三维表面成像系统用于深吸气屏气乳腺癌患者定位与分次内监测的评估

Evaluation of a 3D surface imaging system for deep inspiration breath-hold patient positioning and intra-fraction monitoring.
PubMed 文献 · Radiat Oncol · 2019-07-11

结果显示SGRT与CBCT数据相关性良好,表面与CBCT匹配的相关性(0.61)高于靶区与CBCT匹配(0.44),系统与随机摆位误差差异各方向≤2 mm,SGRT可检测气量BH系统无法发现的意外患者移动。

要点

  • SGRT与CBCT相关性良好,表面匹配相关性强于靶区匹配。
  • 各方向系统与随机摆位误差差异均≤2 mm。
  • SGRT可检测气量BH系统无法发现的意外移动。

核心结论

  • SGRT可用于左侧乳腺癌BH患者的定位与连续监测,使位置误差保持在5 mm以内。

研究设计与样本

  • CBCT扫描分别与靶区和患者表面进行匹配,并采用线性回归分析评估摆位误差差异。
  • 分次内变异性和稳定性通过SGRT测量确定,变异性取单次治疗内不同BH水平间最大差值,稳定性取每次BH起始与结束位置差值。

主要发现

  • SGRT数据与CBCT数据相关性良好,表面与CBCT匹配的相关系数为0.61,靶区与CBCT匹配为0.44。
  • 各方向系统与随机摆位误差差异均≤2 mm。
  • 表面与CBCT匹配的95%一致性界限(均值±2SD)在三个正交方向分别为0.1±3.0、0.6±4.1和0.4±3.4 mm。
  • 气量引导BH的SGRT检测变异性在三个正交方向分别为2.2、2.8和2.3 mm,稳定性分别为-1.0、2.1和1.5 mm。
  • SGRT系统可检测气量BH系统无法检测的意外患者移动。

临床怎么用

  • 可考虑使用SGRT对左侧乳腺癌BH患者进行定位和分次内连续监测,以维持位置误差在5 mm以内。
  • 气量引导BH系统可实现低分次内变异性和良好稳定性,但SGRT可提供其无法检测的额外患者位置信息。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To determine the accuracy of a surface guided radiotherapy (SGRT) system for positioning of breast cancer patients in breath-hold (BH) with respect to cone-beam computed tomography (CBCT). Secondly, to evaluate the thorax position stability during BHs with SGRT, when using an air-volume guidance system. METHODS AND MATERIALS: Eighteen left-sided breast cancer patients were monitored with SGRT during CBCT and treatment, both in BH. CBCT scans were matched on the target volume and the patient surface. The setup error differences were evaluated, including with linear regression analysis. The intra-fraction variability and stability of the air-volume guided BHs were determined from SGRT measurements. The variability was determined from the maximum difference between the different BH levels within one treatment fraction. The stability was determined from the difference between the start and end position of each BH. RESULTS: SGRT data correlated well with CBCT data. The correlation was stronger for surface-to-CBCT (0.61) than target volume-to-CBCT (0.44) matches. Systematic and random setup error differences were ≤ 2 mm in all directions. The 95% limits of agreement (mean ± 2SD) were 0.1 ± 3.0, 0.6 ± 4.1 and 0.4 ± 3.4 mm in the three orthogonal directions, for the surface-to-CBCT matches. For air-volume guided BHs, the variability detected with SGRT was 2.2, 2.8 and 2.3 mm, and the stability - 1.0, 2.1 and 1.5 mm, in three orthogonal directions. Furthermore, the SGRT system could detect unexpected patient movement, undetectable by the air-volume BH system. CONCLUSION: With SGRT, left-sided breast cancer patients can be positioned and monitored continuously to maintain position errors within 5 mm. Low intra-fraction variability and good stability can be achieved with the air-volume BH system, however, additional patient position information is available with SGRT, that cannot be detected with air-volume BH systems.