3T磁共振下塑料间质施源器的图像畸变及其剂量学影响

Image Distortions on a Plastic Interstitial Computed Tomography/Magnetic Resonance Brachytherapy Applicator at 3 Tesla Magnetic Resonance Imaging and Their Dosimetric Impact.
PubMed 文献 · Int J Radiat Oncol Biol Phys · 2017-06-28

研究纳入11例宫颈癌患者,在3T MRI下量化塑料腔内/间质施源器(含塑料针)的畸变,并通过正反读出方向扫描和模拟最坏情况移位评估剂量影响。结果显示施源器位移绝对值第95百分位为0.2-0.75 mm,D98/D90差异最大4.7%/4.3%,膀胱/直肠最受照2 cm³剂量差异最大6.0%/6.3%。结论认为该塑料施源器畸变的剂量学影响有限,3T MRI下施源器重建可行。

要点

  • 3T MRI下塑料施源器位移绝对值第95百分位为0.2-0.75 mm。
  • 正反读出扫描中D98/D90与原计划差异最大4.7%/4.3%。
  • 膀胱/直肠最受照2 cm³剂量差异最大6.0%/6.3%。

核心结论

  • 该塑料施源器(含塑料针)在3T MRI下的畸变对剂量影响有限,用于近距离放疗计划的行施源器重建是可行的。

研究设计与样本

  • 基于11例宫颈癌患者,在临床MRI协议中增加3次扫描:多回波采集计算B0图量化场不均匀性,以及连续的正反读出方向临床序列扫描。

主要发现

  • 感兴趣体积(含宫内装置和卵圆体)的绝对位移第95百分位在全部患者中为0.2至0.75 mm。
  • 正反读出扫描与原计划的D98/D90差异最大为4.7%/4.3%。
  • 膀胱和直肠最受照2 cm³的剂量差异最大为6.0%/6.3%。

关键数值与推荐

  • 施源器位移绝对值第95百分位范围为0.2-0.75 mm。
  • D98/D90差异最大为4.7%/4.3%。
  • 膀胱/直肠最受照2 cm³剂量差异最大为6.0%/6.3%。

临床怎么用

  • 在3T MRI下使用该塑料腔内/间质施源器进行近距离放疗计划时,可基于临床序列进行施源器重建。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To quantify magnetic resonance imaging (MRI) distortions on a plastic intracavitary/interstitial applicator with plastic needles at a field strength of 3 T and to determine the dosimetric impact, using patient data. METHODS AND MATERIALS: For 11 cervical cancer patients, our clinical MRI protocol was extended with 3 scans. From the first scan, a multi-echo acquisition, a map of the magnetic field (B0) was calculated and used to quantify the field inhomogeneity. The expected displacements of the applicator were quantified for the clinical sequence using the measured field inhomogeneity and the clinical sequence's bandwidth. The second and third scan were our routine clinical sequence (duration: <5 minutes each), acquired consecutively using opposing readout directions. The displacement of the applicator between these scans is approximately twice the displacement due to B0 inhomogeneity. The impact of the displacement on the dose was determined by reconstructing the applicator on both scans. The applicator was then shifted and rotated the same distance as the observed displacement to create a worst-case scenario (ie, twice the actual displacement due to B0 inhomogeneity). Next, the dose to 98%/90% (D98/D90) of the clinical target volume at high risk, as well as the dose to the most irradiated 2 cm3 for bladder and rectum, were calculated for the original plan as well as the shifted plan. RESULTS: For a volume of interest containing the intrauterine device and the ovoids the 95th percentile of the absolute displacement ranged between 0.2 and 0.75 mm, over all patients. For all patients, the difference in D98/D90 in the opposing readout scans with the original plan was at most 4.7%/4.3%. For the dose to the most irradiated 2 cm3 of bladder/rectum, the difference was at most 6.0%/6.3%. CONCLUSIONS: The dosimetric impact of distortions on this plastic applicator with plastic needles is limited. Applicator reconstruction for brachytherapy planning purposes is feasible at 3 T MRI.

来源:Int J Radiat Oncol Biol Phys,查看原文。版权归原作者所有。

引用

  1. van Heerden LE, van Kesteren Z, Gurney-Champion OJ, et al. Image Distortions on a Plastic Interstitial Computed Tomography/Magnetic Resonance Brachytherapy Applicator at 3 Tesla Magnetic Resonance Imaging and Their Dosimetric Impact. Int J Radiat Oncol Biol Phys. 2017. PMID: 29280466. doi:10.1016/j.ijrobp.2017.06.016