高场MR引导前列腺放疗中自动形变结构传播的准确性
本研究在高场MR-linac治疗的高危前列腺癌患者中,评估从计划CT和MR扫描自动形变传播结构的准确性。结果显示MR-to-MR传播结构的中位表面距离和Dice相似指数优于CT-MR传播,且MR-MR传播不确定性与观察者内变异程度相似。基于CT和MR的传播均需手动编辑,但MR-to-MR传播结构所需校正更少。
要点
- MR-to-MR传播结构的中位表面距离显著小于CT-MR传播,Dice相似指数更大。
- MR-MR传播不确定性与计划MR上观察者内变异程度相似。
- CT和MR传播均需手动编辑,但MR-to-MR所需校正更少。
核心结论
- 在高场MR-linac治疗的高危前列腺癌患者中,MR-to-MR自动形变传播结构比CT-MR传播更准确,所需手动校正更少。
研究设计与样本
- 对12例接受前列腺及盆腔淋巴结治疗的高危前列腺癌患者,在计划MR和CT上勾画靶区与危及器官,并通过形变配准传播至治疗期间获取的三次T2加权MR扫描。
- 以计划MR上获得的观察者内变异作为地面 truth,在重复扫描上对照手动勾画评估生成结构。
主要发现
- MR-to-MR传播结构相比CT-MR传播具有显著更小的中位表面距离和更大的Dice相似指数。
- MR-MR传播不确定性与观察者内变异在幅度上相似。
- 视觉检查显示形变结构能较好处理源图像与目标图像间器官的小解剖差异,但无法处理大差异。
临床怎么用
- 在高场MR-linac在线自适应放疗中,可优先考虑MR-to-MR形变传播以减少重新计划时间和靶区错误表示风险,但仍需手动编辑。
局限性
- 研究样本量较小(12例),且仅针对高危前列腺癌患者,结论外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: In this study we have evaluated the accuracy of automatic, deformable structure propagation from planning CT and MR scans for daily online plan adaptation for MR linac (MRL) treatment, which is an important element to minimize re-planning time and reduce the risk of misrepresenting the target due to this time pressure. METHODS: For 12 high-risk prostate cancer patients treated to the prostate and pelvic lymph nodes, target structures and organs at risk were delineated on both planning MR and CT scans and propagated using deformable registration to three T2 weighted MR scans acquired during the treatment course. Generated structures were evaluated against manual delineations on the repeated scans using intra-observer variation obtained on the planning MR as ground truth. RESULTS: MR-to-MR propagated structures had significant less median surface distance and larger Dice similarity index compared to CT-MR propagation. The MR-MR propagation uncertainty was similar in magnitude to the intra-observer variation. Visual inspection of the deformed structures revealed that small anatomical differences between organs in source and destination image sets were generally well accounted for while large differences were not. CONCLUSION: Both CT and MR based propagations require manual editing, but the current results show that MR-to-MR propagated structures require fewer corrections for high risk prostate cancer patients treated at a high-field MRL.
引用
- Christiansen RL, Dysager L, Bertelsen AS, et al. Accuracy of automatic deformable structure propagation for high-field MRI guided prostate radiotherapy. Radiat Oncol. 2020. PMID: 32033574. doi:10.1186/s13014-020-1482-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。