头颈癌IMRT计划γ通过率与靶区内气腔及骨结构体积的相关性
结果显示AAA与AXB总体γ通过率无显著差异,但在气腔和骨结构中差异显著;使用AAA时γ通过率与靶区气腔体积自然对数正相关、与骨结构体积自然对数负相关。
要点
- AAA与AXB总体γ通过率无显著差异,但气腔和骨结构区域差异显著。
- 使用AAA时,靶区气腔体积小于约80 cc或骨结构体积大于约6 cc可使γ值低于95%。
- 使用AXB时,γ通过率与气腔或骨结构体积无显著关系。
核心结论
- 头颈IMRT临床QA中,使用不同剂量计算算法时应更关注靶区内气腔和骨结构体积对γ通过率的影响。
主要发现
- AAA与AXB的总体γ值分别为95.6 ± 1.9%和96.2 ± 1.7%(3%/2 mm标准,p > 0.05)。
- AAA与AXB在气腔的γ值分别为86.6 ± 9.4%和98.0 ± 1.7%,在骨结构分别为82.7 ± 13.5%和99.0 ± 1.7%(p < 0.05)。
- 使用AAA时,γ值与靶区气腔体积的自然对数成正比(R2 = 0.674),与骨结构体积的自然对数成反比(R2 = 0.816)。
- 使用AAA时,靶区气腔体积小于约80 cc或骨结构体积大于约6 cc,γ值低于95%。
- 使用AXB时,γ值与气腔或骨结构体积无显著关系。
关键数值与推荐
- γ通过率分析采用3%/2 mm标准。
- 使用AAA时,靶区气腔体积小于约80 cc或骨结构体积大于约6 cc对应γ值低于95%。
临床怎么用
- 头颈IMRT的QA中,使用AAA等算法时应评估靶区内气腔和骨结构体积对γ通过率的影响,必要时关注γ值下降风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Both patient-specific dose recalculation and γ passing rate analysis are important for the quality assurance (QA) of intensity modulated radiotherapy (IMRT) plans. The aim of this study was to analyse the correlation between the γ passing rates and the volumes of air cavities (Vair) and bony structures (Vbone) in target volume of head and neck cancer. METHODS: Twenty nasopharyngeal carcinoma and twenty nasal natural killer T-cell lymphoma patients were enrolled in this study. Nine-field sliding window IMRT plans were produced and the dose distributions were calculated by anisotropic analytical algorithm (AAA), Acuros XB algorithm (AXB) and SciMoCa based on the Monte Carlo (MC) technique. The dose distributions and γ passing rates of the targets, organs at risk, air cavities and bony structures were compared among the different algorithms. RESULTS: The γ values obtained with AAA and AXB were 95.6 ± 1.9% and 96.2 ± 1.7%, respectively, with 3%/2 mm criteria (p > 0.05). There were significant differences (p < 0.05) in the γ values between AAA and AXB in the air cavities (86.6 ± 9.4% vs. 98.0 ± 1.7%) and bony structures (82.7 ± 13.5% vs. 99.0 ± 1.7%). Using AAA, the γ values were proportional to the natural logarithm of Vair (R2 = 0.674) and inversely proportional to the natural logarithm of Vbone (R2 = 0.816). When the Vair in the targets was smaller than approximately 80 cc or the Vbone in the targets was larger than approximately 6 cc, the γ values of AAA were below 95%. Using AXB, no significant relationship was found between the γ values and Vair or Vbone. CONCLUSION: In clinical head and neck IMRT QA, greater attention should be paid to the effect of Vair and Vbone in the targets on the γ passing rates when using different dose calculation algorithms.
引用
- Shen Z, Tan X, Li S, et al. Correlation between the γ passing rates of IMRT plans and the volumes of air cavities and bony structures in head and neck cancer. Radiat Oncol. 2021. PMID: 34289863. doi:10.1186/s13014-021-01861-y