spirometer引导屏气乳腺VMAT在Halcyon上经门静脉成像与表面追踪验证

Spirometer-guided breath-hold breast VMAT verified with portal images and surface tracking.
PubMed 文献 · Radiother Oncol · 2021-01-27

研究评估了7例左侧乳腺癌患者在Halcyon环形机架直线加速器上使用SDX spirometer引导屏气进行VMAT放疗的可重复性与稳定性,采用双深度相机表面扫描和门静脉成像监测呼吸运动。结果显示屏气幅度系统与随机变异性在AP方向低于0.7 mm、SI方向低于1.2 mm(表面监测),门静脉成像误差AP低于1.2 mm、SI低于2.1 mm,75%分次仅需4次屏气完成,82%在10分钟内完成。

要点

  • spirometer引导屏气乳腺VMAT在闭环机架直线加速器上可准确快速完成。
  • 表面监测显示屏气幅度变异性AP<0.7 mm、SI<1.2 mm。
  • 门静脉成像与表面监测偏差相关性有限(R²=0.27/0.38)。

核心结论

  • spirometer引导屏气乳腺VMAT在闭环机架直线加速器上可准确快速实施,腔内表面扫描是监测呼吸运动的有价值技术。

研究设计与样本

  • 前瞻性评估7例左侧乳腺癌患者,采用SDX spirometer屏气配合整合推量VMAT协议在Halcyon治疗,双深度相机表面扫描监测左乳,门静脉成像于切线机架角手动配准规划CT,分析分次间、分次内及屏气内运动,并记录总时间和屏气次数。

主要发现

  • 表面监测(N=130)显示屏气幅度系统与随机变异性AP<0.7 mm、SI<1.2 mm。
  • 门静脉成像(N=140)显示系统与随机误差AP<1.2 mm、SI<2.1 mm。
  • 门静脉成像与表面监测偏差在SI和AP轴相关性有限(R²=0.27/0.38,p<0.01)。
  • 75%分次使用4次屏气完成,82%分次在10分钟内完成。

临床怎么用

  • 在闭环机架直线加速器上可采用spirometer引导屏气VMAT进行左侧乳腺癌放疗,并利用腔内表面扫描监测呼吸运动。
  • 门静脉成像与表面监测偏差相关性有限,提示需结合门静脉成像验证屏气位置。

局限性

  • 样本量仅7例左侧乳腺癌患者,结果需更大样本验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: Fast rotating closed-bore gantry linacs are ideally suited for breath-hold treatments due to reduced imaging and delivery times. We evaluated the reproducibility and stability of spirometer-guided breath-hold breast treatments, using intra-bore surface monitoring and portal imaging on Halcyon (Varian Medical Systems). MATERIALS AND METHODS: Seven left-sided breast cancer patients were treated in breath-hold using the SDX spirometer (Dyn'R) with an integrated boost volumetric arc protocol on Halcyon. A dual depth-camera surface scanning system monitored the left breast. The interfraction, intrafraction and intrabreath-hold motion was determined in the anterior-posterior (AP) and superior-inferior (SI) direction. Portal images (PI), acquired at a tangential gantry angle were manually registered to the planning-CT to determine inter- and intrafraction breath-hold errors for the SI and tangential-anterior-posterior ("AP") axis. Correlations between PI and surface imaging deviations were investigated. To evaluate workflow efficiency, the total time and the number of breath-holds were recorded. RESULTS: Systematic and random variability of breath-hold amplitude was below 0.7 mm for the AP and below 1.2 mm for the SI component as detected by surface monitoring (N = 130). Systematic and random errors retrieved from portal images (N = 140) were below 1.2 mm for the "AP" and 2.1 mm for SI axis. A limited correlation was found between PI and surface monitoring deviations for both the SI and "AP" axes (R2 = 0.27/0.38, p < 0.01). 75% of fractions were completed using four breath-holds and 82% within 10 min. CONCLUSION: Surface imaging indicated spirometer-guided breath-hold VMAT breast radiotherapy can be accurately and quickly performed on a closed-bore gantry linac. Intra-bore surface scanning proved a valuable technique for monitoring breathing motion in closed-bore systems.