提议在调强放疗质量保证流程中加入达成指数

To propose adding index of achievement (IOA) to IMRT QA process.
PubMed 文献 · Radiat Oncol · 2018-06-15

该研究建议在调强放疗质量保证中增加一套包含达成指数、热度过量指数和冷度不足指数的指标体系,以弥补仅用通过/不通过标准评估的不足。通过三维一维模拟剂量分布和8例头颈及5例前列腺临床容积旋转调强放疗病例进行可行性分析,并对30例头颈和30例前列腺患者比较计算与实测数据。结果显示该指数系统比γ方法更能区分计划误差水平,且在γ通过率相同或相似时提供更多信息。

要点

  • 达成指数测量剂量分布与计划的一致程度,热度和冷度指数分别量化过量和欠量的幅度。
  • 在γ通过率相同或相似的多计划间,所提指数能更好区分计划误差。
  • 指数系统不依赖于验收标准,有助于多机构研究中的结果沟通。

核心结论

  • 所提指数系统能更好地评估调强放疗质量保证结果,尤其当多个计划的γ通过率相同或相似时。

研究设计与样本

  • 研究采用三个一维假设剂量分布进行系统分析,并使用8例头颈和5例前列腺患者的临床容积旋转调强放疗质量保证病例评估可行性。
  • 进一步对30例头颈和30例前列腺患者共60例,比较计算与实测数据,并同时使用指数系统和γ方法进行质量保证分析并比较结果。

主要发现

  • γ评估在评价质量保证结果时的敏感性受所用容差标准水平限制。
  • 所提指数倾向于更好地按误差量区分计划,并可定量评估计划中处方剂量的热度和冷度。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: In intensity modulated radiation therapy (IMRT) quality assurance (QA), evaluation of QA result using a pass/non-pass strategy under an acceptance criterion often suffers from lack of information on how good the plan is in absolute manner. In this study, we suggested adding an index system, previously developed for dose painting technique, to current IMRT QA process for better understanding of QA result. METHODS: The index system consists of three indices, index of achievement (IOA), index of hotness (IOH) and index of coldness (IOC). As indicated by its name, IOA does measure the level of agreement. IOH and IOC, on the other hand, measure the magnitude of overdose and underdose, respectively. A systematic analysis was performed with three 1-dimensional hypothetical dose distributions to investigate the characteristics of the index system. The feasibility of the system was also assessed with clinical volumetric modulated arc therapy (VMAT) QA cases from 8 head & neck and 5 prostate patients. In both simulation studies, certain amount of errors was intentionally induced to each dose distribution. Furthermore, we applied the proposed system to compare calculated with actual measured data for a total of 60 patients (30 head & neck and 30 prostate cases). QA analysis was made using both the index system and gamma method, and results were compared. RESULTS: While the gamma evaluation showed limited sensitivity in evaluating QA result depending on the level of tolerance criteria used, the proposed indices tended to better distinguish plans in terms of the amount of errors. Hotness and coldness of prescribed dose in the plan could be evaluated quantitatively by the indices. CONCLUSIONS: The proposed index system provides information with which IMRT QA result would be better evaluated, especially when gamma pass rates are identical or similar among multiple plans. In addition, the independency of the index system on acceptance criteria would help making clear communications among readers of published articles and researchers in multi-institutional studies.

来源:Radiat Oncol,查看原文。版权归原作者所有。

引用

  1. Kim DS, Kim S, Kang SH, et al. To propose adding index of achievement (IOA) to IMRT QA process. Radiat Oncol. 2018. PMID: 29903025. doi:10.1186/s13014-018-1055-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。