低剂量锥形束CT双字典学习重建在头颈放疗图像引导中的评估
本研究回顾性收集42例头颈肿瘤患者的62套CBCT图像,采用SART-DDL算法在减少18%临床原始投影条件下重建低剂量CBCT(D-CBCT),并与SART-TV算法及临床F-CBCT比较。结果显示D-CBCT峰值信噪比较SART-TV高1.7 dB(P<.001),主观评分与F-CBCT无显著差异(P>.05),配准最大平移和旋转差异分别为1.8 mm和1.7度,几何保真度可接受。
要点
- SART-DDL可在减少18%投影条件下重建头颈CBCT,图像质量与临床F-CBCT相当。
- D-CBCT峰值信噪比较SART-TV高1.7 dB(36.36±0.55 dB vs 34.68±0.28 dB,P<.001)。
- D-CBCT与F-CBCT配准最大平移差异1.8 mm、最大旋转差异1.7度。
核心结论
- SART-DDL重建的D-CBCT在头颈患者中图像质量、几何保真度和临床应用能力与F-CBCT相当,且投影数减少18%。
研究设计与样本
- 回顾性研究,随机收集2018年2月至7月四川大学华西医院42例头颈肿瘤患者(平均年龄49.7岁,标准差11.4岁,女12例、男30例)的62套CBCT图像。
- 所有图像集用SART-DDL以减少18%的临床原始投影重建为D-CBCT,并与SART-TV算法定量比较(配对t检验)。
- 由两名肿瘤科医生盲法对D-CBCT与临床F-CBCT进行五级主观评分(配对Wilcoxon秩和检验),并比较两者配准结果及D-CBCT几何保真度。
主要发现
- D-CBCT平均峰值信噪比较SART-TV重建高1.7 dB(36.36±0.55 dB vs 34.68±0.28 dB,P<.001)。
- 所有D-CBCT图像被认定为临床可接受,主观评分与F-CBCT无显著差异(P>.05)。
- 临床配准中D-CBCT与F-CBCT最大平移差异为1.8 mm,最大旋转差异为1.7度。
- D-CBCT的水平、垂直和矢状几何保真度均可接受。
关键数值与推荐
- D-CBCT峰值信噪比36.36±0.55 dB,SART-TV为34.68±0.28 dB,差值1.7 dB(P<.001)。
- D-CBCT与F-CBCT主观评分比较P>.05。
- D-CBCT与F-CBCT配准最大平移差异1.8 mm、最大旋转差异1.7度。
临床怎么用
- SART-DDL可在减少18%投影的条件下用于头颈IGRT的CBCT重建,图像质量与临床F-CBCT相当。
- D-CBCT与F-CBCT配准差异在平移1.8 mm、旋转1.7度以内,几何保真度可接受。
局限性
- 本研究为回顾性单中心研究,仅纳入头颈肿瘤患者,样本量为42例患者共62套图像。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: To develop a low-dose cone beam CT (LD-CBCT) reconstruction method named simultaneous algebraic reconstruction technique and dual-dictionary learning (SART-DDL) joint algorithm for image guided radiation therapy (IGRT) and evaluate its imaging quality and clinical application ability. METHODS: In this retrospective study, 62 CBCT image sets from February 2018 to July 2018 at west china hospital were randomly collected from 42 head and neck patients (mean [standard deviation] age, 49.7 [11.4] years, 12 females and 30 males). All image sets were retrospectively reconstructed by SART-DDL (resultant D-CBCT image sets) with 18% less clinical raw projections. Reconstruction quality was evaluated by quantitative parameters compared with SART and Total Variation minimization (SART-TV) joint reconstruction algorithm with paired t test. Five-grade subjective grading evaluations were done by two oncologists in a blind manner compared with clinically used Feldkamp-Davis-Kress algorithm CBCT images (resultant F-CBCT image sets) and the grading results were compared by paired Wilcoxon rank test. Registration results between D-CBCT and F-CBCT were compared. D-CBCT image geometry fidelity was tested. RESULTS: The mean peak signal to noise ratio of D-CBCT was 1.7 dB higher than SART-TV reconstructions (P < .001, SART-DDL vs SART-TV, 36.36 ± 0.55 dB vs 34.68 ± 0.28 dB). All D-CBCT images were recognized as clinically acceptable without significant difference with F-CBCT in subjective grading (P > .05). In clinical registration, the maximum translational and rotational difference was 1.8 mm and 1.7 degree respectively. The horizontal, vertical and sagittal geometry fidelity of D-CBCT were acceptable. CONCLUSIONS: The image quality, geometry fidelity and clinical application ability of D-CBCT are comparable to that of the F-CBCT for head-and-neck patients with 18% less projections by SART-DDL.