文献速递胸部肿瘤放疗放疗物理与技术

临床4D-CBCT通气测量:图像质量改进提高通气准确性

Ventilation measured on clinical 4D-CBCT: Increased ventilation accuracy through improved image quality.
PubMed 文献 · Radiother Oncol · 2017-11-06

本研究纳入20例非小细胞肺癌患者的4D-CBCT与4D-CT扫描,以4D-CT测得的通气为金标准,评估改进投影校正联合SART重建是否提高4D-CBCT通气测量准确性。结果显示,改进投影校正联合SART重建使平均相关系数从临床标准的0.21显著提高至0.34,连续三次CBCT平均后进一步提高至0.38。

要点

  • 改进投影校正联合SART重建可显著提高4D-CBCT通气测量与4D-CT的相关性。
  • 连续三次CBCT通气平均后相关系数进一步提高至0.38。
  • 该结果可能为提取放疗期间患者呼吸模式动态变化提供基础。

核心结论

  • 改进投影校正联合SART重建提高了4D-CBCT通气测量准确性,可能成为提取放疗期间患者呼吸模式动态变化的踏脚石。

研究设计与样本

  • 研究纳入20例非小细胞肺癌患者的4D-CBCT和4D-CT扫描。
  • 原始CBCT投影经标准或改进投影校正后,分别用FDK算法或SART算法重建。
  • 通气通过DIR计算的Jacobian测量,CBCT与CT比较采用Spearman相关。

主要发现

  • 改进投影校正联合SART重建的平均相关系数为0.34,显著高于临床标准的0.21。
  • 对连续三次CBCT扫描测得的通气取平均后,相关系数进一步提高至0.38。

临床怎么用

  • 在临床4D-CBCT通气测量中,可考虑采用改进投影校正联合SART重建以提高准确性。
  • 对连续多次CBCT通气测量取平均可能进一步改善与4D-CT的相关性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND PURPOSE: Ventilation measured on 4D cone-beam computed tomography (CBCT) using deformable image registration (DIR) may predict specific radiation sensitivity, but the measurement is affected by the current image quality. With 4D computed tomography (CT) measured ventilation acting as a gold standard the current study investigates if image improvements increase the accuracy of 4D-CBCT measured ventilation. MATERIAL AND METHODS: The study consists of 4D-CBCT and 4D-CT scans of 20 non-small-cell lung cancer patients. Raw CBCT projections were subjected to a standard or an improved projection correction and reconstructed by the common FDK-algorithm or the more advanced SART-algorithm. Ventilation was measured as Jacobians calculated from DIR and the comparison between CBCT and CT was done by Spearman correlation. RESULTS: A significant increase in the mean correlation was observed when combining improved projection correction and SART reconstruction (0.34) compared to the clinical standard (0.21). The correlation further increased when averaging ventilation measured from three successive CBCT scans (0.38). CONCLUSION: The study showed that the combination of improved projection correction and the SART reconstruction increased the accuracy of CBCT ventilation and this result can be a stepping stone to extract dynamic changes in respiration pattern of patients during radiotherapy.

来源:Radiother Oncol,查看原文。版权归原作者所有。

引用

  1. Jensen KR, Brink C, Hansen O, et al. Ventilation measured on clinical 4D-CBCT: Increased ventilation accuracy through improved image quality. Radiother Oncol. 2017. PMID: 29122361. doi:10.1016/j.radonc.2017.10.024

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。