基于前瞻性单机构混合分析的前列腺图像引导放疗中锥形束CT计划靶区边界定义的潜在失败风险
本研究在25例接受每日金标定位前列腺IGRT的患者中,比较kV-OBI金标匹配与kV-CBCT软组织匹配的摆位差异及分次内运动。CBCT-ST计算的前列腺PTV边界小于OBI-FM,但Bland-Altman分析显示显著比例偏倚,提示CBCT-ST可能低估真实PTV边界。作者认为确定可靠的CBCT-ST边界至少需一次辅助性金标评估以调查系统误差Σ和随机误差σ。
要点
- CBCT-ST与OBI-FM在前列腺摆位偏差上存在显著比例偏倚。
- CBCT-ST计算的前列腺PTV边界小于OBI-FM计算值,可能被低估。
- 确定可靠CBCT-ST边界至少需一次辅助性金标评估以调查系统误差Σ和随机误差σ。
核心结论
- CBCT-ST计算的前列腺PTV边界与真实PTV边界相比可能被低估。
- 为确定可靠的基于CBCT-ST的PTV边界,至少需要通过至少一次辅助性金标评估来调查在线匹配误差的系统误差Σ和随机误差σ。
研究设计与样本
- 前瞻性单机构混合分析,共分析25例接受前列腺IGRT并每日以植入金标定位的患者。
- 分析1177幅治疗前和治疗后kV-OBI图像,以及1177幅治疗前和206幅治疗后kV-CBCT图像。
- 比较kV-OBI金标匹配(OBI-FM)与kV-CBCT软组织匹配(CBCT-ST)之间的前列腺偏差和分次内运动差异,采用Bland-Altman一致性界限分析。
主要发现
- OBI-FM与CBCT-ST在前列腺前后、上下、左右方向上的平均差异分别为-0.43±1.45 mm、-0.09±1.65 mm、-0.12±0.80 mm,R²分别为0.85、0.88、0.83。
- CBCT-ST获得的分次内运动分别为0.00±1.46 mm、0.02±1.49 mm、0.15±0.64 mm,小于OBI-FM的0.43±1.90 mm、0.12±1.98 mm、0.26±0.80 mm,R²分别为0.42、0.33、0.16。
- Bland-Altman分析显示存在显著比例偏倚。
- CBCT-ST计算的PTV边界分别为1.5 mm、1.4 mm、0.9 mm,小于OBI-FM的3.15 mm、3.66 mm、1.60 mm。
- CBCT-ST的实际PTV边界经OBI-FM值补偿后分别为4.1 mm、4.8 mm、2.2 mm。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The purpose of this study was to evaluate the impact of markerless on-board kilovoltage (kV) cone-beam computed tomography (CBCT)-based positioning uncertainty on determination of the planning target volume (PTV) margin by comparison with kV on-board imaging (OBI) with gold fiducial markers (FMs), and to validate a methodology for the evaluation of PTV margins for markerless kV-CBCT in prostate image-guided radiotherapy (IGRT). METHODS: A total of 1177 pre- and 1177 post-treatment kV-OBI and 1177 pre- and 206 post-treatment kV-CBCT images were analyzed in 25 patients who received prostate IGRT with daily localization by implanted FMs. Intrafractional motion of the prostate was evaluated between each pre- and post-treatment image with these two different techniques. The differences in prostate deviations and intrafractional motions between matching by FM in kV-OBI (OBI-FM) and matching by soft tissues in kV-CBCT (CBCT-ST) were compared by Bland-Altman limits of agreement. Compensated PTV margins were determined and compensated by references. RESULTS: Mean differences between OBI-FM and CBCT-ST in the anterior to posterior (AP), superior to inferior (SI), and left to right (LR) directions were - 0.43 ± 1.45, - 0.09 ± 1.65, and - 0.12 ± 0.80 mm, respectively, with R2 = 0.85, 0.88, and 0.83, respectively. Intrafractional motions obtained from CBCT-ST were 0.00 ± 1.46, 0.02 ± 1.49, and 0.15 ± 0.64 mm, respectively, which were smaller than the results from OBI-FM, with 0.43 ± 1.90, 0.12 ± 1.98, and 0.26 ± 0.80 mm, respectively, with R2 = 0.42, 0.33, and 0.16, respectively. Bland-Altman analysis showed a significant proportional bias. PTV margins of 1.5 mm, 1.4 mm, and 0.9 mm for CBCT-ST were calculated from the values of CBCT-ST, which were also smaller than the values of 3.15 mm, 3.66 mm, and 1.60 mm from OBI-FM. The practical PTV margin for CBCT-ST was compensated with the values from OBI-FM as 4.1 mm, 4.8 mm, and 2.2 mm. CONCLUSIONS: PTV margins calculated from CBCT-ST might be underestimated compared to the true PTV margins. To determine a reliable CBCT-ST-based PTV margin, at least the systemic error Σ and the random error σ for on-line matching errors need to be investigated by supportive preliminary FM evaluation at least once.
引用
- Hirose K, Sato M, Hatayama Y, et al. The potential failure risk of the cone-beam computed tomography-based planning target volume margin definition for prostate image-guided radiotherapy based on a prospective single-institutional hybrid analysis. Radiat Oncol. 2018. PMID: 29880006. doi:10.1186/s13014-018-1043-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。