I期非小细胞肺癌被动散射碳离子放疗的每日CT剂量评估

Dose assessment for patients with stage I non-small cell lung cancer receiving passive scattering carbon-ion radiotherapy using daily computed tomographic images: A prospective study.
PubMed 文献 · Radiother Oncol · 2020-02-07

一项前瞻性研究纳入10例I期非小细胞肺癌患者,采集40组分次前CT图像,比较骨匹配与肿瘤匹配下的剂量分布。结果显示骨匹配下25%分次剂量分布不可接受,肿瘤匹配下仅2.5%不可接受(P<0.001)。结论推荐每日CT对齐用于碳离子放疗。

要点

  • 骨匹配下25%分次剂量分布不可接受,肿瘤匹配仅2.5%
  • 肿瘤匹配下所有患者累积剂量分布均满意
  • 推荐I期NSCLC碳离子放疗采用每日CT对齐

核心结论

  • 基于CT图像的肿瘤匹配在分次和累积剂量分布上均优于传统骨匹配,推荐I期NSCLC碳离子放疗行每日CT对齐。

研究设计与样本

  • 前瞻性研究纳入10例I期非小细胞肺癌患者,采集40组分次前CT图像,与计划CT配准后行骨匹配和肿瘤匹配剂量评估,并使用可变形图像配准生成累积剂量。

主要发现

  • 骨匹配下40组分次剂量分布中25%不可接受,肿瘤匹配下仅2.5%不可接受(P<0.001)。
  • 骨匹配下3例患者累积剂量分布不可接受,肿瘤匹配下全部满意(P<0.001)。
  • 水等效路径长度未见差异。
  • 剂量分布较差患者的所需边界在骨匹配和肿瘤匹配下分别为5.9 mm和4.4 mm。

临床怎么用

  • 对接受碳离子放疗的I期NSCLC患者,建议采用每日CT图像引导的肿瘤匹配进行对齐。
  • 肿瘤匹配较骨匹配可减少分次和累积剂量分布不可接受的比例。

局限性

  • 样本量较小,仅纳入10例患者。
  • 水等效路径长度变化的影响似乎有限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: This study aimed to assess dose distributions for stage I non-small cell lung cancer (NSCLC) with passive scattering carbon-ion radiotherapy (C-ion RT) using daily computed tomography (CT) images. MATERIALS AND METHODS: We enrolled 10 patients with stage I NSCLC and acquired a total of 40 pre-fractional CT image series under the same settings as the planning CT images. These CT images were registered with planning CT images for dose evaluation using both bone matching (BM) and tumor matching (TM). Using deformable image registration, we generated accumulated doses. Moreover, the volumetric dose parameters were compared in terms of tumor coverage and lung exposure and statistical analyses were performed. RESULTS: Overall, 25% of 40 fractional dose distributions were unacceptable with BM, compared with 2.5% with TM (P < 0.001). Using BM, three patients' accumulated dose distributions were unacceptable; however, all were satisfactory with TM (P < 0.001). No differences were observed in water-equivalent path length (WEL). The required margins in patients with poor dose distribution were 5.9 and 4.4 mm for BM and TM, respectively. CONCLUSIONS: This study establishes that CT image-based TM is robust compared with conventional BM for both daily and accumulated dose distributions. The effects of changes in WEL seem to be limited. Hence, daily CT alignment is recommended for patients with stage I NSCLC receiving C-ion RT.