量化3T磁共振系统残余畸变与患者感应磁化率畸变对前列腺癌放疗计划的影响

Quantifying the Effect of 3T Magnetic Resonance Imaging Residual System Distortions and Patient-Induced Susceptibility Distortions on Radiation Therapy Treatment Planning for Prostate Cancer.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2017-10-20

该研究利用3T磁共振扫描仪的系统残余畸变与患者感应磁化率畸变组合位移场,扭曲17例前列腺癌患者的CT图像,并在扭曲图像上优化VMAT计划后转移至原始CT计算新剂量分布。结果显示最大残余平均畸变为3.19 mm(径向距离25 cm处),最大平均患者感应磁化率偏移为5.8 mm(最低带宽122 Hz/像素),但扭曲与未扭曲计划间剂量差异小于0.5%,所有计划质量指标的均值差异90%置信区间均在等效区间(-0.5, 0.5)内。结论认为患者感应磁化率畸变大于系统残余畸变,但对剂量分布的影响不足0.5%。

要点

  • 3T磁共振患者感应磁化率畸变(最大平均5.8 mm)大于系统残余畸变(最大平均3.19 mm)。
  • 扭曲与未扭曲CT计划间剂量差异小于0.5%,所有计划质量指标均值差异90%CI在(-0.5, 0.5)内。
  • 基于3T磁共振的前列腺癌VMAT计划剂量受畸变影响在临床可接受范围内。

核心结论

  • 在研究的勾画区域内,3T闭孔磁共振扫描仪的患者感应磁化率畸变大于使用厂商提供的三维校正后的系统残余畸变,但两者所致剂量误差均低于0.5%。

研究设计与样本

  • 研究利用3T扫描仪的系统残余畸变与患者感应磁化率畸变组合位移场,扭曲17例前列腺癌患者的CT图像。
  • 在扭曲CT图像上优化VMAT计划,再将计划参数转移至原始患者CT图像计算新剂量分布。

主要发现

  • 使用最低带宽122 Hz/像素时,径向距离25 cm处最大残余平均畸变为3.19 mm。
  • 最大平均患者感应磁化率偏移为5.8 mm。
  • 扭曲与未扭曲治疗计划间剂量差异小于0.5%。
  • 所有研究计划质量指标的均值差异90%置信区间均在等效区间(-0.5, 0.5)内。

临床怎么用

  • 对于所研究的前列腺癌勾画区域,3T磁共振系统残余畸变与患者感应磁化率畸变对VMAT剂量分布的影响小于0.5%,提示在该条件下基于3T磁共振制定放疗计划时剂量误差在可接受范围。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To investigate the effect of magnetic resonance system- and patient-induced susceptibility distortions from a 3T scanner on dose distributions for prostate cancers. METHODS AND MATERIALS: Combined displacement fields from the residual system and patient-induced susceptibility distortions were used to distort 17 prostate patient CT images. VMAT dose plans were initially optimized on distorted CT images and the plan parameters transferred to the original patient CT images to calculate a new dose distribution. RESULTS: Maximum residual mean distortions of 3.19 mm at a radial distance of 25 cm and maximum mean patient-induced susceptibility shifts of 5.8 mm were found using the lowest bandwidth of 122 Hz per pixel. There was a dose difference of <0.5% between distorted and undistorted treatment plans. The 90% confidence intervals of the mean difference between the dCT and CT treatment plans were all within an equivalence interval of (-0.5, 0.5) for all investigated plan quality measures. CONCLUSIONS: Patient-induced susceptibility distortions at high field strengths in closed bore magnetic resonance scanners are larger than residual system distortions after using vendor-supplied 3-dimensional correction for the delineated regions studied. However, errors in dose due to disturbed patient outline and shifts caused by patient-induced susceptibility effects are below 0.5%.

来源:Int J Radiat Oncol Biol Phys,查看原文。版权归原作者所有。

引用

  1. Adjeiwaah M, Bylund M, Lundman JA, et al. Quantifying the Effect of 3T Magnetic Resonance Imaging Residual System Distortions and Patient-Induced Susceptibility Distortions on Radiation Therapy Treatment Planning for Prostate Cancer. Int J Radiat Oncol Biol Phys. 2017. PMID: 29229326. doi:10.1016/j.ijrobp.2017.10.021