四维锥形束CT测得的目标运动低估了实际运动

4-Dimensional Cone Beam Computed Tomography-Measured Target Motion Underrepresents Actual Motion.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2018-04-27

研究对8例肺癌患者植入电磁Calypso信标实时测量肿瘤运动,比较4DCBCT重建(FDK或PICCS算法,内部Calypso信号或外部呼吸信号)所代表的运动。结果显示4DCBCT在左右、头脚、前后方向分别有48%、25%、40%的时间低估实际运动,平均低估1.7 mm、4.2 mm、2.5 mm,重建算法与呼吸信号选择对重建运动范围无影响。

要点

  • 4DCBCT在左右、头脚、前后方向分别有48%、25%、40%的时间低估实际靶区运动。
  • 左右、头脚、前后方向平均低估分别为1.7 mm、4.2 mm、2.5 mm。
  • FDK与PICCS算法、Calypso与bellows信号在各方向的重建运动范围均无显著差异。

核心结论

  • 所有4DCBCT扫描均低估了实时靶区运动。
  • 4DCBCT重建所选的重建算法与呼吸信号对重建的运动范围没有影响。

研究设计与样本

  • 8例肺癌患者的气道内植入电磁Calypso信标,实现实时运动测量。
  • 对8例患者中1至2个分次的治疗摆位CBCT投影进行4DCBCT重建,分别采用FDK算法或PICCS方法,并结合内部实时Calypso信标轨迹或外部呼吸信号(bellows腰带)。
  • 将实时信标质心(靶区)运动与4DCBCT重建中分割的信标质心位置进行比较。

主要发现

  • 靶区运动被显著低估,在左右、头脚、前后方向分别有48%、25%、40%的时间4DCBCT运动超过实际运动。
  • 左右、头脚、前后方向的平均低估分别为1.7 mm、4.2 mm、2.5 mm。
  • 左右方向FDK与PICCS比较P=0.47,Calypso与bellows比较P=0.19;头脚方向FDK与PICCS比较P=0.49,Calypso与bellows比较P=0.22;前后方向FDK与PICCS比较P=0.62,Calypso与bellows比较P=0.34,均未显示差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Four-dimensional cone beam computed tomography (4DCBCT) facilitates verification of lung tumor motion before each treatment fraction and enables accurate patient setup in lung stereotactic ablative body radiation therapy. This work aims to quantify the real-time motion represented in 4DCBCT, depending on the reconstruction algorithm and the respiratory signal utilized for reconstruction. METHODS AND MATERIALS: Eight lung cancer patients were implanted with electromagnetic Calypso beacons in airways close to the tumor, enabling real-time motion measurements. 4DCBCT scans were reconstructed from projections for treatment setup CBCT for 1 to 2 fractions of 8 patients with the Feldkamp-Davis-Kress (FDK) algorithm or the prior image constrained compressed sensing (PICCS) method and internal real-time Calypso beacon trajectories or an external respiratory signal (bellows belt). The real-time beacon centroid ("target") motion was compared with beacon centroid positions segmented in the 4DCBCT reconstructions. We tested the hypotheses that (1) the actual target motion was accurately represented in the reconstructions and (2) the reconstruction/respiratory signal combinations performed similarly in the representation of the real-time motion. RESULTS: On average the target motion was significantly underrepresented and exceeded the 4DCBCT motion for 48%, 25%, and 40% of the time in the left-right (LR), superior-inferior (SI), and anterior-posterior (AP) directions, respectively. The average underrepresentation for the LR, SI, and AP direction was 1.7 mm, 4.2 mm, and 2.5 mm, respectively. No difference could be shown between the reconstruction algorithms or respiratory signals in LR direction (FDK vs PICCS: P = .47, Calypso vs bellows: P = .19), SI direction (FDK vs PICCS: P = .49, Calypso vs bellows: P = .22), and AP direction (FDK vs PICCS: P = .62, Calypso vs bellows: P = .34). CONCLUSIONS: The 4DCBCT scans all underrepresented the real-time target motion. The selection of the reconstruction algorithm and respiratory signal for the 4DCBCT reconstruction does not have an impact on the reconstructed motion range.

来源:Int J Radiat Oncol Biol Phys,查看原文。版权归原作者所有。

引用

  1. Steiner E, Shieh CC, Caillet V, et al. 4-Dimensional Cone Beam Computed Tomography-Measured Target Motion Underrepresents Actual Motion. Int J Radiat Oncol Biol Phys. 2018. PMID: 29907487. doi:10.1016/j.ijrobp.2018.04.056