头颈癌放疗固定体位下基于标志点的MRI几何畸变前瞻性分析
结果显示所有标志点中位畸变为1.06 mm(IQR 0.6-1.98),皮肤至皮肤周边标志点畸变高于中心标志点(2.0±1.9 vs. 1.2±1.3 mm,p<0.0001),观察者间ICC=0.84。结论认为MRI几何畸变误差小于2 mm,在放疗计划应用中临床意义不显著。
要点
- 头颈癌放疗固定体位下MRI几何畸变中位值为1.06 mm,误差小于2 mm。
- 皮肤至皮肤周边标志点畸变显著高于中心标志点(2.0±1.9 vs. 1.2±1.3 mm)。
- 左右方向与前后方向畸变无统计学差异(1.5±1.6 vs. 1.6±1.7 mm,p=0.4)。
核心结论
- 在头颈癌放疗计划应用中,MRI几何畸变误差小于2 mm,与金标准CT相比临床意义不显著。
研究设计与样本
- 通过测量解剖标志点(皮肤至皮肤、骨至骨、软组织至软组织)评估几何畸变,共获得430个标志点测量。
- 评估观察者间变异性并计算组内相关系数(ICC)。
主要发现
- 所有扫描中所有标志点的中位畸变为1.06 mm(IQR 0.6-1.98)。
- 左右方向与前后方向畸变无统计学差异(1.5±1.6 vs. 1.6±1.7 mm,p=0.4)。
- 皮肤至皮肤周边标志点畸变高于中心标志点(2.0±1.9 vs. 1.2±1.3 mm,p<0.0001)。
- 四位观察者测量的平均畸变分别为1.05、1.23、1.06和1.05 mm,观察者间无显著差异(p=0.4),ICC=0.84。
临床怎么用
- 在头颈癌放疗固定体位下使用MRI进行计划设计时,几何畸变误差小于2 mm,可支持MRI在放疗计划中的有效应用。
- 需关注皮肤至皮肤周边标志点的畸变较大(2.0±1.9 mm),在勾画周边靶区或危及器官时予以考虑。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Uncertainties related to geometric distortion are a major obstacle for effectively utilizing MRI in radiation oncology. We aim to quantify the geometric distortion in patient images by comparing their in-treatment position MRIs with the corresponding planning CTs, using CT as the non-distorted gold standard. METHODS: Twenty-one head and neck cancer patients were imaged with MRI as part of a prospective Institutional Review Board approved study. MR images were acquired with a T2 SE sequence (0.5 × 0.5 × 2.5 mm voxel size) in the same immobilization position as in the CTs. MRI to CT rigid registration was then done and geometric distortion comparison was assessed by measuring the corresponding anatomical landmarks on both the MRI and the CT images. Several landmark measurements were obtained including; skin to skin (STS), bone to bone, and soft tissue to soft tissue at specific levels in horizontal and vertical planes of both scans. Inter-observer variability was assessed and interclass correlation (ICC) was calculated. RESULTS: A total of 430 landmark measurements were obtained. The median distortion for all landmarks in all scans was 1.06 mm (IQR 0.6-1.98). For each patient 48% of the measurements were done in the right-left direction and 52% were done in the anteroposterior direction. The measured geometric distortion was not statistically different in the right-left direction compared to the anteroposterior direction (1.5 ± 1.6 vs. 1.6 ± 1.7 mm, respectively, p = 0.4). The magnitude of distortion was higher in the STS peripheral landmarks compared to the more central landmarks (2.0 ± 1.9 vs. 1.2 ± 1.3 mm, p < 0.0001). The mean distortion measured by observer one was not significantly different compared to observer 2, 3, and 4 (1.05, 1.23, 1.06 and 1.05 mm, respectively, p = 0.4) with ICC = 0.84. CONCLUSION: MRI geometric distortions were quantified in radiotherapy planning applications with a clinically insignificant error of less than 2 mm compared to the gold standard CT.
引用
- Prospective analysis of in vivo landmark point-based MRI geometric distortion in head and neck cancer patients scanned in immobilized radiation treatment position: Results of a prospective quality assurance protocol. Clin Transl Radiat Oncol. 2017. PMID: 29594224. doi:10.1016/j.ctro.2017.09.003
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。