直肠癌在线自适应MR引导剂量递增的PTV边界评估
本研究在1.5 T MR-Linac上评估中危直肠癌在线自适应MRgRT剂量递增策略中覆盖肿瘤分次内运动所需边界。15例患者接受5×5 Gy新辅助短程放疗,每例共26次3D T2加权MRI。结果示覆盖分次间和分次内GTV运动所需边界分别为17 mm和6 mm,扫描间隔缩短至15分钟或更短时4 mm边界即可满足覆盖。
要点
- 覆盖分次间GTV运动需17 mm边界,分次内运动需6 mm边界。
- 扫描间隔越短所需边界越小,15分钟或更短时4 mm边界可满足覆盖。
- 缩短再计划至剂量投照结束的时间可减小在线自适应MRgRT的靶区边界。
核心结论
- 在中危直肠癌在线自适应MRgRT剂量递增策略中,治疗时间越短,覆盖GTV运动所需边界越小。
- 当再计划与剂量投照结束之间的时间间隔缩短至15分钟时,4 mm边界即可实现充分靶区覆盖。
研究设计与样本
- 前瞻性单臂研究,15例中危直肠癌患者接受新辅助短程放疗5×5 Gy,采用1.5 T MR-Linac在线自适应工作流程。
- 每例患者共采集26次3D T2加权MRI:治疗前1次参考扫描,每个治疗分次5次扫描。
- 在每个扫描上勾画原发肿瘤为GTV,通过等向扩展参考GTV直至覆盖连续GTV超过95%体素来评估靶区覆盖边界,覆盖概率阈值90%定义为充分。
主要发现
- 覆盖分次间GTV运动所需边界为17 mm。
- 覆盖分次内GTV运动所需边界为6 mm。
- 基于扫描间时间间隔的分析显示,时间间隔越短所需边界越小,15分钟或更短的操作仅需4 mm边界即可充分覆盖GTV。
关键数值与推荐
- 靶区覆盖评估标准:等向扩展参考GTV直至覆盖连续GTV超过95%体素。
- 充分边界定义:覆盖概率阈值90%。
- 分次间GTV运动所需边界17 mm,分次内GTV运动所需边界6 mm。
- 扫描间隔15分钟或更短时,4 mm边界可充分覆盖GTV。
临床怎么用
- 在中危直肠癌在线自适应MRgRT剂量递增中,应尽量缩短再计划至剂量投照结束的时间以减小所需边界。
- 若时间间隔可控制在15分钟或更短,可考虑使用4 mm边界覆盖GTV运动。
- 需注意分次间GTV运动所需边界较大(17 mm),在线自适应流程中应予以考虑。
局限性
- 样本量较小,仅15例患者。
- 研究为单中心单臂设计,未设对照组。
- 边界评估基于MRI勾画的GTV,未涉及临床靶区或计划靶区的外扩策略验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: This study assessed the margins needed to cover tumor intrafraction motion during an MR-guided radiotherapy (MRgRT) dose-escalation strategy in intermediate risk rectal cancer. METHODS: Fifteen patients with rectal cancer were treated with neoadjuvant short-course radiotherapy, 5x5 Gy, according to an online adaptive workflow on a 1.5 T MR-linac. Per patient, 26 3D T2 weighted MRIs were made; one reference scan preceding treatment and five scans per treatment fraction. The primary tumor was delineated on each scan as gross tumor volume (GTV). Target coverage margins were assessed by isotropically expanding the reference GTV until more than 95% of the voxels of the sequential GTVs were covered. A margin with a coverage probability threshold of 90% was defined as adequate. Intra- and interfraction margins to cope with the movement of the GTV in the period between scans were calculated to indicate the target volume margins. Furthermore, the margin needed to cover GTV movement was calculated for different time intervals. RESULTS: The required margins to cover inter- and intrafraction GTV motion were 17 mm and 6 mm, respectively. Analysis based on time intervals between scans showed smaller margins were needed for adequate GTV coverage as time intervals became shorter, with a 4 mm margin required for a procedure of 15 min or less. CONCLUSION: The shorter the treatment time, the smaller the margins needed to cover for the GTV movement during an online adaptive MRgRT dose-escalation strategy for intermediate risk rectal cancer. When time intervals between replanning and the end of dose delivery could be reduced to 15 min, a 4 mm margin would allow adequate target coverage.