电磁应答器定位与实时追踪在前列腺癌放疗中金属髋关节假体患者的临床影响
该回顾性研究评估了8例携带金属髋关节假体(MHP)的前列腺癌患者接受根治性放疗时,Calypso电磁应答器(EMT)定位与追踪的可行性。结果显示Calypso系统未检测到MHP干扰,CBCT与Calypso引导一致性良好,EMT间距差异在0.3 cm内,但治疗中EMT中心点24次超出阈值(占20%分次),最大运动在前后方向达0.6 cm。
要点
- Calypso EMT在单侧或双侧MHP患者中可无干扰地定位和追踪前列腺位置。
- CBCT与Calypso引导一致性 optimal,EMT间距差异≤0.3 cm,几何残差在0.2 cm限值内。
- 20%分次出现EMT中心点超阈值,前后方向运动最大达0.6 cm。
核心结论
- Calypso EMT系统可在携带单侧或双侧金属髋关节假体的前列腺癌患者根治性放疗中实现无干扰的定位与实时追踪。
研究设计与样本
- 回顾性评估2016至2018年接受根治性放疗及EMT追踪的8例前列腺癌伴MHP患者(3例双侧、5例单侧),分析计划CT、首次分次CBCT及全部可评估分次(n=120)的EMT位置与治疗中断。
主要发现
- Calypso系统未检测到MHP干扰,每个治疗分次CBCT与Calypso引导一致性 optimal,EMT间距差异在0.3 cm内。
- Calypso测量的EMT至等中心距离与CT数据一致,并经CBCT确认。
- 放疗中EMT中心点24次超出阈值(占所有分次的20%):左右方向5次、前后方向15次、上下方向4次。
- 最大运动在前后轴:单侧MHP患者向前0.6 cm,双侧MHP患者向后0.5 cm。
关键数值与推荐
- 几何残差预期限值≤0.2 cm。
- CT-Calypso或CBCT-Calypso测量的EMT间距差异在0.3 cm内视为一致。
- EMT中心点超出阈值共24次(20%分次),前后方向15次、左右方向5次、上下方向4次。
临床怎么用
- 对于携带金属髋关节假体的前列腺癌患者,可考虑使用Calypso电磁应答器进行定位与实时追踪,无需担心假体干扰。
- 治疗中应关注前后方向的较大位移(可达0.6 cm),必要时结合CBCT验证。
局限性
- 回顾性研究,样本量仅8例,为单中心首次临床经验。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Our purpose was to assess the ability of electromagnetic transponders (EMTs) to localize and track movements in patients with prostate cancer (PCa) with metallic hip prostheses (MHPs) treated with curative radiation therapy (RT). METHODS AND MATERIALS: Data sets of 8 PCa patients with MHPs (3 bilateral and 5 unilateral) treated between 2016 and 2018 with RT and EMT tracking were retrospectively assessed. The distances between the 3 EMTs (apex to left, left to right, right to apex) and the isocenter were calculated both on planning computed tomography (CT) and cone beam CT (CBCT) at the first treatment fraction and compared with data reported by Calypso. EMT position and treatment interruptions triggered by Calypso were analyzed for all evaluable treatment fractions (n = 120). Localization accuracy was quantified by recording the geometric residual value (expected limit ≤0.2 cm) at the RT setup. RESULTS: The Calypso system was able to localize and track prostate position without any detectable interference from MHP. For every treatment fraction, the agreement between the CBCT images and Calypso guidance was optimal, with EMTs always within the defined tolerance (ie, CT-Calypso or CBCT-Calypso measured differences in inter-EMT distances within 0.3 cm). EMT to isocenter distances measured by Calypso reproduced CT data and were confirmed on CBCT scans. During RT, the EMT centroid exceeded the threshold 24 times (20% of all fractions): 5 times in the left-right, 15 times in the anteroposterior, and 4 times in the superoinferior directions. The largest motions recorded were in the anteroposterior axis: 0.6 cm anteriorly and 0.5 cm posteriorly in patients with unilateral and bilateral MHP, respectively. CONCLUSIONS: Our study represents the first clinical experience assessing the localization and tracking accuracy of Calypso EMTs during curative RT of patients with PCa with unilateral or bilateral MHP.