自适应4D锥形束CT首次人体实施:肺癌放疗中更短时间更低剂量

The first-in-human implementation of adaptive 4D cone beam CT for lung cancer radiotherapy: 4DCBCT in less time with less dose.
PubMed 文献 · Radiother Oncol · 2021-05-27

该研究首次在肺癌患者中实施自适应4D锥形束CT(4DCBCT),根据患者实时呼吸信号调节机架旋转速度和kV投影速率。在ADAPT试验中,前四例患者显示成像剂量降低85%,扫描时间缩短至73±12秒(较常规240秒减少70%),对比噪声比从9.2±3.9提升至11.7±4.1。

要点

  • 自适应4DCBCT根据实时呼吸信号调节硬件,显著缩短扫描时间并降低成像剂量。
  • 前四例患者中成像剂量降低85%,扫描时间减少70%至73±12秒。
  • 对比噪声比从常规9.2±3.9改善至11.7±4.1,图像质量提升。

核心结论

  • 自适应4DCBCT首次人体实施可行,可大幅减少肺癌放疗中的扫描时间和成像剂量。

研究设计与样本

  • ADAPT试验中前四例肺癌患者接受自适应4DCBCT扫描,每例在两个治疗分次采集,并与常规4分钟4DCBCT对比。

主要发现

  • ADAPT成像剂量降低85%,扫描时间为73±12秒,较常规240秒减少70%。
  • 对比噪声比从常规4DCBCT的9.2±3.9提升至ADAPT的11.7±4.1。

关键数值与推荐

  • ADAPT技术通过数学优化和外接电路实时调节机架旋转速度和kV投影速率。
  • 图像采用运动补偿重建算法重建。

临床怎么用

  • 该技术有望使肺癌放疗的4DCBCT成像更高效、剂量更低,但需进一步临床验证。

局限性

  • 仅报告前四例患者的初步结果,样本量小。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: We present the first implementation of Adaptive 4D cone beam CT (4DCBCT) that adapts the image hardware (gantry rotation speed and kV projections) in response to the patient's real-time respiratory signal. Adaptive 4DCBCT was applied on lung cancer patients to reduce the scan time and imaging dose in the ADaptive CT Acquisition for Personalised Thoracic imaging (ADAPT) trial. MATERIALS AND METHODS: The ADAPT technology measures the patient's real-time respiratory signal and uses mathematical optimisation and external circuitry attached to the linear accelerator to modulate the gantry rotation speed and kV projection rate to reduce scan times and imaging dose. For each patient, ADAPT scans were acquired on two treatment fractions and reconstructed with a motion compensated reconstruction algorithm and compared to the current state-of-the-art four-minute 4DCBCT acquisition (conventional 4DCBCT). We report on the scan time, imaging dose and image quality for the first four adaptive 4DCBCT patients. RESULTS: The ADAPT imaging dose was reduced by 85% and scan times were 73 ± 12 s representing a 70% reduction compared to the 240 s conventional 4DCBCT scan. The contrast-to-noise ratio was improved from 9.2 ± 3.9 with conventional 4DCBCT to 11.7 ± 4.1 with ADAPT. DISCUSSION: The ADAPT trial represents the first time that gantry rotation speed and projection acquisition have been adapted and optimised in real-time in response to changes in the patient's breathing. ADAPT demonstrates substantially reduced scan times and imaging dose for clinical 4DCBCT imaging that could enable more efficient and optimised lung cancer radiotherapy.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. O'Brien RT, Dillon O, Lau B, et al. The first-in-human implementation of adaptive 4D cone beam CT for lung cancer radiotherapy: 4DCBCT in less time with less dose. Radiother Oncol. 2021. PMID: 34052341. doi:10.1016/j.radonc.2021.05.021

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。