笔束扫描质子治疗中屏气残余运动的剂量学效应:一项实验研究
研究将3例患者CT提取的3D打印肿瘤置入动态拟人呼吸模体,按既往透视评估的屏气期间肿瘤运动设置运动,用辐射变色胶片测量靶区剂量,比较屏气与静止状态下单野均匀剂量(SFUD)和调强质子治疗(IMPT)计划的剂量偏差。SFUD计划γ通过率(3%、3 mm)为85%或更高,IMPT计划降至50%–70%;作者结论认为基于该实验研究SFUD对屏气残余运动稳健,IMPT稳健性较差。
要点
- SFUD计划在静止与屏气病例间剂量偏差小,γ通过率(3%、3 mm)为85%或更高。
- IMPT计划剂量偏差更明显,γ通过率降至50%–70%。
- 作者基于实验研究认为屏气技术对SFUD的屏气内残余运动稳健,IMPT稳健性较差。
核心结论
- 基于该实验研究,屏气技术对SFUD治疗计划的屏气内残余运动稳健。
- IMPT稳健性较差,检测到更大的剂量偏差。
研究设计与样本
- 实验研究:将3例患者CT扫描提取的3D打印肿瘤置入动态拟人呼吸模体。
- 靶区按既往透视评估的每例患者屏气期间肿瘤运动设置运动,用辐射变色胶片测量靶区剂量。
- 分别实施SFUD和IMPT计划,并对每例患者重复无任何运动的实验,以计算静止与屏气病例间的相对剂量偏差。
主要发现
- SFUD计划在静止与屏气病例间显示小剂量偏差,γ通过率(3%、3 mm)为85%或更高。
- IMPT计划剂量偏差更明显,γ通过率降至50%–70%。
关键数值与推荐
- 剂量偏差以γ通过率(3%、3 mm)评价:SFUD为85%或更高,IMPT降至50%–70%。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND PURPOSE: Motion management in the treatment of lung cancer is necessary to assure highest quality of the delivered radiation therapy. In this study, the breath-hold technique is experimentally investigated for pencil beam scanned (PBS) proton therapy, with respect to the dosimetric effect of residual breath-hold motion. MATERIAL AND METHODS: Three-dimensional (3D)-printed tumours extracted from CT scans of three patients were inserted into a dynamic anthropomorphic breathing phantom. The target was set up to move with the individual patient's tumour motion during breath-hold as previously assessed on fluoroscopy. Target dose was measured with radio-chromic film, and both single field uniform dose (SFUD) and intensity-modulated proton therapy (IMPT) plans were delivered. Experiments were repeated for each patient without any motion, to compute the relative dose deviation between static and breath-hold cases. RESULTS: SFUD plans showed small dose deviations between static and breath-hold cases, as evidenced by the gamma pass rate (3%, 3 mm) of 85% or higher. Dose deviation was more evident for IMPT plans, with gamma pass rate reduced to 50-70%. CONCLUSIONS: The breath-hold technique is robust to residual intra-breath-hold motion for SFUD treatment plans, based on our experimental study. IMPT was less robust with larger detected dose deviations.
引用
- Gorgisyan J, Lomax AJ, Rosenschold PMA, et al. The dosimetric effect of residual breath-hold motion in pencil beam scanned proton therapy - An experimental study. Radiother Oncol. 2019. PMID: 31005207. doi:10.1016/j.radonc.2019.01.033
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。