文献速递胸部肿瘤放疗放疗物理与技术

四维计算机断层扫描与四维锥形束计算机断层扫描在放疗中均低估肺部靶区运动

Both four-dimensional computed tomography and four-dimensional cone beam computed tomography under-predict lung target motion during radiotherapy.
PubMed 文献 · Radiother Oncol · 2019-03-14

研究纳入10例接受肺部立体定向消融放疗(SABR)并植入Calypso信标的患者,比较4DCT与4DCBCT测量的靶区运动范围与实时成像及治疗时的靶区运动范围。结果显示4DCT与4DCBCT均显著低估治疗靶区运动范围,且4DCT与4DCBCT在头脚方向上的预测差异有统计学意义(p=0.03)。

要点

  • 4DCT与4DCBCT均显著低估肺部SABR治疗时的靶区运动范围(p<0.0001)。
  • 头脚方向治疗靶区运动范围被4DCT低估1.7倍、被4DCBCT低估1.5倍。
  • 4DCT与4DCBCT在头脚方向预测治疗运动范围差异有统计学意义(p=0.03)。

核心结论

  • 对于肺部SABR患者,4DCT与4DCBCT均显著低估治疗靶区运动范围。

研究设计与样本

  • 10例肺部SABR患者植入Calypso信标,对29个分次(每例1-4分次)进行1分钟CBCT扫描并重建4DCBCT。
  • 比较所有4DCT与4DCBCT分箱的信标质心运动与每个分次的实时成像及治疗信标质心(靶区)运动范围(4倍标准差)。

主要发现

  • 头脚方向4DCT、4DCBCT、成像及治疗运动范围分别为6.3±3.6 mm、7.1±4.5 mm、11.1±7.5 mm和10.9±6.9 mm。
  • 左右及前后方向观察到类似的4DCT与4DCBCT低估。
  • 治疗靶区运动范围在头脚、左右、前后方向被4DCT低估的倍数分别为1.7、1.9、1.7,被4DCBCT低估的倍数分别为1.5、1.6、1.6。
  • 呼气末的均方根误差(RMSE)通常低于吸气末。
  • 4DCBCT与成像及治疗靶区运动的相关性高于4DCT,头脚方向两者差异有统计学意义(p=0.03)。

关键数值与推荐

  • 假设(1)被拒绝(p<0.0001)。
  • 假设(2)在头脚方向显示4DCT与4DCBCT差异有统计学意义(p=0.03)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: To test the hypothesis that 4DCT and 4DCBCT-measured target motion ranges predict target motion ranges during lung cancer SABR. MATERIALS AND METHODS: Ten lung SABR patients were implanted with Calypso beacons. 4DCBCT was reconstructed for 29 fractions (1-4fx/patient) from a 1 min CBCT scan. The beacon centroid motion segmented for all 4DCT and 4DCBCT bins was compared with the real-time imaging and treatment beacon centroid ("target") motion range (4SDs) for each fraction. We tested the hypotheses that (1) 4DCT and 4CBCT predict treatment motion range and (2) there is no difference between 4DCT and 4DCBCT for predicting treatment motion range. Phase-wise root-mean-square errors (RMSEs) between imaging and treatment motion and reconstructed motion (4DCT, 4DCBCT) were calculated. Relationships between motion ranges in 4DCT and 4DCBCT and imaging and treatment motion ranges were investigated for the superior-inferior (SI), left-right (LR) and anterior-posterior (AP) directions. Baseline drifts and amplitude variability were investigated as potential factors leading to motion misrepresentation. RESULTS: SI 4DCT, 4DCBCT, imaging and treatment motion ranges were 6.3 ± 3.6 mm, 7.1 ± 4.5 mm, 11.1 ± 7.5 mm and 10.9 ± 6.9 mm, respectively. Similar 4DCT and 4DCBCT under-predictions were observed in the LR and AP directions. Hypothesis (1) was rejected (p < 0.0001). Treatment target motion range was under-predicted in 4DCT by factors of 1.7, 1.9 and 1.7 and in 4DCBCT by factors of 1.5, 1.6 and 1.6 in the SI, LR, and AP directions, respectively. RMSEs were generally lower for end-exhale than inhale. 4DCBCT showed higher correlations with the imaging and treatment target motion than 4DCT and testing hypothesis (2) a statistically significant difference between 4DCT and 4DCBCT was shown in the SI direction (p = 0.03). CONCLUSION: For lung SABR patients both 4DCT and 4DCBCT significantly under-predict treatment target motion ranges.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Steiner E, Shieh CC, Caillet V, et al. Both four-dimensional computed tomography and four-dimensional cone beam computed tomography under-predict lung target motion during radiotherapy. Radiother Oncol. 2019. PMID: 31015172. doi:10.1016/j.radonc.2019.02.019

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。