分次内前列腺旋转对门控、多叶准直器追踪及治疗床追踪剂量学影响的端到端验证
本研究利用动态骨盆模体结合放射性变色胶片剂量测量,端到端评估门控、MLC追踪和治疗床追踪管理分次内前列腺旋转的剂量学精度。无补偿时旋转使靶区γ通过率中位下降33.2%,三种补偿技术均改善γ通过率和靶区覆盖,但残余旋转相关剂量误差仍存在。
要点
- 分次内旋转无补偿时靶区γ通过率中位下降33.2%。
- 门控、MLC追踪、治疗床追踪分别改善γ通过率13.2%、6.0%、11.1%。
- 三种技术均改善靶区覆盖,门控对危及器官保护更优。
核心结论
- 门控、MLC追踪和治疗床追踪均能改善分次内前列腺旋转下的剂量学精度,但因缺乏旋转校正而仍存在残余运动相关剂量误差。
研究设计与样本
- 使用动态骨盆模体进行端到端测试,模体施加内部横向旋转(最高25°)及耦合垂直和纵向平移,以放射性变色胶片堆叠测量剂量。
- 针对前列腺SBRT伴局灶加量的治疗计划,在三种典型前列腺运动轨迹下分别评估无补偿、门控、MLC追踪和治疗床追踪四种策略。
主要发现
- 无运动补偿时,分次内旋转使前列腺靶区3%(global)/2mm γ通过率中位(范围)下降-33.2%(-68.6%,-4.1%)。
- 运动补偿改善γ通过率:门控13.2%(-0.4%,32.9%),MLC追踪6.0%(-0.8%,27.7%),治疗床追踪11.1%(1.2%,22.9%)。
- 三种补偿技术多数情况下改善靶区覆盖,门控对危及器官保护优于MLC追踪或治疗床追踪。
临床怎么用
- 在前列腺SBRT中应考虑分次内旋转对剂量的影响,门控、MLC追踪或治疗床追踪可用于改善剂量精度,但需注意残余旋转相关误差。
局限性
- 研究为模体实验,未涉及临床患者;三种技术均未进行旋转校正,残余旋转相关剂量误差仍存在。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND PURPOSE: Both gating and tracking can mitigate the deteriorating dosimetric impact of intrafraction translation during prostate stereotactic body radiotherapy (SBRT). However, their ability to manage intrafraction rotation has not yet been thoroughly investigated. The dosimetric accuracy of gating, MLC tracking and couch tracking to manage intrafraction prostate rotation was investigated. MATERIALS AND METHODS: Treatment plans for end-to-end tests of prostate SBRT with focal boosting were generated for a dynamic anthropomorphic pelvis phantom. The phantom applied internal lateral rotation (up to 25°) and coupled vertical and longitudinal translation of a radiochromic film stack that was used for dose measurements. Dose was delivered for each plan while the phantom applied motion according to three typical prostate motion traces without compensation (i), with gating (ii), with MLC tracking (iii) or with couch tracking (iv). Measured doses for the four motion compensation strategies were compared with the planned dose in terms of γ-index analysis, target coverage and organs at risk (OAR) sparing. RESULTS: Intrafraction rotation reduced the 3%(global)/2mm γ-index passing rate (γPR) for the prostate target volume by median (range) -33.2% (-68.6%, -4.1%) when no motion compensation was applied. The use of motion compensation improved the γPR by 13.2% (-0.4%, 32.9%) for gating, by 6.0% (-0.8%, 27.7%) for MLC tracking and by 11.1% (1.2%, 22.9%) for couch tracking. The three compensation techniques improved the target coverage in most cases. Gating showed better OAR sparing than MLC tracking or couch tracking. CONCLUSIONS: Compensation of intrafraction prostate rotation with gating, MLC tracking and couch tracking was investigated experimentally for the first time. All three techniques improved the dosimetric accuracy, but residual motion-related dose errors remained due to the lack of rotation correction.
引用
- De Roover R, Hansen R, Crijns W, et al. Dosimetric impact of intrafraction prostate rotation and accuracy of gating, multi-leaf collimator tracking and couch tracking to manage rotation: An end-to-end validation using volumetric film measurements. Radiother Oncol. 2020. PMID: 33264640. doi:10.1016/j.radonc.2020.11.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。