针灸针理化与力学性能评估
该研究依据ISO 17218:2014标准,对9个品牌0.25 mm×30 mm针灸针进行扫描电镜-能谱分析及穿刺力、强度与针尖锋利度测试。结果显示各品牌针体存在变形、划痕等缺陷,强度测试后针尖均变形;各品牌穿刺力与锋利度存在统计学差异,但均未超过ISO 17218:2014限值。
要点
- 强度测试后所有品牌针尖均出现变形,提示反复操作需关注针尖完整性。
- Uniqmed穿刺力最低(中位数0.033 N,95%CI 0.030-0.035 N),Açomed最高(0.117 N,95%CI 0.109-0.132 N)。
- Dragon锋利度测试力最低(0.026 N,95%CI 0.025-0.029 N),Complementar最高(0.060 N,95%CI 0.056-0.064 N)。
核心结论
- 尽管各品牌间存在统计学显著差异,但没有任何一组超过ISO 17218:2014设定的限值。
- 部分研究未公开品牌名称,导致直接比较困难,本研究结果或可用于未来荟萃分析与ISO 17218:2014修订。
研究设计与样本
- 研究评估9个不同品牌的0.25 mm×30 mm针灸针。
- 采用扫描电镜-能谱(SEM-EDS)在300×放大下进行形态与成分评估。
- 依据ISO 17218:2014,使用Instron 4411在控制针位移条件下进行穿刺性能、强度与针尖锋利度力学测试,穿刺力以牛顿计,并用体视显微镜评估强度测试后的结构变化。
主要发现
- SEM-EDS分析显示多个品牌针体存在变形、划痕和缺陷,尤以Açomed和Complementar明显。
- 强度测试后所有品牌针尖均出现变形。
- Uniqmed穿刺性能测试力最低(中位数0.033 N,95%CI 0.030-0.035 N),Açomed最高(0.117 N,95%CI 0.109-0.132 N)。
- Dragon锋利度测试力最低(0.026 N,95%CI 0.025-0.029 N),Complementar最高(0.060 N,95%CI 0.056-0.064 N)。
关键数值与推荐
- 依据ISO 17218:2014判定,所有品牌穿刺力与锋利度均未超过标准限值。
- Açomed和Complementar在SEM-EDS下可见针体变形、划痕与缺陷。
- 所有品牌在强度测试后针尖均变形。
临床怎么用
- 选择针灸针时可参考品牌间穿刺力与锋利度差异,但均符合ISO 17218:2014限值。
- 操作中注意针尖强度测试后易变形,进针困难或反复操作时应检查针尖完整性。
局限性
- 部分同类研究未公开品牌名称,影响直接比较。
- 关于该主题的研究数量较少。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Acupuncture is a traditional Chinese medicine technique that involves the insertion of filiform needles into the body. Currently, various brands of acupuncture needles are available for purchase and product quality control is intended to ensure the safety of both the patient and the acupuncturist. OBJECTIVE: The purpose of this study was to assess the quality of acupuncture needles using physical-chemical and mechanical tests according to International Organization for Standardization (ISO) standard 17218:2014. METHODS: We evaluated nine different brands of 0.25 mm × 30 mm acupuncture needles. The needles were characterized using scanning electron microscopy-energy dispersive X-ray spectroscopy (SEM-EDS) techniques for morphological and compositional evaluation at 300× magnification. Mechanical tests were used to determine puncture performance, intensity and needle tip sharpness according to ISO 17218:2014 using an Instron 4411 with controlled needle displacement. Puncture force was measured in Newtons, and a stereomicroscope was used to evaluate structural changes after the intensity test. RESULTS: SEM-EDS analyses indicated deformations, scratches and imperfections in various brands of needle, particularly Açomed and Complementar. Following the intensity test, all brands exhibited deformations at the needle tips. The Uniqmed brand had the lowest puncture performance test force (median = 0.033 N, 95% confidence interval (CI) = 0.030-0.035 N), while Açomed had the highest (0.117 N, 95% CI = 0.109-0.132 N). The Dragon brand had the lowest sharpness test force (0.026 N, 95% CI = 0.025-0.029 N), while Complementar had the highest (0.060 N, 95% CI = 0.056-0.064 N). CONCLUSION: Despite statistically significant disparities, no group exceeded the limits set by ISO 17218:2014. Although there are a few research studies on this topic, some authors have opted not to reveal brand names, which makes direct comparisons problematic. As a result, we anticipate that the findings from this study may be used in future meta-analyses and ISO 17218:2014 revisions.
引用
- Dos Santos VAB, Souza TP, Janini ACP, et al. Physicochemical and mechanical evaluation of acupuncture needles. Acupunct Med. 2025. PMID: 40400299. doi:10.1177/09645284251343911
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。