文献速递胸部肿瘤放疗放疗物理与技术

肺癌纵隔淋巴结靶区在摆位成像与治疗实施间的分次内位移

Systematic intrafraction shifts of mediastinal lymph node targets between setup imaging and radiation treatment delivery in lung cancer patients.
PubMed 文献 · Radiother Oncol · 2017-12-16

结果显示纵隔淋巴结靶区存在系统性分次内基线位移,头侧平均2.2 mm、后侧平均1.0 mm,而呼吸运动幅度稳定。

要点

  • 纵隔淋巴结靶区在CBCT与治疗实施间存在头侧2.2 mm、后侧1.0 mm的系统性基线位移。
  • CBCT与治疗实施间的呼吸运动幅度平均差异在各方向均不超过0.2 mm。
  • 基于原发肿瘤在线匹配可能不足以完全覆盖纵隔淋巴结靶区的分次内位移。

核心结论

  • 纵隔淋巴结靶区在CBCT摆位成像与治疗实施之间存在头侧和后侧的系统性分次内基线位移,而分次内运动幅度保持稳定。

主要发现

  • 纵隔淋巴结标记物在CBCT与治疗实施间出现头侧平均2.2 mm(标准差1.8 mm)和后侧平均1.0 mm(标准差1.2 mm)的系统性基线位移。
  • CBCT与治疗实施间的平均呼吸运动幅度在各方向差异均在0.2 mm以内。

临床怎么用

  • 对纵隔淋巴结靶区行在线匹配时,需考虑头侧和后侧的系统性分次内基线位移,评估是否需要扩大靶区外放边界或采用更针对淋巴结靶区的验证成像。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: Internal target motion results in geometrical uncertainties in lung cancer radiotherapy. In this study, we determined the intrafraction motion and baseline shifts of mediastinal lymph node (LN) targets between setup imaging and treatment delivery. MATERIAL AND METHODS: Ten lung cancer patients with 2-4 fiducial markers implanted in LN targets received intensity-modulated radiotherapy with a daily setup cone-beam CT (CBCT) scan used for online soft-tissue match on the primary tumor. At a total of 122 fractions, 5 Hz fluoroscopic kV images were acquired orthogonal to the MV treatment beam during treatment delivery. Offline, the 3D trajectory of the markers was determined from their projected trajectory in the CBCT projections and in the intra-treatment kV images. Baseline shifts and changes in the respiratory motion amplitude between CBCT and treatment delivery were determined from the 3D trajectories. RESULTS: Systematic mean LN baseline shifts of 2.2 mm in the cranial direction (standard deviation (SD): 1.8 mm) and 1.0 mm in the posterior direction (SD: 1.2 mm) occurred between CBCT imaging and treatment delivery. The mean motion amplitudes during CBCT and treatment delivery agreed within 0.2 mm in all directions. CONCLUSIONS: Systematic cranial and posterior intrafraction baseline shifts between CBCT and treatment delivery were observed for mediastinal LN targets. Intrafraction motion amplitudes were stable.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Schmidt ML, Hoffmann L, Møller DS, et al. Systematic intrafraction shifts of mediastinal lymph node targets between setup imaging and radiation treatment delivery in lung cancer patients. Radiother Oncol. 2017. PMID: 29258694. doi:10.1016/j.radonc.2017.11.030

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。