肺癌患者缩短扫描时间的四维锥形束CT定量评估
本研究在15例肺癌患者的44次四维锥形束CT中模拟缩短扫描时间至1分钟,并对12例患者采集20次2分钟扫描,评估图像质量与配准精度。2分钟扫描对配准精度和肿瘤运动检测影响较小,但扫描时间低于2分钟时图像质量明显下降。
要点
- 2分钟四维锥形束CT对多数肺癌患者图像引导放疗的图像质量足够。
- 2分钟扫描对自动配准精度和肿瘤运动检测能力影响较小。
- 扫描时间低于2分钟时自动配准仍多在小误差内,但视觉图像质量明显下降。
核心结论
- 2分钟的四维锥形束CT采集时间在多数肺癌患者中可产生足够图像质量用于图像引导放疗,这一点通过评估模拟和真实采集对自动配准精度的影响得到了定量证明。
研究设计与样本
- 通过仅保留每第二、第三或第四呼吸周期对应的投影图像,在15例肺癌患者的44次四维锥形束CT中模拟缩短至1分钟的扫描时间。
- 另对12例肺癌患者采集了20次2分钟扫描。
主要发现
- 在可获取4分钟扫描配准结果作为参考时,2分钟扫描对配准精度或检测肿瘤运动的能力影响较小。
- 在2分钟采集分析的8次扫描中,6次自动配准精度在2毫米以内;模拟扫描中96.6%的配准差异在2毫米以内。
- 模拟扫描时间低于2分钟时,84.7%的扫描自动配准结果仍在2毫米以内,但视觉图像质量明显下降。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Image guided radiotherapy (IGRT) based on respiration correlated cone-beam CT (4D-CBCT) provides accurate tumour localisation in lung cancer patients by taking into account respiratory motion when deriving setup correction. However, 4D-CBCT scan times are typically longer than for acquisition of 3D-CBCT scans, e.g. 4 min. This work aims to quantitatively evaluate the effect of reduced scan times on 4D-CBCT image quality and registration accuracy in lung cancer patients. METHODS AND MATERIALS: Scan times down to 1 min were simulated by retaining only projection images corresponding to every second, third or fourth respiratory cycle in forty-four 4D-CBCTs from 15 lung cancer patients. In addition twenty 2-minute scans were acquired for 12 lung cancer patients. Image quality was quantified by assessing registration accuracy in the shorter scan times, comparing to the 4-minute scan registration result where available as reference. RESULTS: Use of 2-minute scans had little impact on registration accuracy or ability to detect tumour motion: automatic registration accuracy was within 2 mm in 6/8 scans analysed with 2-minute acquisitions, and 96.6% of registration discrepancies were within 2 mm for the simulated scans. When the scan time simulated was below 2 min, automatic registration results still agreed within 2 mm for 84.7% of scans, however visual image quality was considerably degraded. CONCLUSION: A 4D-CBCT acquisition time of 2 min produces scans of sufficient image quality for IGRT in most lung cancer patients, as demonstrated quantitatively by assessing the impact on automatic registration accuracy in simulated and real acquisitions.
引用
- Bryce-Atkinson A, Marchant T, Rodgers J, et al. Quantitative evaluation of 4D Cone beam CT scans with reduced scan time in lung cancer patients. Radiother Oncol. 2019. PMID: 31015131. doi:10.1016/j.radonc.2019.03.027
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。