文献速递其他结缔组织病

手持测力计评估系统性硬化症相关肌病近端肌力的效度与信度

Validity and reliability of hand-held dynamometry for proximal strength assessment in systemic sclerosis-associated myopathy.
PubMed 文献 · Semin Arthritis Rheum · 2026-04-19

这项前瞻性横断面研究纳入36例系统性硬化症相关肌病(SSc-AM)患者,用MicroFET-2手持测力计(HHD)评估近端等长肌力,并与MMT-8和近端MMT-3比较。复合HHD评分与MMT-8总分(r=0.695,p<0.0001)及近端MMT-3(r=0.736,p<0.0001)强相关,组内信度优异(ICC 0.885-0.985),但髋外展肌组间信度较低(ICC 0.06-0.13)。

要点

  • 复合HHD评分与MMT-8总分及近端MMT-3均呈强相关,提示与常用肌力评估一致。
  • 各近端肌群组内信度优异(ICC 0.885-0.985),但髋外展肌组间信度低(ICC 0.06-0.13)。
  • HHD可定量评估SSc-AM近端等长肌力,但需进一步纵向研究明确反应性和临床用途。

核心结论

  • 手持测力计可对SSc-AM患者近端等长肌力进行定量评估,本研究显示组内重测信度优异。
  • HHD与MMT-8及MMT-3的相关性提示其与常用肌力测量方法一致。

研究设计与样本

  • 前瞻性横断面研究,纳入36例系统性硬化症相关肌病患者。
  • 使用MMT-8、近端MMT-3(颈屈肌、肩外展肌、髋外展肌)和MicroFET-2手持测力计进行肌力测试。
  • 每肌群三次HHD试验取平均值并计算复合HHD评分;8例参与者由第二名检查者重复测试以评估组间信度。

主要发现

  • 复合HHD评分与MMT-8总分强相关(r=0.695,p<0.0001)。
  • 复合HHD评分与近端MMT-3强相关(r=0.736,p<0.0001)。
  • 所有近端肌群组内信度优异(ICC范围0.885-0.985)。
  • 颈屈肌和肩外展肌组间信度可接受(ICC 0.60-0.75),髋外展肌组间信度较低(ICC 0.06-0.13)。

关键数值与推荐

  • 复合HHD评分与MMT-8总分相关系数r=0.695,p<0.0001。
  • 复合HHD评分与近端MMT-3相关系数r=0.736,p<0.0001。
  • 组内信度ICC范围0.885-0.985。
  • 颈屈肌和肩外展肌组间信度ICC 0.60-0.75;髋外展肌组间信度ICC 0.06-0.13。

临床怎么用

  • HHD可作为SSc-AM患者近端肌力的定量评估工具,组内重复测量信度优异。
  • 髋外展肌组间信度低,临床用于髋外展肌评估时需注意稳定性和检查者培训。

局限性

  • 组间信度评估样本量小(n=8)。
  • 髋外展肌组间信度低反映其稳定性要求较高且样本量小。
  • 需进一步纵向研究评估反应性并明确临床用途。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To evaluate the validity and reliability of hand-held dynamometry (HHD) for quantifying proximal muscle strength in patients with systemic sclerosis-associated myopathy (SSc-AM), and to compare HHD with the validated manual muscle testing-8 (MMT-8) and modified three-muscle proximal MMT (MMT-3). METHODS: In this prospective, cross-sectional study, 36 patients with SSc-AM underwent strength testing using MMT-8, a proximal MMT-3 (neck flexors, shoulder abductors, hip abductors) and a quantitative hand-held dynamometry (MicroFET-2). For each muscle group, three HHD trials were averaged, and composite HHD scores were calculated. Eight participants underwent repeat testing by a second examiner to assess inter-rater reliability. Pearson correlation coefficients were calculated to assess the association between HHD and MMT scores. Intra-rater and inter-rater reliability were evaluated using intraclass correlation coefficients (ICCs) derived from two-way mixed effects model with absolute agreement. RESULTS: Composite HHD scores showed strong correlations with MMT-8 total scores (r = 0.695, p < 0.0001) and with the proximal MMT-3 (r = 0.736, p < 0.0001). Intra-rater reliability was excellent across all proximal muscle groups (ICC range 0.885-0.985). Inter-rater reliability was acceptable for neck flexors and shoulder abductors (ICC 0.60-0.75) but lower for hip abductors (ICC 0.06-0.13), reflecting higher stabilization requirements and small sample size (n = 8). CONCLUSION: Hand-held dynamometry provides a quantitative assessment of proximal isometric strength in SSc-AM and demonstrated excellent intra-rater reproducibility in this study. Correlation with MMT-8 and MMT-3 demonstrates alignment with commonly used strength measures. Given known limitations of MMT-based assessment and variable inter-rater reproducibility, further longitudinal studies are needed to evaluate responsiveness and define its clinical utility.

来源:Semin Arthritis Rheum,查看原文。版权归原作者所有。

引用

  1. Paik JJ, Nallapati A, Mecoli CA, et al. Validity and reliability of hand-held dynamometry for proximal strength assessment in systemic sclerosis-associated myopathy. Semin Arthritis Rheum. 2026. PMID: 42068856. doi:10.1016/j.semarthrit.2026.152990

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。