基于传感器的电子脊柱活动度指数在轴性脊柱关节炎中的效度与信度
该重复测量研究纳入40例轴性脊柱关节炎(axSpA)患者,使用惯性测量单元(IMU)传感器测量颈椎和腰椎最大活动范围并构建复合IMU评分。复合IMU评分与BASMI密切相关(r=0.88),各IMU运动测试与BASFI强相关,IMU测试的重测信度(ICC 0.74-0.98)与BASMI(ICC=0.96)相当。
要点
- 复合IMU评分与BASMI高度相关(r=0.88),支持其效度。
- 各IMU运动测试均与BASFI强相关,而BASMI仅部分测试如此。
- IMU测试重测信度良好至优秀(ICC 0.74-0.98),复合区域评分ICC为0.94-0.97。
核心结论
- 使用可穿戴IMU传感器测量颈椎和腰椎活动度在多个平面(包括旋转)上是有效且可靠的评估方法,适用于不同严重程度的axSpA患者。
研究设计与样本
- 重复测量研究设计,纳入40例axSpA患者,使用成对IMU传感器测量颈椎和腰椎最大活动范围,并构建复合IMU评分,同时进行常规测量和身体功能评估。
主要发现
- 复合IMU评分与BASMI密切相关(r=0.88)。
- 常规颈椎旋转和侧屈测试与IMU对应测试密切相关(r=0.85、0.84)。
- 所有IMU运动测试均与BASFI强相关,而BASMI仅部分测试与BASFI强相关。
- 常规测试和IMU测试(除胸廓扩张外)的信度范围从良好到优秀;单个常规测试的重测ICC为0.57-0.91,而每个IMU测试为0.74-0.98。
- 每个复合区域IMU评分的重测信度优秀(ICC=0.94-0.97),与BASMI的信度(ICC=0.96)相当。
临床怎么用
- 可穿戴IMU传感器可作为评估axSpA患者脊柱活动度的有效且可靠工具,尤其适用于需要多平面(包括旋转)评估的场景。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate the validity and reliability of inertial measurement unit (IMU) sensors in the assessment of spinal mobility in axial spondyloarthritis (axSpA). METHODS: A repeated measures study design involving 40 participants with axSpA was used. Pairs of IMU sensors were used to measure the maximum range of movement at the cervical (Cx) and lumbar (Lu) spine. A composite IMU score was defined by combining the IMU measures. Conventional metrology and physical function assessment were performed. Validation was assessed considering the agreement of IMU measures with conventional metrology and correlation with physical function. Reliability was assessed using intra-class correlation coefficients (ICCs). RESULTS: The composite IMU score correlated closely (r = 0.88) with the BASMI. Conventional Cx rotation and lateral flexion tests correlated closely with IMU equivalents (r = 0.85, 0.84). All IMU movement tests correlated strongly with BASFI, while this was true for only some of the BASMI tests. The reliability of both conventional and IMU tests (except for chest expansion) ranged from good to excellent. Test-retest ICCs for individual conventional tests varied between 0.57 and 0.91, in comparison to a range from 0.74 to 0.98 for each of the IMU tests. Each of the composite regional IMU scores had excellent test-retest reliability (ICCs=0.94-0.97), comparable to the reliability of the BASMI (ICC=0.96). CONCLUSION: Cx and Lu spinal mobility measured using wearable IMU sensors is a valid and reliable assessment in multiple planes (including rotation), in patients with a wide range of axSpA severity.
引用
- Gardiner PV, Small D, Muñoz-Esquivel K, et al. Validity and reliability of a sensor-based electronic spinal mobility index for axial spondyloarthritis. Rheumatology (Oxford). 2020. PMID: 32342100. doi:10.1093/rheumatology/keaa122
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。