一种快速生成合成计算机断层扫描用于磁共振成像引导在线自适应再计划的技术:一项探索性研究
本研究开发了一种基于图谱的自动合成计算机断层扫描(sCT)生成技术,用于磁共振成像(MRI)引导放疗的在线自适应再计划。该方法利用每日MRI与参考CT的可变形图像配准,并校正骨刚性和传递随机空气区域。在头颈、胸部、腹部和盆腔4个颅外部位,sCT的平均绝对误差和剂量学分析结果均优于基于体素的CT(bCT)。
要点
- sCT生成方法在4个颅外部位的平均绝对误差均低于bCT。
- gamma分析(1.5%,2 mm)显示sCT通过率高于bCT。
- 该方法可全自动化,适用于在线自适应再计划。
核心结论
- 所提出的方法生成的sCT与基于体密度赋值的方法相比精度相等或更高,且无需多个MRI序列。
- 该方法可完全自动化并适用于在线自适应再计划。
研究设计与样本
- 探索性研究,开发基于图谱的自动sCT生成技术,使用每日MRI与参考CT的可变形图像配准,并在4个颅外部位(头颈、胸部、腹部、盆腔)进行比较。
主要发现
- sCT与bCT的平均绝对误差(MAE)在头颈部为25.5 vs 66.7,胸部为25.9 vs 65.3,腹部为24.8 vs 44.2,盆腔为16.6 vs 47.7。
- gamma分析(1.5%,2 mm)通过率在头颈部为98.7 vs 97.1,胸部为99.1 vs 93.9,腹部为99.5 vs 99.4,盆腔为99.7 vs 99.4。
- MAE和剂量学分析结果均有利于所提出的sCT生成方法。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To develop an automatic, accurate, atlas-based technique for synthetic computed tomography (sCT) generation to be used for online adaptive replanning during magnetic resonance imaging (MRI)-guided radiation therapy (RT). METHODS AND MATERIALS: The proposed method uses deformable image registration (DIR) of daily MRI and reference computed tomography (CT) with additional corrections to maintain bone rigidity and to transfer random air regions by thresholding. The DIR is performed with constraints on the bony structures using a special algorithm of ADMIRE (Elekta). The air regions are delineated from low-signal regions on the daily MRI and forced to air density. The bone regions in the MRI (already determined from the CT) are separated from the air regions because both bone and air have low signal density in MRI. All these steps are automated. The generated sCT is compared with reference CT and the alternative voxel-based CT (bCT) for 4 extracranial sites (head and neck, thorax, abdomen, pelvis) in terms of mean absolute error (MAE), gamma analysis of 3-dimensional doses, and dose volume histogram parameters. RESULTS: Both MAE and dosimetric analysis results were favorable for the proposed sCT generation method. The average MAE for the sCT/bCT were 25.5/66.7, 25.9/65.3, 24.8/44.2 and 16.6/47.7 for head and neck, thorax, abdomen, and pelvis, respectively, and the gamma analysis (1.5%, 2 mm) yielded 98.7/97.1, 99.1/93.9, 99.5/99.4, 99.7/99.4, respectively, for those sites. CONCLUSIONS: The proposed method generates equal or more accurate sCT than those from the bulk density assignment, without the need for multiple MRI sequences. This method can be fully automated and applicable for online adaptive replanning.
引用
- Ahunbay EE, Thapa R, Chen X, et al. A Technique to Rapidly Generate Synthetic Computed Tomography for Magnetic Resonance Imaging-Guided Online Adaptive Replanning: An Exploratory Study. Int J Radiat Oncol Biol Phys. 2018. PMID: 30550817. doi:10.1016/j.ijrobp.2018.12.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。