金标kV正交成像在前列腺癌每日IG对齐中比软组织CBCT更精准

2D kV orthogonal imaging with fiducial markers is more precise for daily image guided alignments than soft-tissue cone beam computed tomography for prostate radiation therapy.
PubMed 文献 · Adv Radiat Oncol · 2017-05-04

结果显示kV-FM在AP、LR、SI方向的重测可重复性均优于ST-CBCT,误差为亚毫米级;ST-CBCT对齐变异导致正常结构剂量显著增加并降低PTV剂量。结论支持每日kV-FM成像以提高前列腺外照射剂量递送精度。

要点

  • kV-FM在AP、LR、SI方向的组内重测可重复性均优于ST-CBCT。
  • kV-FM重测误差为亚毫米级,ST-CBCT误差可达毫米级。
  • ST-CBCT对齐变异导致正常结构剂量增加并降低PTV剂量。

核心结论

  • 基于kV-FM的每日IG对齐在前列腺癌IMRT中比ST-CBCT更精准。
  • ST-CBCT对齐不一致引入的误差较大,支持每日kV-FM成像以提高外照射剂量递送精度。

研究设计与样本

  • 每张kV和CBCT图像由同一医师两次离线对齐至DRR或计划CT,评估组内重测可重复性。
  • 分析256张kV和142张CBCT图像,并由另一医师在图像子集上重复重测分析以验证结果。

主要发现

  • kV-FM的组内重测可重复性Pearson相关系数95%CI:AP 0.987-0.991,LR 0.955-0.969,SI 0.971-0.980。
  • ST-CBCT的组内重测可重复性Pearson相关系数95%CI:AP 0.804-0.877,LR 0.877-0.924,SI 0.791-0.869。
  • kV-FM重测误差:AP 0.4±0.7 mm,RL 0.4±1.0 mm,SI 0.5±0.7 mm。
  • ST-CBCT重测误差:AP 2.1±2.2 mm,LR 1.3±1.4 mm,SI 1.2±1.8 mm。
  • kV-FM与ST-CBCT的平均位移差异:AP 0.3±3.8 mm,LR -1.1±8.5 mm,SI -2.0±3.7 mm。

关键数值与推荐

  • 剂量体积直方图显示ST-CBCT IG对齐相关的重测变异导致许多患者正常结构剂量显著增加并降低PTV剂量。

临床怎么用

  • 前列腺癌每日IG-IMRT对齐可优先采用kV-FM成像以提高对齐可重复性。
  • 使用ST-CBCT进行每日对齐时需注意其重测变异可能增加正常结构剂量并降低PTV剂量。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The hypothesis is that 2-dimensional kV orthogonal imaging with fiducial markers (kV-FM) and soft-tissue cone beam computed tomography (ST-CBCT) are equally reproducible for daily positional alignments for image guided (IG) intensity modulated radiation therapy (IMRT) for prostate cancer. METHODS AND MATERIALS: Ten patients undergoing definitive treatment for prostate cancer with IG-IMRT were imaged daily with kV-FM and ST-CBCT. For each acquired kV and CBCT image, offline alignments to the digitally reconstructed radiograph or planning CT, respectively, were made twice by the same physician to assess intraobserver test-retest reproducibility. The 256 kV and 142 CBCT images were analyzed, and the test-retest analysis was performed again on a subset of images by another physician to verify the results. RESULTS: The results demonstrated that kV-FM had better intraobserver test-retest reproducibility in the anterior-posterior (AP; 95% confidence interval [CI] Pearson correlation coefficient [r], 0.987-0.991), left-right (LR; 95% CI r, 0.955-0.969), and superior-inferior (SI; 95% CI r, 0.971-0.980) directions for daily IG alignments compared with ST-CBCT (AP: 95% CI r, 0.804-0.877; LR: 95% CI r, 0.877-0.924; SI: 95% CI r, 0.791-0.869). Errors associated with intraobserver test-retest reproducibility were submillimeter with kV-FM (AP: 0.4 ± 0.7 mm; RL: 0.4 ± 1.0 mm; SI: 0.5 ± 0.7 mm) compared with ST-CBCT (AP: 2.1 ± 2.2 mm; LR: 1.3 ± 1.4 mm; SI: 1.2 ± 1.8 mm). The mean shift differences between kV-FM and ST-CBCT were 0.3 ± 3.8 mm for AP, -1.1 ± 8.5 mm for LR, and -2.0 ± 3.7 mm for SI. Dose-volume histograms were generated and showed that test-retest variability associated with ST-CBCT IG-alignments resulted in significantly increased dose to normal structures and a reduced planning target volume dose in many patients. CONCLUSIONS: The kV-FM-based daily IG alignment for IMRT of prostate cancer is more precise than ST-CBCT, as assessed by a physician's ability to reproducibly align images. Given the magnitude of the error introduced by inconsistency in making ST-CBCT alignments, these data support a role for daily kV imaging of FM to enhance the precision of external beam dose delivery to the prostate.