兆伏级拓扑图用于螺旋断层放疗快速患者对准的临床开发与评估
本研究开发并评估了一种基于兆伏级(MV)拓扑图的快速患者定位工具,用于螺旋断层放疗。结果显示MV拓扑图与MVCT临床性能等效,且采集时间显著更短,可作为基于骨性标志患者对准的替代或补充工具。
要点
- MV拓扑图与MVCT在骨盆和头颈患者中平移对准差异等效。
- 一对MV拓扑图采集时间最多比MVCT(粗扫模式)少12.7倍,覆盖更长纵向长度。
- MV拓扑图可作为螺旋断层放疗中基于骨性标志患者对准的替代或补充工具。
核心结论
- MV拓扑图在骨盆和头颈患者中显示出与标准MVCT等效的临床性能,且采集时间显著更短。
研究设计与样本
- MV拓扑图成像协议在机架角0度和90度静态采集2个正交图像,离线重建并与计划CT生成的数字重建拓扑图配准。
主要发现
- 骨盆患者(MVCT减MV拓扑图)对准差异幅度(标准差)为-0.3±2.1、-0.8±2.4、1.6±1.7 mm。
- 头颈患者(MVCT减MV拓扑图)对准差异幅度(标准差)为0.6±1.2、0.8±4.2、1.6±2.6 mm。
- 两种成像技术线性回归系数在骨盆横向、纵向、垂直方向分别为1.18、1.10、0.94,头颈分别为0.86、0.63、0.38。
- 一对MV拓扑图采集时间最多比MVCT(粗扫模式)少12.7倍,同时覆盖更长纵向长度。
临床怎么用
- MV拓扑图可用于螺旋断层放疗中基于骨性标志的患者快速对准,作为MVCT的替代或补充工具。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To develop and evaluate a fast patient localization tool using megavoltage (MV)-topogram on helical tomotherapy. METHODS AND MATERIALS: Eighty-one MV-topogram pairs for 18 pelvis patients undergoing radiation were acquired weekly under an institutional review board-approved clinical trial. The MV-topogram imaging protocol requires 2 orthogonal acquisitions at static gantry angles of 0 degrees and 90 degrees for a programed scan length. A MATLAB based in-house software was developed to reconstruct the MV-topograms offline. Reference images (digitally reconstructed topograms, digitally reconstructed topograms) were generated using the planning computed tomography and tomotherapy geometry. The MV-topogram based alignment was determined by registering the MV-topograms to the digitally reconstructed topogram using bony landmark on commercial MIM software. The daily shifts in 3 translational directions determined from MV-topograms were compared with the megavoltage computed tomography (MVCT) based patient shifts. Linear-regression and two one-sided tests equivalence tests were performed to investigate the relation and equivalence between the 2 techniques. Seventy-eight MV-topogram pairs for 19 head and neck patients were included to validate the finding. RESULTS: The magnitudes of alignment differences of (MVCT - MV-topogram) (and standard deviations) were -0.3 ± 2.1, -0.8 ± 2.4, and 1.6 ± 1.7 mm for pelvis and 0.6 ± 1.2, 0.8 ± 4.2, 1.6 ± 2.6 mm for head and neck; the linear-regression coefficients between 2 imaging techniques were 1.18, 1.10, 0.94, and 0.86, 0.63, 0.38 in the lateral, longitudinal, vertical directions for pelvis and head and neck, respectively. The acquisition time for a pair of MV-topograms was up to 12.7 times less than MVCT scans (coarse scan mode) while covering longer longitudinal length. CONCLUSIONS: MV-topograms showed equivalent clinical performance to the standard MVCT with significantly less acquisition time for pelvis and H&N patients. The MV-topogram can be used as an alternative or complimentary tool for bony landmark-based patient alignment on tomotherapy.
引用
- Qi XS, Chu FI, Zhang Z, et al. Clinical Development and Evaluation of Megavoltage Topogram for Fast Patient Alignment on Helical Tomotherapy. Adv Radiat Oncol. 2020. PMID: 33305096. doi:10.1016/j.adro.2020.05.014