IROC肺与脊柱模体照射失败模式差异分析

Differences in the Patterns of Failure Between IROC Lung and Spine Phantom Irradiations.
PubMed 文献 · Pract Radiat Oncol · 2020-05-13

该研究分析2012年1月至2018年12月间IROC移动肺模体和SBRT脊柱模体照射失败案例,肺模体1052例中116例失败,脊柱模体263例中42例失败。两种模体失败模式不同,均具临床相关性,凸显独立端到端剂量审计的价值。

要点

  • 脊柱模体失败主因是系统性剂量误差,占60%。
  • 两种模体失败模式不同,均可能反映患者治疗中的误差。

核心结论

  • IROC移动肺模体与SBRT脊柱模体存在两种不同的失败模式,肺模体多数因定位误差失败,脊柱模体多数因系统性剂量误差失败。
  • 这些误差具有临床相关性,可能在实际患者治疗中表现为错误,凸显独立端到端剂量审计的价值。

研究设计与样本

  • 回顾性分析2012年1月至2018年12月所有IROC移动肺模体和SBRT脊柱模体照射失败记录,失败定义为机构未达到IROC计划剂量与实际剂量一致性的既定标准。
  • 共分析158例失败案例:肺模体1052例照射中116例失败,脊柱模体263例照射中42例失败;分析报告包括点剂量不一致、剂量剖面和伽马分析。

主要发现

  • 肺模体失败需要7个类别描述,脊柱模体失败需要4个类别描述;两种模体均存在系统性剂量误差和定位误差。
  • 脊柱模体失败中多数(60%)为系统性剂量误差,任何方向的定位误差仅占14%。

临床怎么用

  • 放射治疗科应关注肺SBRT中头脚方向定位误差的管控,以及脊柱SBRT中系统性剂量误差的管控。
  • 独立端到端剂量审计有助于发现社区放疗中出现的误差类型,可指导质量改进。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Our purpose was to investigate and classify the reasons why institutions fail the Imaging and Radiation Oncology Core (IROC) stereotactic body radiation therapy (SBRT) spine and moving lung phantoms, which are used to credential institutions for clinical trial participation. METHODS AND MATERIALS: All IROC moving lung and SBRT spine phantom irradiation failures recorded from January 2012 to December 2018 were evaluated in this study. A failure was a case where the institution did not meet the established IROC criteria for agreement between planned and delivered dose. We analyzed the reports for all failing irradiations, including point dose disagreement, dose profiles, and gamma analyses. Classes of failure patterns were created and used to categorize each instance. RESULTS: There were 158 failing cases analyzed: 116 of 1052 total lung irradiations and 42 of 263 total spine irradiations. Seven categories were required to describe the lung phantom failures, whereas 4 were required for the spine. Types of errors present in both phantom groups included systematic dose and localization errors. Fifty percent of lung failures were due to a superior-inferior localization error, that is, error in the direction of major motion. Systematic dose errors, however, contributed to only 22% of lung failures. In contrast, the majority (60%) of spine phantom failures were due to systematic dose errors, with localization errors (in any direction) accounting for only 14% of failures. CONCLUSIONS: There were 2 distinct patterns of failure between the IROC moving lung and SBRT spine phantoms. The majority of the lung phantom failures were due to localization errors, whereas the spine phantom failures were largely attributed to systematic dose errors. Both of these errors are clinically relevant and could manifest as errors in patient cases. These findings highlight the value of independent end-to-end dosimetry audits and can help guide the community in improving the quality of radiation therapy by focusing attention on where errors manifest in the community.

来源:Pract Radiat Oncol,查看原文。版权归原作者所有。

引用

  1. Edward SS, Alvarez PE, Taylor PA, et al. Differences in the Patterns of Failure Between IROC Lung and Spine Phantom Irradiations. Pract Radiat Oncol. 2020. PMID: 32413413. doi:10.1016/j.prro.2020.04.004

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。