基于MRI的放射治疗计划中伪CT生成:基于图像块、基于图谱和体密度方法的比较

Pseudo-CT Generation for MRI-Only Radiation Therapy Treatment Planning: Comparison Among Patch-Based, Atlas-Based, and Bulk Density Methods.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2018-10-16

研究纳入39例接受容积旋转调强放疗的前列腺癌患者,比较新提出的非局部均值图像块法(PBM)与基于图谱法(ABM)、体密度法(BDM)生成伪CT的成像与剂量学精度。PBM的平均绝对误差为41.1 HU,平均误差为-1.1 HU,剂量体积直方图差异小,前列腺计划靶区、直肠和膀胱的剂量不确定度显著低于ABM和BDM,平均γ通过率99.5%显著高于ABM的94.9%和BDM的96.1%。作者结论认为PBM不确定度低,小于ABM和BDM,且不太可能有临床意义。

要点

  • PBM的平均绝对误差为41.1 HU,平均误差为-1.1 HU。
  • PBM的γ通过率99.5%显著高于ABM的94.9%和BDM的96.1%。
  • PBM在前列腺计划靶区、直肠和膀胱的剂量不确定度显著低于ABM和BDM。

核心结论

  • 所提出的PBM在参考CT上计划的剂量中提供低不确定度,这些不确定度小于ABM和BDM,且不太可能有临床意义。

研究设计与样本

  • 39例接受容积旋转调强放疗的前列腺癌患者,除计划CT外采集T2加权MRI,用PBM、ABM和BDM三种方法从MRI生成伪CT,PBM在训练队列中通过特征提取和近似最近邻搜索进行,并在验证队列中评估成像和剂量学终点。

主要发现

  • PBM的平均绝对误差为41.1 HU,平均误差为-1.1 HU。
  • PBM的剂量体积直方图差异为前列腺计划靶区V95%为0.7%、直肠V70Gy为0.5%、膀胱V50Gy为0.2%。
  • 与ABM和BDM相比,PBM在前列腺计划靶区(70-78 Gy)、直肠(8.5-29 Gy、40-48 Gy和61-73 Gy)和膀胱(12-78 Gy)提供显著更低的剂量不确定度。
  • PBM的平均γ通过率(99.5%)显著高于ABM(94.9%)或BDM(96.1%)。

关键数值与推荐

  • PBM的平均绝对误差为41.1 HU,平均误差为-1.1 HU。
  • PBM的γ通过率99.5%显著高于ABM的94.9%和BDM的96.1%。
  • PBM在前列腺计划靶区(70-78 Gy)、直肠(8.5-29 Gy、40-48 Gy、61-73 Gy)和膀胱(12-78 Gy)的剂量不确定度显著低于ABM和BDM。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Methods have been recently developed to generate pseudo-computed tomography (pCT) for dose calculation in magnetic resonance imaging (MRI)-only radiation therapy. This study aimed to propose an original nonlocal mean patch-based method (PBM) and to compare this PBM to an atlas-based method (ABM) and to a bulk density method (BDM) for prostate MRI-only radiation therapy. MATERIALS AND METHODS: Thirty-nine patients received a volumetric modulated arc therapy for prostate cancer. In addition to the planning computed tomography (CT) scans, T2-weighted MRI scans were acquired. pCTs were generated from MRIs using 3 methods: an original nonlocal mean PBM, ABM, and BDM. The PBM was performed using feature extraction and approximate nearest neighbor search in a training cohort. The PBM accuracy was evaluated in a validation cohort by using imaging and dosimetric endpoints. Imaging endpoints included mean absolute error and mean error between Hounsfield units of the pCT and the reference CT (CTref). Dosimetric endpoints were based on dose-volume histograms calculated from the CTref and the pCTs for various volumes of interest and on 3-dimensional gamma analyses. The PBM uncertainties were compared with those of the ABM and BDM. RESULTS: The mean absolute error and mean error obtained from the PBM were 41.1 and -1.1 Hounsfield units. The PBM dose-volume histogram differences were 0.7% for prostate planning target volume V95%, 0.5% for rectum V70Gy, and 0.2% for bladder V50Gy. Compared with ABM and BDM, PBM provided significantly lower dose uncertainties for the prostate planning target volume (70-78 Gy), the rectum (8.5-29 Gy, 40-48 Gy, and 61-73 Gy), and the bladder (12-78 Gy). The PBM mean gamma pass rate (99.5%) was significantly higher than that of ABM (94.9%) or BDM (96.1%). CONCLUSIONS: The proposed PBM provides low uncertainties with dose planned on CTref. These uncertainties were smaller than those of ABM and BDM and are unlikely to be clinically significant.