文献速递泌尿系结石

比较机器人辅助透视引导与超声引导经皮肾镜取石术肾造瘘的前瞻性随机单盲临床试验

A Randomized, Single-Blind Clinical Trial Comparing Robotic-Assisted Fluoroscopic-Guided with Ultrasound-Guided Renal Access for Percutaneous Nephrolithotomy.
PubMed 文献 · J Urol · 2022-05-13

该随机单盲试验纳入2020年1月至2021年5月行输尿管镜辅助mini-PCNL(微通道经皮肾镜取石术)的大肾结石患者71例(超声组35例,机器人辅助透视引导组36例)。机器人辅助透视引导组穿刺次数更少、穿刺时间更短,但单次穿刺成功率两组无显著差异。作者认为机器人辅助透视引导肾造瘘在mini-PCNL中可能具有进一步应用潜力。

要点

  • 机器人辅助透视引导组单次穿刺成功率为50.0%,超声组为34.3%,两组差异无统计学意义(p=0.2)。
  • 机器人辅助透视引导组平均穿刺次数1.83次,显著少于超声组的2.51次(p=0.025)。
  • 机器人辅助透视引导组穿刺时间中位数5.5分钟,短于超声组的8.0分钟(p=0.049)。

核心结论

  • 机器人辅助透视引导肾造瘘在mini-PCNL中可能具有进一步应用潜力。

研究设计与样本

  • 随机单盲临床试验,采用区组随机化(在线软件),纳入2020年1月至2021年5月行输尿管镜辅助mini-PCNL的大肾结石患者,共71例(超声组35例,机器人辅助透视引导组36例)。
  • 机器人辅助透视引导组使用X线自动针靶向系统进行透视引导肾造瘘;PCNL由住院医师使用气动碎石系统通过16.5Fr/17.5Fr通道完成。

主要发现

  • 单次穿刺成功率超声组34.3% vs 机器人辅助透视引导组50.0%(p=0.2),两组无显著差异。
  • 因靶向困难导致住院医师无法建立通道的比例:超声组14.3% vs 机器人辅助透视引导组0%(p=0.025)。
  • 平均穿刺次数机器人辅助透视引导组1.83次 vs 超声组2.51次(p=0.025),机器人辅助透视引导组显著更少。
  • 穿刺时间中位数机器人辅助透视引导组5.5分钟 vs 超声组8.0分钟(p=0.049),机器人辅助透视引导组更短。
  • 术后3个月结石清除率机器人辅助透视引导组83.3% vs 超声组70.6%(p=0.26),两组无显著差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: We conducted a randomized, single-blind clinical trial comparing the surgical outcomes of robotic-assisted fluoroscopic-guided (RAF group) and ultrasound-guided (US group) renal access in mini-percutaneous nephrolithotomy (PCNL). MATERIALS AND METHODS: We recruited patients who underwent mini-PCNL with ureteroscopic assistance for large renal stones between January 2020 and May 2021. Block randomization was performed using online software. Automated needle target with x-ray was used for fluoroscopic-guided renal access in the RAF group. PCNL was performed by residents using a pneumatic lithotripsy system with 16.5Fr/17.5Fr tracts. The primary outcome was single puncture success, and the secondary outcomes were stone-free rate, complication rate, parameters measured during renal access and fluoroscopy time. RESULTS: In total, 71 patients (35 in US group, 36 in RAF group) were enrolled. No difference was seen in the single puncture success rate between the US and RAF groups (34.3% and 50.0%, p=0.2). In 14.3% cases in the US group vs no cases in the RAF group, the resident was unable to obtain access due to difficult targeting (p=0.025). The mean number of needle punctures was significantly fewer, and the median duration of needle puncture was shorter in the RAF group (1.83 vs 2.51 times, p=0.025; 5.5 vs 8.0 minutes, p=0.049, respectively). The stone-free rate at 3 months after surgery was 83.3% and 70.6% in the RAF and US groups, respectively (p=0.26). Multivariate analysis revealed that RAF guidance reduced the mean number of needle punctures by 0.73 times (p=0.021). CONCLUSIONS: RAF renal access in mini-PCNL may have further potential applications in this field.

来源:J Urol,查看原文。版权归原作者所有。

引用

  1. Taguchi K, Hamamoto S, Okada A, et al. A Randomized, Single-Blind Clinical Trial Comparing Robotic-Assisted Fluoroscopic-Guided with Ultrasound-Guided Renal Access for Percutaneous Nephrolithotomy. J Urol. 2022. PMID: 35549460. doi:10.1097/JU.0000000000002749

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。