无需染色、面向体外受精实验室使用的自动化精子分析
研究纳入35份新鲜精液和25份洗涤后样本,开发无需染色即可同步测量精子浓度、活动力和形态的自动化系统。结果显示浓度、活动力和形态测量的组内相关系数均高于0.97,Passing-Bablok回归与Bland-Altman分析表明自动化测量与放大图像离线人工基准可互换。结论认为该系统重现性和可靠性高,设计用于体外受精实验室对活样本进行常规定量精子分析。
要点
- 自动化系统无需染色即可同步测量精子浓度、活动力和形态。
- 浓度、活动力和形态测量的组内相关系数均高于0.97。
- 自动化测量与放大图像离线人工基准分析结果可互换。
核心结论
- 自动化精子分析显示出高重现性和可靠性。
- 该系统设计用于体外受精实验室对活样本进行常规定量精子分析。
研究设计与样本
- 研究纳入35份新鲜精液和25份洗涤后样本,来自接受生育检查的男性伴侣。
- 系统利用稳健精子追踪测量浓度和活动力,利用低对比度图像分割测量活精子形态。
- 采用组内相关系数评估重现性,采用Passing-Bablok回归和Bland-Altman分析评估可靠性。
主要发现
- 浓度、活动力和形态自动化测量的组内相关系数均高于0.97。
- 回归和Bland-Altman分析表明自动化测量与放大图像离线人工基准可互换。
- 对精液和洗涤后样本的进一步分析以及前向和非前向活动力测量也显示高重现性和可靠性。
临床怎么用
- 作者结论指出该系统设计用于体外受精实验室对活样本进行常规定量精子分析。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Computer-aided sperm analysis is typically used in andrology labs, not in in vitro fertilization labs, which requires staining for sperm morphology measurement. In in vitro fertilization labs, sperm analysis still relies on manual observation and suffers from subjectivity and inconsistency. We developed a system for automated measurement of sperm concentration, motility, and morphology without the need for sperm staining. The reproducibility and reliability of the system were evaluated. MATERIALS AND METHODS: Thirty-five fresh semen and 25 washed samples were obtained from male partners attending for fertility investigations. Sperm concentration, motility, and morphology were automatically measured simultaneously, leveraging robust sperm tracking for concentration and motility measurement and low contrast image segmentation for morphology measurement of live sperm. Reproducibility of sperm measurements was evaluated by intraclass correlation coefficients. Reliability of sperm measurement was evaluated by Passing and Bablok regression analysis and Bland-Altman analysis. RESULTS: Automated measurement of concentration, motility, and morphology had intraclass correlation coefficients higher than 0.97. The regression and Bland-Altman analysis indicated that automated measurement and off-line manual benchmarking with zoomed-in images were interchangeable. Further analysis on semen and washed samples and the measurement on progressive and nonprogressive motility also showed high reproducibility and reliability. CONCLUSIONS: Automated sperm analysis revealed high reproducibility and reliability. The system is designed for routine use in in vitro fertilization labs to perform quantitative sperm analysis on live samples.
引用
- Dai C, Shan G, Jahangiri S, et al. Staining-free, Automated Sperm Analysis for In Vitro Fertilization Lab Use. J Urol. 2022. PMID: 36097845. doi:10.1097/JU.0000000000002903
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。