文献速递神经康复骨科与运动康复

MERLIN居家上肢康复机器人系统在卒中患者中的可用性研究

A usability study in patients with stroke using MERLIN, a robotic system based on serious games for upper limb rehabilitation in the home setting.
PubMed 文献 · J Neuroeng Rehabil · 2021-02-23

本研究评估了基于严肃游戏的MERLIN机器人系统在9例不同恢复阶段卒中患者居家上肢康复中的可用性。患者先接受1周中心训练,随后居家使用2周。结果显示系统满意度高(SUS均值71.94%),Fugl-Meyer运动功能总分较干预前有中度改善(p=0.002),且安全性、可行性和激励性良好。

要点

  • MERLIN居家上肢康复系统在卒中患者中满意度高(SUS均值71.94%)。
  • 干预后Fugl-Meyer运动功能总分较前中度改善(p=0.002)。
  • 系统安全、可行、有激励性,支持居家远程康复延伸。

核心结论

  • 基于严肃游戏的MERLIN系统用于卒中患者居家上肢康复是安全、有用、可行且具有激励性的。

研究设计与样本

  • 招募9例处于亚急性、短期慢性或长期慢性恢复阶段且上肢/手功能障碍的卒中患者,使用MERLIN系统3周(1周在IMIBIC训练,2周居家)。
  • 干预后采用系统可用性量表(SUS)、适配内在动机量表(IMI)、魁北克辅助技术满意度评估(QUEST)和ArmAssist可用性评估问卷评估可用性。
  • 干预前后评估上肢运动功能临床结局。

主要发现

  • 9例患者均完成研究。
  • 可用性评估显示高满意度:SUS均值71.94%(SD=16.38),QUEST均值3.81(SD=0.38),适配IMI均值6.12(SD=1.36)。
  • ArmAssist问卷平均6分(满分7分),表明MERLIN非常直观、易学易用。
  • Fugl-Meyer运动功能总分从干预前到干预后显示中度改善(p=0.002)。
  • 改良Ashworth量表结局无显著变化(p=0.169)。

关键数值与推荐

  • SUS均值71.94%(SD=16.38)。
  • QUEST均值3.81(SD=0.38)。
  • 适配IMI均值6.12(SD=1.36)。
  • ArmAssist问卷平均6分(满分7分)。
  • Fugl-Meyer运动功能总分干预前后比较p=0.002;改良Ashworth量表p=0.169。

临床怎么用

  • MERLIN系统可用于卒中患者居家上肢康复,以延伸机构康复并增加和延长治疗。
  • 系统直观易学易用且具有激励性,适合在家庭环境中开展远程康复。

局限性

  • 样本量小(9例),未设对照组。
  • 研究为可用性研究,临床结局改善需更大样本验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Neuroscience and neurotechnology are transforming stroke rehabilitation. Robotic devices, in addition to telerehabilitation, are increasingly being used to train the upper limbs after stroke, and their use at home allows us to extend institutional rehabilitation by increasing and prolonging therapy. The aim of this study is to assess the usability of the MERLIN robotic system based on serious games for upper limb rehabilitation in people with stroke in the home environment. METHODS: 9 participants with a stroke in three different stages of recovery (subacute, short-term chronic and long-term chronic) with impaired arm/hand function, were recruited to use the MERLIN system for 3 weeks: 1 week training at the Maimonides Biomedical Research Institute of Cordoba (IMIBIC), and 2 weeks at the patients' homes. To evaluate usability, the System Usability Scale (SUS), Adapted Intrinsic Motivation Inventory (IMI), Quebec User Evaluation of Satisfaction with assistive Technology (QUEST), and the ArmAssist Usability Assessment Questionnaire were used in the post-intervention. Clinical outcomes for upper limb motor function were assessed pre- and post-intervention. RESULTS: 9 patients participated in and completed the study. The usability assessment reported a high level of satisfaction: mean SUS score 71.94 % (SD = 16.38), mean QUEST scale 3.81 (SD = 0.38), and mean Adapted IMI score 6.12 (SD = 1.36). The results of the ArmAssist Questionnaire showed an average of 6 out of 7, which indicates that MERLIN is extremely intuitive, easy to learn and easy to use. Regarding clinical assessment, the Fugl-Meyer scores showed moderate improvements from pre- to post-intervention in the total score of motor function (p = 0.002). There were no significant changes in the Modified Ashworth scale outcomes (p = 0.169). CONCLUSIONS: This usability study indicates that home-based rehabilitation for upper limbs with the MERLIN system is safe, useful, feasible and motivating. Telerehabilitation constitutes a major step forward in the use of intensive rehabilitation at home. Trial registration ClinicalTrials.gov, NCT04405609. Registered 06 January 2020-Retrospectively registered, https://clinicaltrials.gov/ct2/show/NCT04405609.

来源:J Neuroeng Rehabil,查看原文。版权归原作者所有。

引用

  1. Guillén-Climent S, Garzo A, Muñoz-Alcaraz MN, et al. A usability study in patients with stroke using MERLIN, a robotic system based on serious games for upper limb rehabilitation in the home setting. J Neuroeng Rehabil. 2021. PMID: 33622344. doi:10.1186/s12984-021-00837-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。