文献速递泌尿生殖与妇科肿瘤放疗放疗物理与技术

基于半自动烧入基准标记的前列腺放疗MR-only工作流程准确性

Accuracy of an MR-only workflow for prostate radiotherapy using semi-automatically burned-in fiducial markers.
PubMed 文献 · Radiat Oncol · 2021-02-19

该研究评估了使用半自动烧入基准标记(FM)的MR-only工作流程的准确性,并与CT/MR工作流程进行比较。

要点

  • MR-only工作流程靶区勾画图像配准的IOE比CT/MR工作流程小3倍
  • 所有观察者正确识别全部FM的患者比例为90%(18/20)
  • MR-only与CT/MR工作流程的CBCT配准在RL、CC和AP方向等效,仅旋转随机误差有显著差异

核心结论

  • 使用带烧入FM的sCT的MR-only工作流程优于当前CT/MR工作流程,CT-MR配准的观察者间误差小3倍,且CT与sCT参考扫描的CBCT配准结果相当。

研究设计与样本

  • 在20例患者上确定基于FM的CT-to-MR图像配准和基于软组织的MR-to-MR图像配准的观察者间误差(IOE)。
  • 在20例患者上确定所得FM位置的观察者间变异。
  • 在26例患者上回顾性将CBCT配准到带烧入FM的sCT,并与CT-CBCT配准比较。

主要发现

  • MR-only工作流程靶区勾画图像配准的IOE比CT/MR工作流程小3倍。
  • 所有观察者正确识别全部FM的患者为18/20(90%)。
  • FM质心(COM)位置的IOE在CC方向在CT层厚(2.5 mm)以内,AP方向低于1.0 mm,RL方向低于1.5 mm。
  • MR-only工作流程与CT/MR工作流程的IGRT位置验证配准在RL、CC和AP方向等效,但旋转随机误差有显著差异。

关键数值与推荐

  • FM质心(COM)位置的IOE在CC方向在CT层厚(2.5 mm)以内。
  • FM质心(COM)位置的IOE在AP方向低于1.0 mm,在RL方向低于1.5 mm。
  • 所有观察者正确识别全部FM的患者为18/20(90%)。

临床怎么用

  • MR-only工作流程使用sCT替代CT,可减少放疗链条中的系统误差,并避免CT扫描。
  • 半自动烧入FM到sCT的方法可行,且基于FM的配准精度高,可用于靶区勾画和IGRT位置验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The benefit of MR-only workflow compared to current CT-based workflow for prostate radiotherapy is reduction of systematic errors in the radiotherapy chain by 2-3 mm. Nowadays, MRI is used for target delineation while CT is needed for position verification. In MR-only workflows, MRI based synthetic CT (sCT) replaces CT. Intraprostatic fiducial markers (FMs) are used as a surrogate for the position of the prostate improving targeting. However, FMs are not visible on sCT. Therefore, a semi-automatic method for burning-in FMs on sCT was developed. Accuracy of MR-only workflow using semi-automatically burned-in FMs was assessed and compared to CT/MR workflow. METHODS: Thirty-one prostate cancer patients receiving radiotherapy, underwent an additional MR sequence (mDIXON) to create an sCT for MR-only workflow simulation. Three sources of accuracy in the CT/MR- and MR-only workflow were investigated. To compare image registrations for target delineation, the inter-observer error (IOE) of FM-based CT-to-MR image registrations and soft-tissue-based MR-to-MR image registrations were determined on twenty patients. Secondly, the inter-observer variation of the resulting FM positions was determined on twenty patients. Thirdly, on 26 patients CBCTs were retrospectively registered on sCT with burned-in FMs and compared to CT-CBCT registrations. RESULTS: Image registration for target delineation shows a three times smaller IOE for MR-only workflow compared to CT/MR workflow. All observers agreed in correctly identifying all FMs for 18 out of 20 patients (90%). The IOE in CC direction of the center of mass (COM) position of the markers was within the CT slice thickness (2.5 mm), the IOE in AP and RL direction were below 1.0 mm and 1.5 mm, respectively. Registrations for IGRT position verification in MR-only workflow compared to CT/MR workflow were equivalent in RL-, CC- and AP-direction, except for a significant difference for random error in rotation. CONCLUSIONS: MR-only workflow using sCT with burned-in FMs is an improvement compared to the current CT/MR workflow, with a three times smaller inter observer error in CT-MR registration and comparable CBCT registration results between CT and sCT reference scans. Trial registry Medical Research Involving Human Subjects Act (WMO) does apply to this study and was approved by the Medical Ethics review Committee of the Academic Medical Center. Registration number: NL65414.018.18. Date of registration: 21-08-2018.

来源:Radiat Oncol,查看原文。版权归原作者所有。

引用

  1. Goudschaal K, Beeksma F, Boon M, et al. Accuracy of an MR-only workflow for prostate radiotherapy using semi-automatically burned-in fiducial markers. Radiat Oncol. 2021. PMID: 33608008. doi:10.1186/s13014-021-01768-8

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。