自适应多通道FES神经假体:学习控制与自动步态评估
该研究旨在开发一种学习型全腿支撑FES神经假体,控制膝关节屈伸及踝关节背屈跖屈的拮抗肌对,并建立连续评估膝踝关节角度的方法。在4例不完全性脊髓损伤患者中,3例可生成个体化刺激模式,其中2例足部平均角度、1例膝关节平均角度有约4°的轻微改善。
要点
- 学习型全腿支撑FES神经假体可为不完全性SCI患者生成个体化刺激模式。
- 3例可适配者中2例足部平均角度、1例膝关节平均角度改善约4°。
- 严重步态波动者可致步态相位检测误触发,需排除。
核心结论
- 所提方法可实现对受试者当前步态的直接适配,初步试验显示神经假体可生成个体化刺激模式。
研究设计与样本
- 首次测试纳入4例不完全性脊髓损伤患者,评估学习型全腿支撑FES神经假体的刺激模式生成与关节角度变化。
主要发现
- 4例中3例生成个体化刺激模式,另1例因步态相位检测误触发被排除。
- 3例中2例足部平均角度、1例膝关节平均角度有轻微改善,幅度约4°,见于背屈峰值、跖屈峰值或膝屈曲峰值时刻。
关键数值与推荐
- 采用基于生理关节角度参考带的迭代学习控制器(ILC),不依赖个体步长持续时间。
- 通过将健康受试者生理关节角度形变匹配个体病理步态,实现膝与足关节角度的每步连续评估。
临床怎么用
- 该神经假体有望为不完全性SCI患者提供个体化多通道FES步态支持,但需先解决步态波动者的触发时序问题。
局限性
- 样本量小(4例),仅3例完成适配,结果为初步测试。
- 膝关节角度重置敏感性、显著步态波动者的时序问题及ILC增益自动调谐尚未解决。
- 未与常规多通道FES神经假体进行长期康复效果比较。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: FES (Functional Electrical Stimulation) neuroprostheses have long been a permanent feature in the rehabilitation and gait support of people who had a stroke or have a Spinal Cord Injury (SCI). Over time the well-known foot switch triggered drop foot neuroprosthesis, was extended to a multichannel full-leg support neuroprosthesis enabling improved support and rehabilitation. However, these neuroprostheses had to be manually tuned and could not adapt to the persons' individual needs. In recent research, a learning controller was added to the drop foot neuroprosthesis, so that the full stimulation pattern during the swing phase could be adapted by measuring the joint angles of previous steps. METHODS: The aim of this research is to begin developing a learning full-leg supporting neuroprosthesis, which controls the antagonistic muscle pairs for knee flexion and extension, as well as for ankle joint dorsi- and plantarflexion during all gait phases. A method was established that allows a continuous assessment of knee and foot joint angles with every step. This method can warp the physiological joint angles of healthy subjects to match the individual pathological gait of the subject and thus allows a direct comparison of the two. A new kind of Iterative Learning Controller (ILC) is proposed which works independent of the step duration of the individual and uses physiological joint angle reference bands. RESULTS: In a first test with four people with an incomplete SCI, the results showed that the proposed neuroprosthesis was able to generate individually fitted stimulation patterns for three of the participants. The other participant was more severely affected and had to be excluded due to the resulting false triggering of the gait phase detection. For two of the three remaining participants, a slight improvement in the average foot angles could be observed, for one participant slight improvements in the averaged knee angles. These improvements where in the range of 4circat the times of peak dorsiflexion, peak plantarflexion, or peak knee flexion. CONCLUSIONS: Direct adaptation to the current gait of the participants could be achieved with the proposed method. The preliminary first test with people with a SCI showed that the neuroprosthesis can generate individual stimulation patterns. The sensitivity to the knee angle reset, timing problems in participants with significant gait fluctuations, and the automatic ILC gain tuning are remaining issues that need be addressed. Subsequently, future studies should compare the improved, long-term rehabilitation effects of the here presented neuroprosthesis, with conventional multichannel FES neuroprostheses.
引用
- Müller P, Del Ama AJ, Moreno JC, et al. Adaptive multichannel FES neuroprosthesis with learning control and automatic gait assessment. J Neuroeng Rehabil. 2020. PMID: 32111245. doi:10.1186/s12984-020-0640-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。