MR引导三分配加速部分乳腺照射改善同侧乳腺与胸壁保护

Improved Ipsilateral Breast and Chest Wall Sparing With MR-Guided 3-fraction Accelerated Partial Breast Irradiation: A Dosimetric Study Comparing MR-Linac and CT-Linac Plans.
PubMed 文献 · Adv Radiat Oncol · 2021-01-22

本研究对10例乳腺癌患者比较MR-Linac前向调强放疗与CT-Linac容积旋转调强放疗在新型三分配APBI方案中的剂量学差异。MR PTV3mm与CT PTV3mm计划质量相当,但MR PTV3mm较CT PTV10mm显著改善同侧乳腺和胸壁保护。

要点

  • MR PTV3mm与CT PTV3mm计划靶区覆盖和OAR剂量无显著差异
  • MR PTV3mm较CT PTV10mm显著降低同侧乳腺V12.3Gy和胸壁V24Gy
  • 剂量学差异的临床相关性尚不明确

核心结论

  • MR与CT计划在均采用3mm PTV外扩时计划质量相当,MR PTV3mm较CT PTV10mm改善同侧乳腺和胸壁保护。

研究设计与样本

  • 研究纳入10例乳腺癌患者,按NSABP B39/RTOG 0413协议勾画靶区与危及器官。
  • 比较MR PTV3mm与CT PTV3mm计划、MR PTV3mm与CT PTV10mm计划,采用Wilcoxon配对符号秩检验。
  • APBI方案为3次照射,临床靶区24.6 Gy、PTV 23.4 Gy,OAR约束从现有5次APBI协议缩放。

主要发现

  • MR PTV3mm与CT PTV3mm平均PTV分别为84.3±51.9 cm3和82.6±55 cm3(P=.5),PTV V23.4Gy、均匀性指数、适形指数和R50相似。
  • MR PTV3mm与CT PTV10mm平均PTV分别为84.3±51.9 cm3和131.7±74.4 cm3(P=.002),PTV V23.4Gy分别为99%±0.9%和97.6%±1.4%(P=.03)。
  • MR PTV3mm较CT PTV10mm同侧乳腺V12.3Gy为34.8%±12.7% vs 44.4%±10.9%(P=.002),胸壁V24Gy为8.5±5.5 cm3 vs 21.8±14.9 cm3(P=.004)。

关键数值与推荐

  • PTV剂量覆盖指标V23.4Gy:MR PTV3mm为99%±0.9%,CT PTV10mm为97.6%±1.4%(P=.03)。
  • 同侧乳腺V12.3Gy:MR PTV3mm为34.8%±12.7%,CT PTV10mm为44.4%±10.9%(P=.002)。
  • 胸壁V24Gy:MR PTV3mm为8.5±5.5 cm3,CT PTV10mm为21.8±14.9 cm3(P=.004)。

临床怎么用

  • 在3次APBI中,MR引导实时 cine MRI 靶区监测可支持3mm PTV外扩,获得与CT 3mm计划相当的靶区覆盖和OAR剂量。
  • 若临床采用10mm PTV外扩的CT计划,改用MR引导3mm外扩可改善同侧乳腺和胸壁剂量保护。

局限性

  • 剂量学差异的临床相关性尚不明确。
  • 研究仅纳入10例患者,为剂量学比较,未报告临床结局。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: External beam accelerated partial breast irradiation (APBI) is subject to treatment uncertainties that must be accounted for through planning target volume (PTV) margin. We hypothesize that magnetic resonance-guided radiation therapy with reduced PTV margins enabled by real-time cine magnetic resonance image (MRI) target monitoring results in better normal tissue sparing compared with computed tomography (CT)-guided radiation therapy with commonly used clinical PTV margins. In this study, we compare the plan quality of ViewRay MRIdian Linac forward planned intensity modulated radiation therapy and TrueBeam volumetric modulated arc therapy for a novel 3-fraction APBI schedule. METHODS AND MATERIALS: Targets and organs at risk (OARs) were segmented for 10 patients with breast cancer according to NSABP B39/RTOG 0413 protocol. A 3 mm margin was used to generate MR PTV3mm and CT PTV3mm plans, and a 10 mm margin was used for CT PTV10mm. An APBI schedule delivering 24.6 Gy to the clinical target volume and 23.4 Gy to the PTV in 3 fractions was used. OAR dose constraints were scaled down from existing 5-fraction APBI protocols. Target and OAR dose-volume metrics for the following data sets were analyzed using Wilcoxon matched-pairs signed-rank test: (1) MR PTV3mm versus CT PTV3mm plans and (2) MR PTV3mm versus CT PTV10mm. RESULTS: Average PTVs were 84.3 ± 51.9 cm3 and 82.6 ± 55 cm3 (P = .5) for MR PTV3mm and CT PTV3mm plans, respectively. PTV V23.4Gy, dose homogeneity index, conformity index (CI), and R50 were similar. There was no meaningful difference in OAR metrics, despite MR PTV3mm being larger than the CT PTV3mm in 70% of the patients. Average PTVs for MR PTV3mm and CT PTV10mm plans were 84.3 ± 51.9 cm3 and 131.7 ± 74.4 cm3, respectively (P = .002). PTV V23.4Gy was 99% ± 0.9% versus 97.6% ± 1.4% (P = .03) for MR PTV3mm and CT PTV10mm, respectively. Dose homogeneity index, CI, and R50 were similar. MR PTV3mm plans had better ipsilateral breast (V12.3Gy, 34.8% ± 12.7% vs 44.4% ± 10.9%, P = .002) and chest wall sparing (V24Gy, 8.5 ± 5.5 cm3 vs 21.8 ± 14.9 cm3, P = .004). CONCLUSIONS: MR- and CT-based planning systems produced comparable plans when a 3 mm PTV margin was used for both plans. As expected, MR PTV3mm plans produced better ipsilateral breast and chest wall sparing compared with CT PTV10mm. The clinical relevance of these differences in dosimetric parameters is not known.