儿童康复中标准化与日常生活条件下基于传感器的步速估算准确性与比较
结果显示传感器步速估算的测量误差在标准化条件下为0.01±0.07 m/s、日常生活条件下为0.04±0.06 m/s,且多数儿童在测试情境与日常生活中并未采用相同步速。
要点
- 单个踝部惯性传感器可在标准化与日常生活条件下较准确估算儿童步速。
- 多数儿童步行测试速度与日常生活步速不一致。
- 建议传感器置于受累较轻侧,佩戴矫形器时则置于矫形器侧。
核心结论
- 基于单个踝部惯性传感器的步速估算法在标准化与日常生活条件下均能较准确测量儿童步速,且多数儿童在两种条件下的步速并不相同。
研究设计与样本
- 通过将基于传感器的步速估算与参考方法在两种条件下比较来确定准确性,并比较个体在两种条件下的步速。
主要发现
- 测量误差在标准化条件下为0.01±0.07 m/s,在日常生活条件下为0.04±0.06 m/s。
- 多数儿童在测试情境中采用的步速与日常生活不同。
关键数值与推荐
- 测量误差:标准化条件0.01±0.07 m/s;日常生活条件0.04±0.06 m/s。
- 推荐将传感器置于受累较轻侧;若儿童佩戴矫形器,则置于佩戴矫形器的一侧。
临床怎么用
- 可在标准化步行测试和儿童习惯环境中使用单个踝部惯性传感器测量步速,以评估康复后的步行表现。
- 传感器佩戴位置按受累较轻侧或矫形器侧选择,有助于提高日常生活中的佩戴时间与数据质量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Gait speed is a widely used outcome measure to assess the walking abilities of children undergoing rehabilitation. It is routinely determined during a walking test under standardized conditions, but it remains unclear whether these outcomes reflect the children's performance in daily life. An ankle-worn inertial sensor provides a usable opportunity to measure gait speed in the children's habitual environment. However, sensor-based gait speed estimations need to be accurate to allow for comparison of the children's gait speed between a test situation and daily life. Hence, the first aim of this study was to determine the measurement error of a novel algorithm that estimates gait speed based on data of a single ankle-worn inertial sensor in children undergoing rehabilitation. The second aim of this study was to compare the children's gait speed between standardized and daily life conditions. METHODS: Twenty-four children with walking impairments completed four walking tests at different speeds (standardized condition) and were monitored for one hour during leisure or school time (daily life condition). We determined accuracy by comparing sensor-based gait speed estimations with a reference method in both conditions. Eventually, we compared individual gait speeds between the two conditions. RESULTS: The measurement error was 0.01 ± 0.07 m/s under the standardized and 0.04 ± 0.06 m/s under the daily life condition. Besides, the majority of children did not use the same speed during the test situation as in daily life. CONCLUSION: This study demonstrates an accurate method to measure children's gait speed during standardized walking tests and in the children's habitual environment after rehabilitation. It only requires a single ankle sensor, which potentially increases wearing time and data quality of measurements in daily life. We recommend placing the sensor on the less affected side, unless the child wears one orthosis. In this latter case, the sensor should be placed on the side with the orthosis. Moreover, this study showed that most children did not use the same speed in the two conditions, which encourages the use of wearable inertial sensors to assess the children's walking performance in their habitual environment following rehabilitation.
引用
- Rast FM, Aschwanden S, Werner C, et al. Accuracy and comparison of sensor-based gait speed estimations under standardized and daily life conditions in children undergoing rehabilitation. J Neuroeng Rehabil. 2022. PMID: 36195950. doi:10.1186/s12984-022-01079-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。