左侧乳腺癌6013次深吸气屏气的稳定性与重复性
本研究分析103例左侧乳腺癌患者接受表面引导深吸气屏气(DIBH)放疗的6013次屏气,评估屏气内稳定性和分次间重复性。屏气水平中位标准差0.3 mm,分次间呼吸幅度最大差异平均1.3 mm,提示光学表面扫描可实现稳定可靠的DIBH治疗。
要点
- 屏气内稳定性好:屏气水平中位标准差0.3 mm,中位绝对线性幅度偏差0.4 mm。
- 分次间重复性良好:所有患者所有分次呼吸幅度最大差异平均1.3 mm。
- 表面引导系统联合视听反馈与自动门控可实现稳定可靠的DIBH放疗。
核心结论
- 光学表面扫描的临床整合使左侧乳腺癌表面引导DIBH放疗在常规临床中实现稳定可靠的治疗实施。
研究设计与样本
- 分析103例接受左侧乳腺癌放疗并使用表面引导DIBH技术的患者,每治疗分次连续测量患者垂直运动,通过视听反馈和自动门控控制束流。
主要发现
- 共分析束流开启期间的6013次屏气。
- 门控窗距自由呼吸基线(最大呼气)的平均幅度为15.8 mm(95%CI 8.5-30.6 mm),宽度为3.5 mm(95%CI 2-4.3 mm)。
- 屏气内稳定性指标:DIBH期间屏气水平的中位标准差为0.3 mm(95%CI 0.1-0.9 mm),中位绝对屏气内线性幅度偏差为0.4 mm(95%CI 0.01-2.1 mm)。
- 重复性测试显示分次间可靠性良好,所有患者所有分次呼吸幅度的最大差异平均为1.3 mm(95%CI 0.5-2.6 mm)。
临床怎么用
- 临床可采用光学表面扫描联合视听反馈与自动门控实施左侧乳腺癌DIBH放疗,以获得稳定的屏气水平和良好的分次间重复性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Patients with left-sided breast cancer frequently receive deep inspiration breath-hold (DIBH) radiotherapy to reduce the risk of cardiac side effects. The aim of the present study was to analyze intra-breath-hold stability and inter-fraction breath-hold reproducibility in clinical practice. MATERIAL AND METHODS: Overall, we analyzed 103 patients receiving left-sided breast cancer radiotherapy using a surface-guided DIBH technique. During each treatment session the vertical motion of the patient was continuously measured by a surface guided radiation therapy (SGRT) system and automated gating control (beam on/off) was performed using an audio-visual patient feedback system. Dose delivery was automatically triggered when the tracking point was within a predefined gating window. Intra-breath-hold stability and inter-fraction reproducibility across all fractions of the entire treatment course were analyzed per patient. RESULTS: In the present series, 6013 breath-holds during beam-on time were analyzed. The mean amplitude of the gating window from the baseline breathing curve (maximum expiration during free breathing) was 15.8 mm (95%-confidence interval: [8.5-30.6] mm) and had a width of 3.5 mm (95%-CI: [2-4.3] mm). As a measure of intra-breath-hold stability, the median standard deviation of the breath-hold level during DIBH was 0.3 mm (95%-CI: [0.1-0.9] mm). Similarly, the median absolute intra-breath-hold linear amplitude deviation was 0.4 mm (95%-CI: [0.01-2.1] mm). Reproducibility testing showed good inter-fractional reliability, as the maximum difference in the breathing amplitudes in all patients and all fractions were 1.3 mm on average (95%-CI: [0.5-2.6] mm). CONCLUSION: The clinical integration of an optical surface scanner enables a stable and reliable DIBH treatment delivery during SGRT for left-sided breast cancer in clinical routine.
引用
- Reitz D, Walter F, Schönecker S, et al. Stability and reproducibility of 6013 deep inspiration breath-holds in left-sided breast cancer. Radiat Oncol. 2020. PMID: 32448224. doi:10.1186/s13014-020-01572-w
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。