加速直线加速器机架旋转的超快单次屏气锥形束CT肺癌成像

Ultrafast single breath-hold cone-beam CT lung cancer imaging with faster linac gantry rotation.
PubMed 文献 · Radiother Oncol · 2019-03-16

该概念验证研究将机架旋转速度提升至18°/s,使单次屏气锥形束CT(CBCT)采集时间限制在10-20秒,并与标准CBCT(3°/s)比较图像质量、剂量和配准精度。超快CBCT在高对比度区域图像质量可接受,剂量输出降低60-83%,配准精度保持优于1毫米,与标准CBCT无显著配准差异。作者认为该技术有望在单次屏气内实现分次内成像与治疗结合。

要点

  • 机架加速至18°/s使单次屏气CBCT采集时间限于10-20秒。
  • 超快CBCT在高对比度区域图像质量可接受,对比度因欠采样降低约80%。
  • 剂量输出降低60-83%,配准精度保持优于1毫米,与标准CBCT无显著差异。

核心结论

  • 加速机架旋转至18°/s的超快高对比度肺部肿瘤CBCT成像在单次屏气10-20秒内显示出有希望的结果。
  • 如此快速的成像时间结合快速治疗递送,可能为单次屏气相位内的分次内联合成像与治疗铺平道路,从而减少残余运动并提高治疗精度和患者舒适度。
  • 更快的机架旋转可在时间至关重要的场景(姑息患者护理、总体减少不确定性、提高患者吞吐量)设定具有巨大临床影响的基准。

研究设计与样本

  • 概念验证研究,遵循客户验收测试分析图像质量,测量CT剂量指数,并用偏心滚珠轴承体模确定配准精度。
  • 使用胸部体模中不同形状的肿瘤模拟插件模拟一名肺部肿瘤患者。
  • 通过信噪比、对比噪声比和插件几何形状量化图像质量,在代表性位置测量剂量。
  • 在预定位置扫描插件确定配准精度,在计划CT上执行手动、自动(临床软件)和客观自动(内部开发)配准,比较结果与应用位移之间的偏移。

主要发现

  • 超快CBCT在高对比度区域图像质量可接受,尽管因欠采样导致对比度降低约80%。
  • 在所示设置中剂量输出显著降低60-83%,变化归因于机架制动特性。
  • 配准精度保持优于1毫米,客观自动配准的平均位移误差为0.0±0.2毫米。
  • 超快CBCT与标准CBCT之间未显示显著的配准差异。

局限性

  • 本研究为加速机架旋转至18°/s的首次测试。
  • 图像质量分析基于客户验收测试和体模,而非临床患者图像。
  • 对比度因欠采样降低约80%。
  • 剂量输出降低60-83%的变化归因于机架制动特性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Lung tumors treated with hypo-fractionated deep-inspiration breath-hold stereotactic body radiotherapy benefit from fast imaging and treatment. Single breath-hold cone-beam-CT (CBCT) could reduce motion artifacts and improve treatment precision. Thus, gantry speed was accelerated to 18°/s, limiting acquisition time to 10-20 s. Image quality, dosimetry and registration accuracy were compared with standard-CBCT (3°/s). METHODS AND MATERIALS: For proof-of-concept, image quality was analyzed following customer acceptance tests, CT-dose index measured, and registration accuracy determined with an off-centered ball-bearing-phantom. A lung-tumor patient was simulated with differently shaped tumor-mimicking inlays in a thorax-phantom. Signal-to-noise-ratio, contrast-to-noise-ratio and geometry of the inlays quantified image quality. Dose was measured in representative positions. Registration accuracy was determined with inlays scanned in pre-defined positions. Manual, automatic (clinical software) and objective-automatic (in-house-developed) registration was performed on planning-CT, offsets between results and applied shifts were compared. RESULTS: Image quality of ultrafast-CBCT was adequate for high-contrast areas, despite contrast-reduction of ∼80% due to undersampling. Dose-output was considerably reduced by 60-83% in presented setup; variations are due to gantry-braking characteristics. Registration accuracy was maintained better than 1 mm, mean displacement errors were 0.0 ± 0.2 mm with objective-automatic registration. Ultrafast-CBCT showed no significant registration differences to standard-CBCT. CONCLUSIONS: This study of first tests with faster gantry rotation of 18°/s showed promising results for ultrafast high-contrast lung tumor CBCT imaging within single breath-hold of 10-20 s. Such fast imaging times, in combination with fast treatment delivery, could pave the way for intra-fractional combined imaging and treatment within one breath-hold phase, and thus mitigate residual motion and increase treatment accuracy and patient comfort. Even generally speaking, faster gantry rotation could set a benchmark with immense clinical impact where time matters most: palliative patient care, general reduction in uncertainty, and increase in patient throughput especially important for emerging markets with high patient numbers.