颅内放疗专用固定装置的分次内与分次间分析
本研究评估了新型颅内放疗固定装置在两家机构中的分次内与分次间运动、患者舒适度及放射技师信心。结果显示分次间平移运动均值<2 mm(SD<4 mm)、旋转<0.5°(SD<1.5°),分次内平移<0.4 mm(SD<0.6 mm)、旋转0.3°(SD<0.5°);患者舒适度良好,技师模具准备中位时间9 min(范围6–12 min)。
要点
- 分次间平移运动均值<2 mm(SD<4 mm),旋转<0.5°(SD<1.5°)。
- 分次内平移运动均值<0.4 mm(SD<0.6 mm),旋转0.3°(SD<0.5°)。
- 患者舒适度良好,技师模具准备中位时间9 min(范围6–12 min)。
核心结论
- 该固定装置在颅内放疗中有效可行,结合每日在线影像可实现颅内及立体定向治疗所需的亚毫米级精度。
研究设计与样本
- 多中心研究,纳入荷兰HagaZiekenhuis 17例脑转移瘤患者和意大利IRCCS Ospedale Sacro Cuore Don Calabria 16例患者(10例脑转移瘤、6例原发性中枢神经系统肿瘤)。
- 中位靶体积在IRCCS为436 cc(范围3.2–1628 cc),在Haga为4.58 cc(范围0.4–27.19 cc)。
- IRCCS中位处方剂量30 Gy(范围27–60 Gy),中位分次数10次(范围3–30次);Haga中位处方剂量21 Gy(范围8–21 Gy),中位分次数1次(范围1–3次)。
- CT模拟时组装固定装置并访谈患者舒适度,记录模拟摆位时间和技师操作;治疗前CBCT与计划CT行颅骨自动刚性配准并移床,治疗后额外CBCT与治疗前CBCT配准识别分次内运动。
- 共126次CBCT由放疗医师/医学物理师离线分析,在线配准均由经验丰富的放疗医师或技师验证。
主要发现
- 所有患者对固定装置舒适度反应积极,报告良好至高度满意。
- 技师对锁定机制和卡扣的使用反应积极。
- 两家机构定位不确定度结果可比,分次间平移运动均值<2 mm(SD<4 mm),旋转<0.5°(SD<1.5°);分次内平移运动<0.4 mm(SD<0.6 mm),旋转0.3°(SD<0.5°)。
关键数值与推荐
- 分次间平移运动均值<2 mm(SD<4 mm),旋转<0.5°(SD<1.5°)。
- 分次内平移运动均值<0.4 mm(SD<0.6 mm),旋转0.3°(SD<0.5°)。
- 技师组装固定装置中位时间9 min(范围6–12 min)。
- 共分析126次CBCT。
临床怎么用
- 该固定装置可结合每日在线CBCT影像用于颅内及立体定向放疗,以实现亚毫米级定位精度。
- 患者舒适度和技师准备时间可接受,支持该装置在临床日常实践中使用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Immobilization devices are crucial to minimize patient positioning uncertainties in radiotherapy (RT) treatments. Accurate inter and intra-fraction motions is particularly important for intracranial and stereotactic radiation treatment which require high precision in dose delivery. Recently, a new immobilization device has been developed specifically for the radiation treatment of intracranial malignancies. To date, no data are available on the use of this device in daily clinical practice. The aim of this study is to investigate the intra and inter-fraction variations, patient comfort and radiographer confidence of the immobilization system from two distinct institutions: HagaZiekenhuis, Den Haag, Netherlands and IRCCS Ospedale Sacro Cuore Don Calabria, Negrar, Italy. MATERIAL AND METHOD: Sixteen patients (10 diagnosed with brain metastases and 6 with primary central nervous systemic tumor) from IRCCS Ospedale Sacro Cuore Don Calabria and 17 patients (all diagnosed with brain metastases tumor) from HagaZiekenhuis were included in this study. The median target volume was 436 cc (range 3.2-1628 cc) and 4.58 cc (range 0.4-27.19 cc) for IRCCS and Haga, respectively. For patients treated in IRCCS Sacro Cuore Don Calabria, the median dose prescription was 30 Gy (range 27-60 Gy) and median number of fractions 10 (range 3-30). In Haga the median dose prescription was 21 Gy (range 8-21 Gy) and the median number of fraction was 1 (range 1-3). The immobilization device was assembled during CT simulation. A short interview to the patient regarding the device's comfort level was conducted at the end of the simulation procedure. Additionally, simulation setup time and radiographer (RTT) procedures (i.e. mask preparation) were evaluated. Prior to radiation treatment delivery, an automatic rigid match on the cranial bones between cone beam computed tomography (CBCT) and planning-CT was performed. A couch shift was performed subsequently. An extra post-treatment CBCT was acquire after the treatment delivery. This post-treatment CBCT was matched with pre-treatment CBCT to identify any possible intra-fraction motion. All online matches were validated by experienced radiation oncologist or RTT. A total of 126 CBCT's were analyzed offline by radiation oncologist/medical physicist. The data of the pre-treatment CBCT match was used to quantify inter-fraction motion. The post-treatment CBCT was matched with pre-treatment CBCT to identify any possible intra-fraction motion. RESULTS: During the molding of the mask, all patients responded positive to the comfort. Median time required by the RTTs to assemble the immobilization system was 9 min (range 6-12 min). In terms of comfort, all patients reported a good-to high level of satisfaction. The RTTs also respond positively towards the use of the locking mechanism and clips. Results of positioning uncertainties were comparable between the two institutes. The mean inter-fraction motion for all translational and rotational directions were < 2 mm (SD < 4 mm) and < 0.5°(SD < 1.5°), respectively, while the mean intra-fraction motions were < 0.4 mm (SD < 0.6 mm) and 0.3° (SD < 0.5°). CONCLUSIONS: This study demonstrates the efficacy and feasibility of the immobilization device in the intracranial radiation treatment. Both patient comfort and preparation time by RTTs are considered adequate. In combination with online daily imaging procedure, this device can achieve submillimeter accuracy required for intracranial and stereotactic treatments.