文献速递泌尿生殖与妇科肿瘤放疗放疗物理与技术

基于MRI的定量磁化率成像与无监督机器学习自动检测植入的低剂量率近距离治疗粒子

MRI-based automated detection of implanted low dose rate (LDR) brachytherapy seeds using quantitative susceptibility mapping (QSM) and unsupervised machine learning (ML).
PubMed 文献 · Radiother Oncol · 2018-09-19

本研究构建含20枚碘-125链状假粒子和钙化的前列腺模体,在3T和1.5T磁共振上采用三维梯度回波序列和优化的定量磁化率成像技术,利用无监督机器学习算法(K-means和K-medoids聚类)确定粒子数量、质心和方向。基于定量磁化率成像的方法生成的粒子阳性对比与钙化显著不同,3T和1.5T估计的粒子质心与标准CT检测算法完全一致(最大差异0.7 mm),粒子方向与真实方向高度相关(R>0.98)。

要点

  • 优化的定量磁化率成像方法为粒子生成高质量阳性对比,与钙化对比显著不同,可通过阈值区分。
  • 3T和1.5T磁共振数据估计的粒子质心与标准CT检测算法完全一致,最大差异0.7 mm。
  • 估计的粒子方向与真实方向高度相关(R>0.98)。

核心结论

  • 所提出的基于MRI的工作流程能够准确稳健地定位粒子(位置和方向),在复杂粒子配置上验证后,有潜力替代目前广泛应用的基于CT的工作流程。

研究设计与样本

  • 构建含20枚碘-125链状假粒子和钙化的前列腺模体,在3T和1.5T磁共振扫描仪上采用三维梯度回波序列,应用优化的定量磁化率成像技术生成粒子阳性对比,并使用无监督机器学习算法(K-means和K-medoids聚类)确定粒子数量、质心和方向。

主要发现

  • 优化的定量磁化率成像方法生成的粒子阳性对比与钙化显著不同,可通过阈值轻松区分。
  • 3T和1.5T磁共振数据估计的粒子质心与标准CT检测算法完全一致,最大差异为0.7 mm。
  • 估计的粒子方向与真实方向高度相关(R>0.98)。

临床怎么用

  • 作者认为,所提出的基于MRI的工作流程在复杂粒子配置上验证后,有潜力替代目前广泛应用的基于CT的工作流程。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: Permanent seed brachytherapy is an established treatment option for localized prostate cancer. Currently, post-implant dosimetry is performed on CT images despite challenging target delineation due to limited soft tissue contrast. This work aims to develop an MRI-only workflow for post-implant dosimetry of prostate brachytherapy seeds. MATERIAL AND METHODS: A prostate mimicking phantom containing twenty stranded I-125 dummy seeds and calcifications was constructed. A three-dimensional gradient-echo MR sequence was employed on 3T and 1.5T MR scanners. An optimized quantitative susceptibility mapping (QSM) technique was applied to generate positive contrast for the seeds and calcifications. Seed numbers, centroids, and orientations were determined using unsupervised machine learning algorithms (K-means and K-medoids clustering). The geometrical seed positions and the resulting dose distribution were compared to the clinical CT-based approach. RESULTS: The optimized QSM-based method generated high quality positive contrast for the seeds that were significantly different from that for calcifications and could be easily differentiated by thresholding. The estimated seed centroids from both 3T and 1.5T MR data were in perfect agreement with the standard CT-based seed detection algorithm (maximum difference of 0.7 mm). The estimated seed orientations were highly correlated with the actual orientations (R > 0.98). CONCLUSIONS: The proposed MRI-based workflow enabling an accurate and robust means to localize the seeds (position and orientation) upon validation on complex seed configurations, has the potential to replace the current widely practiced CT-based workflow.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Nosrati R, Soliman A, Safigholi H, et al. MRI-based automated detection of implanted low dose rate (LDR) brachytherapy seeds using quantitative susceptibility mapping (QSM) and unsupervised machine learning (ML). Radiother Oncol. 2018. PMID: 30243670. doi:10.1016/j.radonc.2018.09.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。