组峰值温度31.70±2.609°C vs 非温控组44.37±3.318°C(P<0.01),平均温度27.40±2.107°C vs 35.9±1.921°C(P<0.01)。结论认为PT-Scope能精确监测并将肾内温度控制在安全阈值内。
要点
- PT-Scope测温与经皮肾造瘘测温一致,可精确监测肾内温度。
- 温控组峰值温度31.70±2.609°C,非温控组44.37±3.318°C(P<0.01)。
- 温控组平均温度27.40±2.107°C,非温控组35.9±1.921°C(P<0.01)。
核心结论
- PT-Scope已证明能够精确监测并将肾内温度控制在安全阈值内。
研究设计与样本
- 建立猪测温模型,观察不同钬激光功率(10、20、30 W)和灌注速率(0、20、50 mL/min)下的温度变化,并用经皮肾造瘘测温和PT-Scope测量。
- 建立猪肾结石模型,将猪随机分为温控组(PT-Scope连接具有温度调节的灌注-吸引平台)和非温控组(无温度调节),比较两组肾内温度差异。
主要发现
- 在各种激光功率和灌注速率下,PT-Scope测温精确,与经皮肾造瘘测温一致。
- PT-Scope获得的温度反映距光纤尖端约0.05 cm处的温度,而靠近光纤尖端的温度显著更高。
- 温控组峰值温度为31.70±2.609°C,非温控组为44.37±3.318°C(P<0.01)。
- 温控组平均温度为27.40±2.107°C,非温控组为35.9±1.921°C(P<0.01)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate the efficacy of a novel temperature control flexible ureteroscope system in the precise monitoring and control of intrarenal temperature (IRT) during ureteroscopy. METHODS: We developed a novel temperature control flexible ureteroscope system (PT-Scope), including a temperature-monitoring ureteroscope and a irrigation-suction platform for temperature regulation. A porcine thermometry model was established to observe temperature changes under varying holmium laser powers (10, 20, 30 W) and irrigation rates (0, 20, 50 mL/min), utilizing percutaneous nephrostomy thermometry and PT-Scope measurements, with subsequent evaluation of temperature variations at different distances from the laser fiber tip. A porcine kidney stone model was established while porcine was randomly assigned to two groups: In the temperature control group, PT-Scope was connected to the irrigation-suction platform with temperature regulation, while in the nontemperature control group without temperature regulation. Comparative analysis was performed to evaluate differences in IRT between the two groups. RESULTS: Across various laser powers and irrigation rates, the temperature measurement capability of the PT-Scope was precise, demonstrating consistency with percutaneous nephrostomy temperature measurements. The temperature obtained from the PT-Scope reflect the temperature approximately 0.05 cm away from the fiber tip, whereas temperatures close to fiber tip were significantly higher. The peak temperature of the temperature control group vs nontemperature control group were 31.70 ± 2.609°C and 44.37 ± 3.318 °C, respectively (P < 0.01). The mean temperature of the temperature control group vs nontemperature control group was 27.40 ± 2.107 °C vs 35.9 ± 1.921 °C (P < 0.01). CONCLUSION: PT-Scope has demonstrated the capability to precisely monitor and control IRT within a safe threshold.
引用
- Yang J, Li Z, Lai C, et al. An In Vivo Assessment of a Novel Temperature Control Flexible Ureteroscope System for Monitoring and Controlling Intrarenal Temperature During Flexible Ureteroscopy. Urology. 2024. PMID: 39002847. doi:10.1016/j.urology.2024.07.011