OMERACT 巨细胞动脉炎体模项目:验证用于诊断巨细胞动脉炎的3D打印超声训练体模

OMERACT GCA phantom project: validation of a 3D-printed ultrasound training phantom for diagnosis of giant cell arteritis.
PubMed 文献 · RMD Open · 2025-02-06

该研究让来自12个欧洲国家的28名超声专家在盲法下依据OMERACT巨细胞动脉炎超声定义检查腋动脉和颞动脉3D打印体模,两轮中体模正常/异常分类正确率均>81%,Fleiss kappa第一轮0.80(95%CI 0.78至0.81)、第二轮0.74(95%CI 0.72至0.75),IMT测量的组内相关系数(ICC 1.1)两轮均为0.98(95%CI 0.98至0.99)。

要点

  • 超声专家使用3D打印腋动脉和颞动脉体模,依据OMERACT巨细胞动脉炎超声定义进行检查。
  • 体模正常/异常分类正确率两轮均>81%,Fleiss kappa第一轮0.80、第二轮0.74。
  • IMT测量的ICC 1.1两轮均为0.98,评分者内信度中位Cohen kappa为0.83、中位ICC为0.78。

核心结论

  • 研究显示超声专家在应用OMERACT巨细胞动脉炎超声定义及测量腋动脉和颞动脉3D打印体模的IMT方面具有高信度。
  • 该体模可能有助于临床医生在培训中评估疑似或确诊巨细胞动脉炎患者的大动脉。

研究设计与样本

  • 来自12个欧洲国家的28名参与者获得八套腋动脉和颞浅动脉(包括常见分支、额支和顶支)体模,按预定方案盲法检查并依据OMERACT巨细胞动脉炎超声定义评估。
  • 由于内膜-中膜复合体勾画困难,体模的顶支被修改,并进行了第二轮检查。
  • 测量IMT并将体模分类为正常或血管炎性。

主要发现

  • 两轮中体模被正确分类为正常/异常的病例均>81%。
  • Fleiss kappa第一轮为0.80(95%CI 0.78至0.81),第二轮为0.74(95%CI 0.72至0.75)。
  • IMT测量的组内相关系数(ICC 1.1)两轮均为0.98(95%CI 0.98至0.99)。
  • 评分者内信度良好,中位Cohen kappa为0.83,中位ICC为0.78。

关键数值与推荐

  • 体模正常/异常分类正确率两轮均>81%。
  • Fleiss kappa第一轮0.80(95%CI 0.78至0.81),第二轮0.74(95%CI 0.72至0.75)。
  • IMT测量ICC 1.1两轮均为0.98(95%CI 0.98至0.99)。
  • 评分者内信度中位Cohen kappa为0.83,中位ICC为0.78。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Ultrasonography is crucial for diagnosing giant cell arteritis (GCA); however, training opportunities are rare. This study tested the reliability of ultrasonography findings and measurements of the intima-media thickness (IMT) among ultrasonography experts by using phantoms of the axillary (AA) and temporal arteries (TA) created with high-resolution 3D printing. METHODS: Twenty-eight participants from 12 European countries received eight sets of phantoms of the AA and the superficial TA (including common, frontal and parietal branches), which were examined in a blinded fashion according to a predefined protocol and evaluated based on Outcome Measures in Rheumatology (OMERACT) GCA ultrasound definitions. Due to difficulties with the delineation of the intima-media complex, the parietal branch of the phantoms was modified, and a second round was conducted. The IMT was measured, and phantoms were classified as normal or vasculitic. RESULTS: In both rounds, the phantoms were correctly classified as normal/abnormal in >81% of cases yielding a Fleiss' kappa of 0.80 (95% CI 0.78 to 0.81) in round 1 and 0.74 (95% CI 0.72 to 0.75) in round 2. IMT measurements revealed an intraclass correlation coefficient (ICC 1.1) of 0.98 (95% CI 0.98 to 0.99) in both rounds. Intrarater reliability was good with a median Cohens Kappa of 0.83 and median ICC of 0.78. CONCLUSION: The study demonstrated high reliability among ultrasound experts in applying the OMERACT ultrasound definitions for GCA and in measuring the IMT using a 3D-printed phantom of the AA and TA. This phantom could assist clinicians in training to assess the large arteries of patients with suspected or established GCA.

来源:RMD Open,查看原文。版权归原作者所有。