文献速递泌尿生殖与妇科肿瘤放疗放疗物理与技术

宫颈癌外照射与近距离治疗间乙状结肠复杂形变建模

Modeling Complex Deformations of the Sigmoid Colon Between External Beam Radiation Therapy and Brachytherapy Images of Cervical Cancer.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2020-02-03

约束对称薄板样条鲁棒点匹配(sTPS-RPM)平均DSC达0.96且逆一致性误差<1 mm,优于其他方法。乙状结肠在两次治疗间存在大而复杂的形变,D1cm3和D2cm3热点位置可能不一致。

要点

  • 约束sTPS-RPM配准乙状结肠平均DSC为0.96,逆一致性误差<1 mm。
  • BT模器插入导致乙状结肠大形变,刚性配准平均DSC仅0.26。
  • BT计划乙状结肠D1cm3和D2cm3热点形变至EBRT解剖平均位移27.1 mm。

核心结论

  • EBRT与BT间乙状结肠存在大而复杂的形变,D1cm3和D2cm3热点在整个治疗中不太可能处于同一位置。
  • 约束sTPS-RPM似乎是处理BT模器插入所致大形变的首选方法。

研究设计与样本

  • 比较五种配准方法将乙状结肠轮廓从BT传播至EBRT解剖:刚性配准、商业混合方法ANAtomically CONstrained Deformation Algorithm、RayStation控制ROI表面投影、经典和约束sTPS-RPM。
  • 用DSC、距离一致性和Hausdorff距离比较配准性能;对55例患者报告形变方法传播BT计划D1cm3和D2cm3至EBRT解剖的位移和差异。

主要发现

  • 每例患者均观察到乙状结肠大而复杂的形变。
  • 刚性配准乙状结肠对齐差,平均DSC为0.26。
  • ANAtomically CONstrained Deformation Algorithm平均DSC为0.57;仅用乙状结肠表面作为控制ROI平均DSC提高至0.79。
  • 经典和约束sTPS-RPM平均DSC分别为0.95和0.96,平均逆一致性误差<1 mm;约束sTPS-RPM提供更真实的形变和更好的表面拓扑。

关键数值与推荐

  • BT解剖上乙状结肠计划平均(范围)D1cm3为13.4 Gy(1-24.1),D2cm3为12.2 Gy(1-21.5)。
  • 使用约束sTPS-RPM将乙状结肠从BT形变至EBRT解剖,这些热点平均(范围)位移为27.1 mm(6.8-81)。

临床怎么用

  • 约束sTPS-RPM可用于将EBRT剂量映射至BT解剖,以支持个性化BT计划优化。
  • 应关注乙状结肠在EBRT与BT间的形变,避免假设D1cm3和D2cm3热点位置固定。

局限性

  • 本研究为回顾性分析,样本量有限。
  • 原文未报告其他配准方法传播D1cm3和D2cm3的位移数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: In this study, we investigated registration methods for estimating the large interfractional sigmoid deformations that occur between external beam radiation therapy (EBRT) and brachytherapy (BT) for cervical cancer. METHODS AND MATERIALS: Sixty-three patients were retrospectively analyzed. The sigmoid colon was delineated on 2 computed tomography images acquired during EBRT (without applicator) and BT (with applicator) for each patient. Five registration approaches were compared to propagate the contour of the sigmoid from BT to EBRT anatomies: rigid registration, commercial hybrid (ANAtomically CONstrained Deformation Algorithm), controlling ROI surface projection of RayStation, and the classical and constrained symmetrical thin-plate spline robust point matching (sTPS-RPM) methods. Deformation of the sigmoid due to insertion of the BT applicator was reported. Registration performance was compared by using the Dice similarity coefficient (DSC), distance to agreement, and Hausdorff distance. The 2 sTPS-RPM methods were compared by using surface triangle quality criteria between deformed surfaces. Using the deformable approaches, the BT dose of the sigmoid was deformed toward the EBRT anatomy. The displacement and discrepancy between the deformable methods to propagate the planned D1cm3 and D2cm3 of the sigmoid from BT to EBRT anatomies were reported for 55 patients. RESULTS: Large and complex deformations of the sigmoid were observed for each patient. Rigid registration resulted in poor sigmoid alignment with a mean DSC of 0.26. Using the contour to drive the deformation, ANAtomically CONstrained Deformation Algorithm was able to slightly improve the alignment of the sigmoid with a mean DSC of 0.57. Using only the sigmoid surface as controlling ROI, the mean DSC was improved to 0.79. The classical and constrained sTPS-RPM methods provided mean DSCs of 0.95 and 0.96, respectively, with an average inverse consistency error <1 mm. The constrained sTPS-RPM provided more realistic deformations and better surface topology of the deformed sigmoids. The planned mean (range) D1cm3 and D2cm3 of the sigmoid were 13.4 Gy (1-24.1) and 12.2 Gy (1-21.5) on the BT anatomy, respectively. Using the constrained sTPS-RPM to deform the sigmoid from BT to EBRT anatomies, these hotspots had a mean (range) displacement of 27.1 mm (6.8-81). CONCLUSIONS: Large deformations of the sigmoid were observed between the EBRT and BT anatomies, suggesting that the D1cm3 and D2cm3 of the sigmoid would unlikely to be at the same position throughout treatment. The proposed constrained sTPS-RPM seems to be the preferred approach to manage the large deformation due to BT applicator insertion. Such an approach could be used to map the EBRT dose to the BT anatomy for personalized BT planning optimization.