文献速递泌尿生殖与妇科肿瘤放疗放疗物理与技术

1.5T MR-Linac上MR引导放疗期间前列腺分次内运动评估

Prostate intrafraction motion during the preparation and delivery of MR-guided radiotherapy sessions on a 1.5T MR-Linac.
PubMed 文献 · Radiother Oncol · 2020-07-03

本研究在1.5T MR-Linac上对5例低中危前列腺癌患者行20×3.1 Gy在线自适应放疗,每分次全程采集3D cine-MR并用软组织对比追踪算法自动测定前列腺分次内运动。结果显示平移95%置信区间均在临床应用的5 mm margins内,仅4/100分次需要针对分次内运动进行计划自适应,提示5 mm margins足够,并可能通过cine-MR引导的分次内计划自适应进一步缩小。

要点

  • 5例低中危前列腺癌患者行20×3.1 Gy在线自适应MR-Linac放疗,全程cine-MR自动追踪前列腺运动。
  • 前列腺平移95%置信区间均在临床5 mm margins内,仅4/100分次需分次内计划自适应。
  • cine-MR引导的分次内计划自适应可能进一步缩小margins。

核心结论

  • 在1.5T MR-Linac上全程cine-MR追踪前列腺分次内运动是可行的,临床5 mm margins对这些治疗足够,并可能通过cine-MR引导的分次内计划自适应进一步缩小。

研究设计与样本

  • 5例低中危前列腺癌患者接受20×3.1 Gy分次在线自适应MR-Linac放疗,每分次开始采集初始MRI,传播并适配计划MRI轮廓后重新优化计划并照射,beam-on同时启动3D cine-MR采集至全程结束,用既往验证的软组织对比追踪算法测定前列腺分次内运动,并用4D phantom确定方法绝对精度。

主要发现

  • 在约45 min全on-couch期间完全自动测定前列腺运动,未发现误配准。
  • 前列腺平移95%置信区间均在临床应用的5 mm margins内。
  • 100分次中仅4分次需要针对分次内运动进行计划自适应。
  • 使用4D phantom确定了该方法的绝对精度。

临床怎么用

  • 本研究表明在1.5T MR-Linac在线自适应前列腺放疗中,临床应用的5 mm margins足以覆盖分次内运动。
  • 可考虑利用cine-MR引导的分次内计划自适应进一步缩小margins。

局限性

  • 研究仅纳入5例低中危前列腺癌患者,样本量小。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate prostate intrafraction motion using MRI during the full course of online adaptive MR-Linac radiotherapy (RT) fractions, in preparation of MR-guided extremely hypofractionated RT. MATERIAL AND METHODS: Five low and intermediate risk prostate cancer patients were treated with 20 × 3.1 Gy fractions on a 1.5T MR-Linac. Each fraction, initial MRI (Pre) scans were obtained at the start of every treatment session. Pre-treatment planning MRI contours were propagated and adapted to this Pre scan after which plan re-optimization was started in the treatment planning system followed by dose delivery. 3D Cine-MR imaging was started simultaneously with beam-on and acquired over the full beam-on period. Prostate intrafraction motion in this cine-MR was determined with a previously validated soft-tissue contrast based tracking algorithm. In addition, absolute accuracy of the method was determined using a 4D phantom. RESULTS: Prostate motion was completely automatically determined over the full on-couch period (approx. 45 min) with no identified mis-registrations. The translation 95% confidence intervals are within clinically applied margins of 5 mm, and plan adaption for intrafraction motion was required in only 4 out of 100 fractions. CONCLUSION: This is the first study to investigate prostate intrafraction motions during entire MR-guided RT sessions on an MR-Linac. We have shown that high quality 3D cine-MR imaging and prostate tracking during RT is feasible with beam-on. The clinically applied margins of 5 mm have proven to be sufficient for these treatments and may potentially be further reduced using intrafraction plan adaptation guided by cine-MR imaging.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. de Muinck Keizer DM, Kerkmeijer LGW, Willigenburg T, et al. Prostate intrafraction motion during the preparation and delivery of MR-guided radiotherapy sessions on a 1.5T MR-Linac. Radiother Oncol. 2020. PMID: 32622779. doi:10.1016/j.radonc.2020.06.044