本研究以30例类风湿关节炎患者为对象,对22个手关节行两次ARTHUR V.2.0扫描并与风湿病专家评估对照,DIANA V.2.0按EULAR-OMERACT量表(0-3)评分。ARTHUR成功扫描660个关节中的564个(成功率85.5%);系统与专家在滑膜肥厚和Doppler活动的二分类一致性分别为87.3%和91.2%,作者认为表现令人鼓舞但需进一步优化特定关节。

要点

  • ARTHUR V.2.0扫描660个手关节中成功564个,总体成功率85.5%。
  • 系统与专家滑膜肥厚二分类一致率87.3%、Doppler活动91.2%。
  • 作者认为需进一步改进以优化特定关节的表现。

核心结论

  • ARTHUR V.2.0与DIANA V.2.0的可重复性与文献报道的专家内部一致性相当,与人类评估者的一致性令人鼓舞,但需进一步改进以优化特定关节的表现。

研究设计与样本

  • 研究纳入30例类风湿关节炎患者,每例对22个手关节连续行两次ARTHUR V.2.0扫描,并以风湿病专家评估为参考标准,专家对自动化系统结果设盲。
  • DIANA V.2.0按EULAR-OMERACT量表(0-3分)对滑膜肥厚和Doppler活动评分,一致性采用加权Cohen kappa、精确一致率(PEA)、接近一致率(PCA)及二分类结局(健康≤1 vs 疾病≥2)评价。
  • 比较包括机器人内部重复性及其与专家风湿病医师和设盲独立评估者的一致性。

主要发现

  • ARTHUR成功扫描660个关节中的564个,总体成功率85.5%。
  • 机器人内部一致性:滑膜肥厚PEA 63.0%、PCA 93.0%、二分类90.5%、kappa 0.54;Doppler PEA 74.8%、PCA 93.7%、二分类88.1%、kappa 0.49。
  • ARTHUR+DIANA与风湿病专家一致性:滑膜肥厚PEA 57.9%、PCA 92.9%、二分类87.3%、kappa 0.38;Doppler PEA 77.3%、PCA 94.2%、二分类91.2%、kappa 0.44。
  • 与独立评估者一致性:滑膜肥厚PEA 49.0%、PCA 91.2%、二分类80.0%、kappa 0.39;Doppler PEA 62.6%、PCA 94.4%、二分类88.1%、kappa 0.48。

关键数值与推荐

  • 滑膜肥厚和Doppler活动采用EULAR-OMERACT量表(0-3分)评分。
  • 二分类结局按Global OMERACT-EULAR Synovitis Scoring定义:健康≤1分,疾病≥2分。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 作者指出需进一步改进以优化特定关节的表现。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To evaluate the agreement and repeatability of an automated robotic ultrasound system (ARTHUR V.2.0) combined with an AI model (DIANA V.2.0) in assessing synovial hypertrophy (SH) and Doppler activity in rheumatoid arthritis (RA) patients, using an expert rheumatologist's assessment as the reference standard. METHODS: 30 RA patients underwent two consecutive ARTHUR V.2.0 scans and rheumatologist assessment of 22 hand joints, with the rheumatologist blinded to the automated system's results. Images were scored for SH and Doppler by DIANA V.2.0 using the EULAR-OMERACT scale (0-3). The agreement was evaluated by weighted Cohen's kappa, percent exact agreement (PEA), percent close agreement (PCA) and binary outcomes using Global OMERACT-EULAR Synovitis Scoring (healthy ≤1 vs diseased ≥2). Comparisons included intra-robot repeatability and agreement with the expert rheumatologist and a blinded independent assessor. RESULTS: ARTHUR successfully scanned 564 out of 660 joints, corresponding to an overall success rate of 85.5%. Intra-robot agreement for SH: PEA 63.0%, PCA 93.0%, binary 90.5% and for Doppler, PEA 74.8%, PCA 93.7%, binary 88.1% and kappa values of 0.54 and 0.49. Agreement between ARTHUR+DIANA and the rheumatologist: SH (PEA 57.9%, PCA 92.9%, binary 87.3%, kappa 0.38); Doppler (PEA 77.3%, PCA 94.2%, binary 91.2%, kappa 0.44) and with the independent assessor: SH (PEA 49.0%, PCA 91.2%, binary 80.0%, kappa 0.39); Doppler (PEA 62.6%, PCA 94.4%, binary 88.1%, kappa 0.48). CONCLUSIONS: ARTHUR V.2.0 and DIANA V.2.0 demonstrated repeatability on par with intra-expert agreement reported in the literature and showed encouraging agreement with human assessors, though further refinement is needed to optimise performance across specific joints.